Article Abstract

Statement Of Problem: While intraoral scanners (IOS) demonstrate high accuracy in single-tooth and partial-arch applications, their reliability in full-arch scans with varying edentulous spans remains uncertain. Increasing span length may compromise scanning accuracy and prosthetic outcomes.

Objectives: To assess the influence of edentulous span length on the accuracy of two IOS devices Primescan (Dentsply Sirona) and Trios 3 (3Shape) in complete-arch scenarios. The hypothesis was that accuracy would decline with increasing span length, with performance differences between scanners.

Material And Methods: An in vitro study was conducted using a maxillary complete-arch model with a central incisor implant. Twelve span-length scenarios (FA1-FA12) were created by sequentially removing adjacent teeth. Each condition was scanned 10 times using both IOS devices (n = 240). A laboratory scanner (Identica blue, MEDIT) provided reference scans. Trueness (vs. reference) and precision (intra-group) were evaluated using root mean square (RMS) error. Scans were aligned via automatic and local best-fit in Medit Design. Statistical analysis included two-way mixed ANOVA, repeated measures ANOVA, and linear regression with Bonferroni correction (α = 0.004).

Results: Trios 3 demonstrated lower overall RMS values (0.296 ± 0.043 mm) than Primescan (0.338 ± 0.049 mm), with significant differences in 9 of 12 scenarios (p < 0.001). Correlation with span length was stronger for Primescan (R² = 0.892) than Trios 3 (R² = 0.674). Both devices showed high reliability.

Conclusions: Increasing edentulous span length negatively impacts scanner accuracy. Devices demonstrating consistent performance are preferable for reliable full-arch digital workflows.

Download full-text PDF

Source
http://dx.doi.org/10.3290/j.qi.b6394659DOI Listing

Publication Analysis

Top Keywords

span length
24
edentulous span
12
influence edentulous
8
scanner accuracy
8
vitro study
8
increasing span
8
ios devices
8
span
6
length
6
accuracy
6

Similar Publications

The neural representations of prior information about the state of the world are poorly understood. Here, to investigate them, we examined brain-wide Neuropixels recordings and widefield calcium imaging collected by the International Brain Laboratory. Mice were trained to indicate the location of a visual grating stimulus, which appeared on the left or right with a prior probability alternating between 0.

View Article and Find Full Text PDF

Order parameters in membranes: Following Joachim Seelig's path.

Biochim Biophys Acta Biomembr

August 2025

Institut Européen de Chimie et Biologie, Univ. Bordeaux, CNRS, INSERM, IECB, US1, UAR 3033, F-33600 Pessac, France; Institute of Chemistry & Biology of Membranes & Nanoobjects, Univ. Bordeaux, CNRS, Bordeaux INP, CBMN, UMR 5248, F-33600 Pessac, France. Electronic address:

Following the publication of biological membrane models in the 1970s, Joachim Seelig was the first to experimentally demonstrate the dynamic nature of these membranes. He conducted the first ssNMR experiments to measure the order parameters of the CD (H) bond of lipids deuterium-labelled, showing a fairly fluid membrane interior. Since then, the order parameters of the CD, CH and CC bonds have been measured.

View Article and Find Full Text PDF

Background: Visuospatial working memory (VSWM) is critical for academic achievement, particularly in mathematics. The Corsi Block-Tapping Test (CBTT) is one of the most widely used tools to assess VSWM, traditionally relying on span length as the primary performance indicator. However, recent research suggests that additional metrics, such as accuracy and reaction time (RT), may offer complementary insights.

View Article and Find Full Text PDF

We present a scaffold-level genome assembly from a specimen (encrusting sponge; Porifera; Demospongiae; Chondrillida; Halisarcidae). The genome sequence is 195.70 megabases in span.

View Article and Find Full Text PDF

Background: Carpometacarpal (CMC) arthroplasty can provide symptomatic relief and improvement in function in patients with CMC osteoarthritis. The study purpose was to identify demographic, 30-day outcome, and reimbursement trends in CMC arthroplasty between 2012 and 2021.

Methods: This was a retrospective cohort study of adult patients undergoing CMC arthroplasty from January 1, 2012, to December 31, 2021, using the National Surgical Quality Improvement Program database.

View Article and Find Full Text PDF