Purpose: To investigate the potential patient risk and interactions between a prototype implantable pressure monitoring device and a 3T clinical magnetic resonance imaging (MRI) machine to guide device design towards MR Conditional safety approval.

Materials And Methods: The pressure monitor device contained a catheter-mounted piezo-resistive pressure sensor, rechargeable battery, wireless communication system, and inductive pickup coil. Standard testing methods were used to guide experiments to investigate static field induced force and torque, radiofrequency (RF)-induced heating, image artifacts, and the MR's effect on device function. The specific clinical application of intracranial pressure monitoring was considered. RF-induced heating experiments were supported by numerical modeling of the RF body coil, the device, and experimental phantom.

Results: Sensing catheter lead length and configuration was an important component of the device design. A short 150 mm length catheter produced a heating effect of less than 2°C and a long 420 mm length catheter caused heating of 7.2°C. Static magnetic field interactions were below standard safety risk levels and the MR did not interfere with device function; however, artifacts have the potential to interfere with image quality.

Conclusion: Investigation of MR interactions at the prototype stage provides useful implantable device design guidance and confidence that an implantable pressure monitor may be able to achieve MR Conditional safety approval.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jmri.24909DOI Listing

Publication Analysis

Top Keywords

implantable pressure
12
pressure monitoring
12
device design
12
device
9
monitoring device
8
interactions prototype
8
conditional safety
8
pressure monitor
8
rf-induced heating
8
device function
8

Similar Publications

Incision capsular bag herniation during subluxated lens surgery: mechanisms, management, and prevention.

Front Med (Lausanne)

August 2025

Eye Center, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, China.

Purpose: To investigate the mechanism, intraoperative characteristics, management, and prevention of incision capsular bag herniation (ICBH), a previously unreported complication during cataract surgery in eyes with lens subluxation.

Methods: A retrospective observational case series was conducted on five male patients who developed ICBH during phacoemulsification with intraocular lens (IOL) implantation between January 2019 and December 2024. Among 867 subluxated-lens surgeries performed during this period, the estimated incidence of ICBH was 0.

View Article and Find Full Text PDF

Triply periodic minimal surfaces have garnered significant interest in the field of biomaterial scaffolds due to their unique structural properties, including a high surface-to-volume (S/V) ratio, tunable permeability, and the potential for enhanced biocompatibility. Bone scaffolds necessitate specific features to effectively support tissue regeneration. This study examines the permeability and active cell proliferation area of advanced Triply Periodic Minimal Surface (TPMS) lattice structures, focusing on a novel lattice design.

View Article and Find Full Text PDF

Amorphous silicon resistors enable smaller pixels in photovoltaic retinal prosthesis.

J Neural Eng

September 2025

Hansen Experimental Physics Laboratory, Stanford University, 452 Lomita Mall, Stanford, California, 94305, UNITED STATES.

Clinical trials of the photovoltaic subretinal prosthesis PRIMA demonstrated feasibility of prosthetic central vision with resolution matching its 100 μm pixel width. To improve prosthetic acuity further, pixel size should be decreased. However, there are multiple challenges, one of which is related to accommodating a compact shunt resistor within each pixel that discharges the electrodes between stimulation pulses and helps increase the contrast of the electric field pattern.

View Article and Find Full Text PDF

Objective: The aim of this study was to ascertain whether a prosthetic meniscus adapts pressures to native tibial contact pressures in a canine stifle joint after total medial meniscectomy.

Study Design: Ex vivo biomechanical experimental study.

Sample Population: Seven cadaveric hindlimbs of seven large-breed dogs.

View Article and Find Full Text PDF

Background: To describe clinical features, risk factors and outcomes of out-of-the-bag (OTB) intraocular lens (IOL) dislocation in dead bag syndrome (DBS).

Methods: Retrospective review of a single-surgeon series of eyes with IOLs that developed OTB IOL dislocation, managed at Singapore National Eye Centre (January 2014-December 2021), with a minimum of 6 months of follow-up. Eyes with OTB IOL dislocation following secondary IOL implantation and intraoperative capsule complications were excluded.

View Article and Find Full Text PDF