A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Multi-Institutional Survey of 4-Year Ultrasound Quality Control Findings. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Objectives: Periodic quality control (QC) testing of ultrasound (US) imaging systems is essential to ensure and maintain image quality and safety. The study aims to analyze QC findings from medical physics annual surveys of modern clinical US systems in a multi-institutional survey.

Methods: QC results from annual surveys between 2018 and 2021 were retrospectively collected from 12 medical physicists from 11 institutions or consulting companies (hereafter referred to as sites). QC tests were classified as scanner- and transducer-related tests, where scanner-related tests also included evaluation of diagnostic workstation display monitors and routine QC programs if applicable. Test methodology and pass/fail criteria were established by each site. QC findings were defined as results requiring follow-up action by service engineers, vendors, or other personnel. The percentage of annual findings was calculated for each site and averaged across the years of data contributed. Furthermore, findings across all sites were aggregated for an overall analysis.

Results: QC data from a total of 1069 scanner tests and 4542 transducer tests were collected. The average annual percentage of scanner-related findings varied from 0 to 40.8% among all sites (first quartile 8.7%, median 18.2%, and third quartile 34.1%), while that of transducer-related findings ranged between 0.5% and 19.9% (first quartile 3.9%, median 7.2%, and third quartile 14.5%). For scanner-related tests, the top 3 categories of findings were associated with physical and mechanical integrity (171 findings, 74.7%), quantitative testing of the scanner display monitor (40 findings, 17.5%), and scanner port creating artifacts (10 findings, 4.4%). The top 3 findings for transducers related to uniformity and artifact (227 findings, 79.9%), physical and mechanical integrity (44 findings, 15.5%), and sensitivity in fundamental mode (9 findings, 3.2%).

Conclusions: Medical physics annual surveys revealed considerable actionable findings of modern US systems, which inform the status of current US QC practices in the United States and may guide future standardization and recommendation of QC tests.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jum.16763DOI Listing

Publication Analysis

Top Keywords

findings
16
annual surveys
12
quality control
8
medical physics
8
physics annual
8
scanner-related tests
8
third quartile
8
physical mechanical
8
mechanical integrity
8
tests
7

Similar Publications