Repeatability and reproducibility of anterior segment measurements with a new full-range SS-OCT and comparison with an AS-OCT.

Photodiagnosis Photodyn Ther

Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University; NHC Key laboratory of Myopia and Related Eye Diseases; Key Laboratory of Myopia and Related Eye Diseases, Chinese Academy of Medical Sciences, Shanghai, China; Shanghai Research Center of Ophthalmology and Optometry

Published: August 2025


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: To evaluate the intraobserver repeatability and interobserver reproducibility of anterior segment parameters measured by the new YG-100K full-range swept-source optical coherence tomographer (SS-OCT) and compare them with values from the validated CASIA 2 anterior segment optical coherence tomographer (AS-OCT).

Methods: In a randomized sequence, the first operator conducted three consecutive measurements with both devices, while a second operator independently conducted three consecutive measurements with only the YG-100K. Each device's built-in software automatically identified the scleral spur and angle recess. Measured parameters included central corneal thickness (CCT), anterior chamber depth (ACD), angle-to-angle (ATA) distance, anterior chamber width (ACW), lens vault (LV), crystalline lens rise (CLR), lens thickness (LT), trabecular iris angle (TIA), angle open distance (AOD) and trabecular iris space area (TISA). The within-subject standard deviation (Sw), test-retest (TRT) repeatability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated. Bland-Altman plots were assessed inter-device agreement.

Results: The YG-100K demonstrated excellent repeatability and reproducibility, evidenced by low Sw, TRT, and CoV values, and ICCs nearing 1.000 (intraobserver ICCs: 0.918-0.998; interobserver ICCs: 0.972-0.998). Comparative analysis revealed good agreement for CCT and ACD. However, ATA and ACW were significantly lower, whereas LT, LV and CLR were slightly higher with YG-100K. Angle measurements between the two devices showed wide 95 % limits of agreement (LoA).

Conclusions: This new full-range SS-OCT exhibited high repeatability and reproducibility for anterior segment measurements, confirming its reliability. However, significant differences and wide 95 % LoAs compared to the CASIA 2 suggested that these devices were not interchangeable.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.pdpdt.2025.104763DOI Listing

Publication Analysis

Top Keywords

anterior segment
16
repeatability reproducibility
12
reproducibility anterior
12
segment measurements
8
full-range ss-oct
8
optical coherence
8
coherence tomographer
8
conducted three
8
three consecutive
8
consecutive measurements
8

Similar Publications

To evaluate long‑term outcomes of corneal patch grafting (CPG) and to determine prognostic factors for anatomical and functional success. This retrospective study included 35 eyes from 35 patients who underwent CPG between April 2016 and September 2022 at Adana City Training and Research Hospital. Collected data included age, sex, preoperative and postoperative best-corrected visual acuity (BCVA), graft localization and size, anterior segment findings, graft survival, secondary surgical procedures, and rates of anatomical and functional success.

View Article and Find Full Text PDF

Background: The practice patterns for patient positioning, surgical techniques, and challenges faced by ophthalmologists during eye surgery on patients with kyphosis in India are yet unknown.

Methods: A cross-sectional online survey was conducted through Google Forms amongst practicing ophthalmic surgeons over two months and communicated across email lists and social media networks of state and regional ophthalmological associations of India in 2022.

Results: Fifty-two ophthalmologists responded (mean age 48.

View Article and Find Full Text PDF

Background: The m.3243A>G mutation in the MT-TL1 gene is the most common mtDNA mutation. The mutation can lead to a spectrum of conditions, including diabetes, hearing loss, heart and muscle involvement, encephalopathy and epilepsy, gastrointestinal problems, and vision impairment, often occurring concurrently-collectively referred to as MELAS (mitochondrial encephalopathy lactic acidosis and stroke-like episodes) syndrome.

View Article and Find Full Text PDF

Background: Recent trials of large core thrombectomy have shown that our traditional understanding of infarct characteristics and reperfusion benefit may be incomplete for patients with acute ischemic stroke (AIS). The Alberta Stroke Program Early CT Score (ASPECTS) has wide inter-rater variability, and modern studies have also shown that reperfusion therapies can benefit some patients regardless of the ASPECTS. Reproducible imaging metrics that account for the degree of hypo-attenuation on non-contrast computed tomography (NCCT) may be better suited to guide treatments.

View Article and Find Full Text PDF

Background: Fragmented QRS complex (f-QRS) on a 12-lead electrocardiogram (ECG) has been associated with myocardial scars. However, its diagnostic accuracy for detecting myocardial scars assessed by cardiac magnetic resonance (CMR) imaging remains uncertain.

Objective: To evaluate the diagnostic performance of f-QRS for detecting myocardial scars assessed by 3.

View Article and Find Full Text PDF