Publications by authors named "Bu S Park"

Objective: To enhance RF safety when implantable medical devices are located within the body coil but outside the imaging region by using a secondary resonator (SR) to reduce electric fields, the corresponding specific absorption rate (SAR), and temperature change during MRI.

Materials And Methods: This study was conducted using numerical simulations with an American Society for Testing and Materials (ASTM) phantom and adult human models of Ella and Duke from Virtual Family Models, along with corresponding experimental results of temperature change obtained using the ASTM phantom. The circular SR was designed with an inner diameter of 150 mm and a width of 6 mm.

View Article and Find Full Text PDF

Objective: To improve sensitivity and uniformity of MR images obtained using a phased array RF coil, an inductively coupled secondary resonator with RF detuning circuits at 300 MHz was designed.

Materials And Methods: A secondary resonator having detuning circuits to turn off the resonator during the transmit mode was constructed. The secondary resonator was located at the opposite side of the four-channel phased array to improve sensitivity and uniformity of the acquired MR images.

View Article and Find Full Text PDF

Objective: Improve F magnetic resonance imaging uniformity of perfluorocarbon (PFC)-labeled cells by using a secondary inductive resonator tuned to 287 MHz to enhance the induced radio frequency (RF) magnetic field (B) at 7.05 T.

Materials And Methods: Following Faraday's induction law, the sign of induced B made by the secondary resonator can be changed depending on the tuning of the resonator.

View Article and Find Full Text PDF

This study describes the MRI-related radio frequency (RF) safety evaluation of breast tissue expander devices to establish safety criteria. Numerical simulations and experimental measurements were performed at 64 MHz with a gel phantom containing a breast expander. Additionally, computational modeling was performed (64 and 128 MHz) with an adult female model, containing a virtually implanted breast tissue expander device for four imaging landmark positions.

View Article and Find Full Text PDF

This study investigates the use of pads with high dielectric constant (HDC) materials to alter electromagnetic field distributions in patients during magnetic resonance imaging (MRI). The study was performed with numerical simulations and phantom measurements. An initial proof-of-concept and validation was performed using a phantom at 64 MHz, showing increases of up to 10% in electromagnetic field when using distilled water as the high dielectric material.

View Article and Find Full Text PDF

Materials with high dielectric constant (HDC) have been used in high field MRI to decrease specific absorption rate (SAR), increase magnetic field intensity, and increase signal-to-noise ratio. In previous studies, the HDC materials were placed inside the RF coil decreasing the space available. This study describes an alternative approach that considers an HDC-based sleeve placed outside the RF coil.

View Article and Find Full Text PDF

Rapamycin, produced by the soil bacterium Streptomyces hygroscopicus, has the ability to suppress the immune system and is used as an antifungal, anti-inflammatory, antitumor, and immunosuppressive agent. In an attempt to increase the productivity of rapamycin, mutagenesis of wild-type Streptomyces hygroscopicus was performed using ultraviolet radiation, and the medium composition was optimized using glycerol (which is one of the cheapest starting substrates) by applying Plackett-Burman design and response surface methodology. Plackett-Burman design was used to analyze 14 medium constituents: M100 (maltodextrin), glycerol, soybean meal, soytone, yeast extract, (NH4)2SO4, L-lysine, KH2PO4, K2HPO4, NaCl, FeSO4·7H2O, CaCO3, 2-(N-morpholino) ethanesulfonic acid, and the initial pH level.

View Article and Find Full Text PDF

We present a quadrature volume coil designed for brain imaging of a macaque monkey fixed in a sphinx position (facing down the bore) within a stereotactic frame at 3 T, where the position of the monkey and presence of the frame preclude use of existing coils. Requirements include the ability to position and remove the coil without disturbing the position of the monkey in the frame. A saddle coil and a solenoid were combined on a modified cylindrical former and connected in quadrature as to produce a homogeneous circularly polarized field throughout the brain.

View Article and Find Full Text PDF

Direct imaging of a histological slice is challenging. The vast difference in dimension between planar size and the thickness of histology slices would require an RF coil to produce a uniform RF magnetic (B1) field in a 2D plane with minimal thickness. In this work a novel RF coil designed specifically for imaging a histology slice was developed and tested.

View Article and Find Full Text PDF