Article Abstract

Background: The mercury thermometer (MT) was considered the reference standard for the evaluation of body temperature; however, since April 2009, it has no longer been available in Italy.

Objective: This study aimed to evaluate the accuracy of digital thermometer (DT) and galinstan thermometer (GT) in comparison with the MT.

Methods: We prospectively recruited 284 children (age, 1 month to 17 years; mean, 8.5 years) seen in the emergency department of a tertiary pediatric hospital between November and December 2010. For each patient, body temperature was measured sequentially in the right axilla in a randomized fashion using DT, GT, and MT. Fever was defined as an axillary temperature of 37.5°C or greater. The temperature readings with DT and GT were compared statistically with those of MT (reference standard).

Results: No statistically significant difference in mean temperature was found between MT and GT in pairwise comparison (P = 0.06), whereas significant differences were found between MT and DT (P < 0.001) and between GT and DT (P < 0.001). Sensitivity and specificity of DT were 67.5% and 98.0%, respectively. When considering children with a temperature of 39°C or higher, the false-negative rate was 65.4% with DT and 30.8% with GT.

Conclusions: Although both DT and GT had good specificity and positive predictive value compared with MT, GT had higher sensitivity and a lower rate of false-negative rates. Galinstan thermometer is more accurate in the measurement of body temperature compared with DT or MT.

Download full-text PDF

Source
http://dx.doi.org/10.1097/PEC.0b013e3182809c29DOI Listing

Publication Analysis

Top Keywords

body temperature
16
galinstan thermometer
12
thermometer accurate
8
measurement body
8
temperature
8
accurate digital
4
digital measurement
4
body
4
temperature children
4
children background
4

Similar Publications

Tires are complex polymeric materials composed of rubber elastomers (both natural and synthetic), fillers, steel wire, textiles, and a range of antioxidant and curing systems. These constituents are distributed differently among the various tire parts, which are classified based on their function and proximity to the rim. This study presents a rapid and sensitive approach for the characterization of tire components using mild thermal desorption/pyrolysis (TDPy) coupled to direct analysis in real-time mass spectrometry (DART-MS).

View Article and Find Full Text PDF

The Essence of Nature Can be the Simplest (6)-Lifespan: Determined by Extracellular Fenton Chemistry.

Chem Biodivers

September 2025

State Key Laboratory for Conservation and Utilization of Bio-Resources in Yunnan & Yunnan Key Laboratory of Basic Research and Innovative Application for Green Biological Production, Key Laboratory for Microbial Resources of the Ministry of Education, School of Life Sciences, Yunnan University, Kunm

Understanding the determinants of lifespan is a central objective in biology. Lifespan is shaped by dynamic, stage-specific changes in metabolism, energy allocation, and genome integrity. Heart rate serves as a physiological marker that reflects both life stage and metabolic state.

View Article and Find Full Text PDF

Severe burns are a major global health concern, and are associated with long-term physical and psychological impairments, multi-organ dysfunction, and substantial morbidity and mortality. While burn injuries in adults trigger systemic immuno-metabolic alterations-characterized by white adipose tissue browning, elevated resting energy expenditure, widespread catabolism, and inflammation-these adaptive responses are considerably impaired in older adults, with molecular mechanisms behind these differences remaining largely unclear. As a key regulator of systemic metabolism, investigating the pathological role of adipose tissue (AT) postburn may reveal novel targets that could potentially improve patient outcomes.

View Article and Find Full Text PDF

Purpose: To analyze the effects of ablation interruption on ablation depths and clinical refractive outcomes to characterize the impact of ambient temperature changes and ablation interruption on ocular surface temperature (OST) during excimer laser ablation.

Methods: This prospective study was conducted on laser ablations in polymethylmethacrylate (PMMA) plates and porcine corneas to simulate laser in situ keratomileusis (LASIK) treatments using the EX500 laser (Alcon Laboratories, Inc) at ambient temperatures of 18, 20, and 22 °C. Ablation interruption was performed for 1, 2, 3, 4, and 5 seconds at the 10th second of the treatment of -9.

View Article and Find Full Text PDF

Volatile organic compounds (VOCs) are carbon-based chemicals characterized by high vapor pressure and low boiling points under standard temperature and pressure conditions. VOCs are categorized as exogenous or endogenous, depending on their source. Endogenous VOCs are metabolic byproducts eliminated via respiration.

View Article and Find Full Text PDF