Purpose: To address the unmet need for a cross-platform, multiparametric relaxometry technique to facilitate data harmonization across different sites.

Methods: A simultaneous T and T mapping technique, 3D quantification using an interleaved Look-Locker acquisition sequence with a T preparation pulse (3D-QALAS), was implemented using the open-source vendor-agnostic Pulseq platform. The technique was tested on four 3 T scanners from two vendors across two sites, evaluating cross-scanner, cross-software version, cross-site, and cross-vendor variability. The cross-vendor reproducibility was assessed using both the vendor-native and Pulseq-based implementations. A National Institute of Standards and Technology/International Society for Magnetic Resonance in Medicine system phantom and three human subjects were evaluated. The acquired T and T maps from the different 3D-QALAS runs were compared using linear regression, Bland-Altman plots, coefficient of variation (CV), and intraclass correlation coefficient (ICC).

Results: Pulseq-QALAS demonstrated high linearity (R = 0.994 for T, R = 0.999 for T) and correlation (ICC = 0.99 [0.98-0.99]) against temperature-corrected NMR reference values in the system phantom. Compared to vendor-native sequences, the Pulseq implementation showed significantly higher reproducibility in phantom T values (CV, 2.3% vs. 17%; p < 0.001), and improved T reproducibility (CV, 3.4% vs. 4.9%; p = 0.71, not significant). The Pulseq implementation reduced cross-vendor variability to a level comparable to cross-scanner (within-vendor) variability. In vivo, Pulseq-QALAS exhibited reduced cross-vendor variability, particularly for T values in gray matter with a twofold reduction in variability (CV, 2.3 vs. 5.9%; p < 0.001).

Conclusion: An identical implementation across different scanners and vendors, combined with consistent reconstruction and fitting pipelines, can improve relaxometry measurement reproducibility across platforms.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC12204796PMC
http://dx.doi.org/10.1002/mrm.30566DOI Listing

Publication Analysis

Top Keywords

multiparametric relaxometry
8
system phantom
8
vendor-agnostic multiparametric
4
relaxometry improves
4
improves cross-platform
4
cross-platform reproducibility
4
reproducibility purpose
4
purpose address
4
address unmet
4
unmet cross-platform
4

Similar Publications

Multi-compartment diffusion-relaxation MR signal representation in the spherical 3D-SHORE basis.

Comput Biol Med

September 2025

Laboratorio de Procesado de Imagen (LPI), ETSI Telecomunicación, Universidad de Valladolid, Valladolid, Spain. Electronic address:

Modelling the diffusion-relaxation magnetic resonance (MR) signal obtained from multi-parametric sequences has recently gained immense interest in the community due to new techniques significantly reducing data acquisition time. A preferred approach for examining the diffusion-relaxation MR data is to follow the continuum modelling principle that employs kernels to represent the tissue features, such as the relaxations or diffusion properties. However, constructing reasonable dictionaries with predefined signal components depends on the sampling density of model parameter space, thus leading to a geometrical increase in the number of atoms per extra tissue parameter considered in the model.

View Article and Find Full Text PDF

Background: Quantitative mapping of the longitudinal relaxation rate (R1=1/T1) is a major building block for several multiparametric MRI protocols intended for adaptive radiation therapy planning. The implementation of these protocols is challenging in anatomical sites that experience large physiological motion.

Purpose: To implement and validate a motion-resolved quantitative T1 mapping method on a 1.

View Article and Find Full Text PDF

Objectives: To assess the diagnostic performance of splanchnic hemodynamic parameters measured with 4D flow MRI for diagnosis of severe portal hypertension (PH; hepatic venous pressure gradient (HVPG) ≥ 12 mmHg), and prediction of liver decompensation in patients with chronic liver disease.

Materials And Methods: Our single-center prospective study (2018-2022) enrolled patients undergoing transjugular liver biopsy with wedge HVPG measurement for abdominal 4D flow MRI (MRI-HVPG interval = 28 ± 18 (1-78) days). 4D flow MRI was acquired at 1.

View Article and Find Full Text PDF

Anterior cruciate ligament (ACL) injuries lead to an increased risk of osteoarthritis (OA). However, efforts to diagnose OA before irreversible changes to the joint occur remain limited. In this work, we utilized both quantitative MRI (qMRI) and displacements under applied loading MRI (dualMRI) to determine if relaxometry measures derived from qMRI and strains derived from dualMRI correlate with patient-reported outcomes at six months post unilateral ACL reconstruction.

View Article and Find Full Text PDF

Exploring scenarios for implementing fast quantitative MRI.

Eur J Radiol Open

June 2025

Copernicus Institute of Sustainable Development, Innovation Studies Section, Utrecht University, the Netherlands.

Purpose: MRI waitlists and discomfort from long scanning sessions are significant problems in clinical radiology. Novel multiparametric quantitative MRI techniques (qMRI) for radiological imaging enable acquisition of full-brain data within minutes to address these problems. While technical and clinical work is advancing, there has been limited research on implementing fast qMRI.

View Article and Find Full Text PDF