Article Abstract

Purpose: Spinal navigation systems require pre- and/or intra-operative 3-D imaging, which expose young patients to harmful radiation. We assessed a scoliosis-specific MRI-protocol that provides T2-weighted MRI and AI-generated synthetic-CT (sCT) scans, through deep learning algorithms. This study aims to compare MRI-based synthetic-CT spinal navigation to CT for safety and accuracy of pedicle screw planning and placement at thoracic and lumbar levels.

Methods: Spines of 5 cadavers were scanned with thin-slice CT and the scoliosis-specific MRI-protocol (to create sCT). Preoperatively, on both CT and sCT screw trajectories were planned. Subsequently, four spine surgeons performed surface-matched, navigated placement of 2.5 mm k-wires in all pedicles from T3 to L5. Randomization for CT/sCT, surgeon and side was performed (1:1 ratio). On postoperative CT-scans, virtual screws were simulated over k-wires. Maximum angulation, distance between planned and postoperative screw positions and medial breach rate (Gertzbein-Robbins classification) were assessed.

Results: 140 k-wires were inserted, 3 were excluded. There were no pedicle breaches > 2 mm. Of sCT-guided screws, 59 were grade A and 10 grade B. For the CT-guided screws, 47 were grade A and 21 grade B (p = 0.022). Average distance (± SD) between intraoperative and postoperative screw positions was 2.3 ± 1.5 mm in sCT-guided screws, and 2.4 ± 1.8 mm for CT (p = 0.78), average maximum angulation (± SD) was 3.8 ± 2.5° for sCT and 3.9 ± 2.9° for CT (p = 0.75).

Conclusion: MRI-based, AI-generated synthetic-CT spinal navigation allows for safe and accurate planning and placement of thoracic and lumbar pedicle screws in a cadaveric model, without significant differences in distance and angulation between planned and postoperative screw positions compared to CT.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00586-025-09151-xDOI Listing

Publication Analysis

Top Keywords

spinal navigation
16
postoperative screw
12
screw positions
12
safety accuracy
8
scoliosis-specific mri-protocol
8
ai-generated synthetic-ct
8
synthetic-ct spinal
8
planning placement
8
placement thoracic
8
thoracic lumbar
8

Similar Publications

The integration of mobile health (mHealth) technologies is transforming neurosurgery. Despite its potential, many uses remain unrealized due to the unique challenges and complexity of developing mHealth technology. While neurosurgeons bring invaluable clinical expertise and an understanding of patient needs, the technical intricacies of application development often require collaboration with developers and computer scientists, a process that can feel unfamiliar and difficult to navigate.

View Article and Find Full Text PDF

The purpose of this study is to propose a standardized classification of minimally invasive cervical pedicle screw (MICEPS) fixation according to the levels instrumented and the extent of the construct, thereby facilitating reproducible surgical planning and technique. We developed a three-tiered MICEPS classification with a specific surgical algorithm based on anatomic levels and construct length: Type 1, subaxial cervical fixation; Type 2, subaxial cervical to proximal thoracic fixation; and Type 3, subaxial cervical to T3/4 cervicothoracic stabilization. All techniques employ O-arm intraoperative navigation and preserve posterior tension-band integrity.

View Article and Find Full Text PDF

This editorial examines the feasibility and ethical implications of integrating robotic-assisted orthopaedic surgery into South Africa's healthcare system. While such technologies are well-established in high-income settings, enhancing surgical precision, reproducibility and postoperative outcomes, their widespread adoption in resource-limited contexts poses considerable logistical, financial and ethical challenges. The South African health landscape, marked by stark disparities between public and private sectors, provides a unique setting in which to assess whether such innovations should be prioritised over more pressing service delivery needs.

View Article and Find Full Text PDF

Expanding implant options in robot-assisted thoracolumbar spine surgery through novel customized instrumentation.

J Orthop

December 2025

Department of Spine Anesthesia, Manipal Comprehensive Spine Care Centre, Manipal Hospital, Old Airport Road, Bangalore, Karnataka, 560017, India.

Background: Robot-assisted pedicle screw placement demonstrates superior accuracy over conventional techniques, yet proprietary ecosystems limit implant compatibility and surgical choice. Can a custom screwdriver enable third-party instrumentation while maintaining navigational precision within the Mazor X Stealth Edition robotic system?

Methods: This retrospective comparative single-center study analysed 100 consecutive patients undergoing robot-assisted thoracolumbar fusion using custom screwdriver instrumentation (July-December 2024) versus 100 historical controls using standard proprietary instrumentation (October 2023-June 2024). The custom screwdriver had shaft dimensions matched to the robotic arm guide and navigation tracker compatibility.

View Article and Find Full Text PDF

Purpose: The spinal surgery robot can provide precise preoperative or intraoperative surgical planning and guide the implementation of the surgical plan under real-time navigation monitoring. Researchers have recently begun exploring the use of robotic systems to assist in performing Percutaneous Endoscopic Lumbar Discectomy (PELD). However, prior studies have primarily focused on robotic-assisted transforaminal puncture.

View Article and Find Full Text PDF