Background: To evaluate how the Heli-FX EndoAnchor system (Medtronic Vascular, Santa Rosa, CA, USA) influences the Real Achieved Sealing Zone (RASZ) compared to Endovascular Aneurysm Repair (EVAR) with Endurant (Medtronic Vascular, Santa Rosa, CA, USA) alone in a matched and unmatched real-world cohort.

Methods: In a retrospective, single-center study (2015-2022), 221 Abdominal Aortic Aneurysm patients underwent Endurant stent graft (Medtronic Vascular, Santa Rosa, CA, USA) placement either alone (n = 190) or with adjunctive Heli-FX EndoAnchors (n = 31). Preoperative and postoperative computed tomography angiography scans were evaluated using a standardized measurement technique (closed-polygon area, center lumen line distances) to quantify the Target Anticipated Sealing Zone (TASZ) and the RASZ. Primary endpoints were changes in RASZ length and area compared to TASZ. Secondary endpoints included type 1A and any endoleak rates, overall reinterventions. Statistical analysis included unadjusted and adjusted Cox proportional hazard regression models, receiver operating characteristic (ROC) curves, and propensity score matching.

Results: Heli-FX patients presented significantly more "hostile" proximal sealing zones, featuring shorter TASZ length (11 vs. 17 mm; P = 0.003), higher infrarenal angulation (45° vs. 18°; P < 0.001). Despite this hostile anatomy, the Heli-FX group exhibited a substantial increase in RASZ area (+48.3 mm; P = 0.001) and length (+5.8 mm; P = 0.003) compared to standard EVAR patients. ROC analysis indicated distinct RASZ cutoffs for predicting type 1A endoleak in each group (27.7 mm and 4.5 mm for Heli-FX vs. 49.5 mm and 9.5 mm for Endurant). In the 24 TASZ-features-matched couples, Heli-FX patients showed increased RASZ area (+42 vs -31 mm over TASZ; P = 0.003), and length (+5 vs -3 mm over TASZ; P = 0.003).

Conclusion: EndoAnchors substantially improve RASZ length and surface area compared to EVAR with the same stent graft alone. Routine device-specific RASZ assessment could refine treatment algorithms and pave the way for more personalized and durable endovascular solutions.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.avsg.2025.04.119DOI Listing

Publication Analysis

Top Keywords

medtronic vascular
12
vascular santa
12
santa rosa
12
rosa usa
12
proximal sealing
8
endovascular aneurysm
8
aneurysm repair
8
sealing zone
8
rasz
8
stent graft
8

Similar Publications

Durable Long-Term Outcomes After Early Transcatheter Arterialization of the Deep Veins.

JACC Case Rep

September 2025

University Hospitals Harrington Heart & Vascular Institute, Cleveland, Ohio, USA. Electronic address:

Background: Up to 20% of patients with chronic limb-threatening ischemia (CLTI) are not eligible for conventional revascularization options despite having severe symptoms of nonhealing ulcers and gangrene. Transcatheter arterialization of the deep veins (TADV) has shown promising results in this challenging subset of patients.

Case Summary: We present the long-term outcomes of 4 patients with no-option CLTI who were at risk of major below-the-knee amputation and were treated with TADV with the LimFlow System (Inari Medical).

View Article and Find Full Text PDF

This report explores the challenges of using the Woven EndoBridge (WEB) (Terumo, Tokyo, Japan) for a middle cerebral artery bifurcation aneurysm and introduces a novel technical approach not previously described in the literature. In a 75-year-old male patient, a stiff microguidewire was used to guide a microcatheter to the distal branch of the middle cerebral artery bifurcation aneurysm to modify the bifurcation angle, ensuring the WEB's stable placement. The patient's postoperative course was uneventful, with magnetic resonance angiogram (MRA) confirming the aneurysm's complete disappearance.

View Article and Find Full Text PDF

Severe, untreated tricuspid regurgitation is associated with worse clinical outcomes. While isolated tricuspid valve (TV) surgery has been linked to poor long-term outcomes, transcatheter TV therapies, including edge-to-edge repair and transcatheter tricuspid valve replacement (TTVR), have emerged as effective alternatives and have been shown to improve outcomes, leading to their regulatory approval in the United States. Conduction system abnormalities are commonly seen among patients undergoing TTVR due to the close proximity of the atrioventricular node and the His bundle to the TV annulus.

View Article and Find Full Text PDF

The efficacy of the remote monitoring system following transcatheter aortic valve replacement.

Cardiovasc Revasc Med

August 2025

Department of Cardiothoracic Surgery, NYU Langone Health, NY, United States of America. Electronic address:

Background: Transcatheter aortic valve replacement (TAVR) has become a cornerstone in the management of aortic valve disease. However, delayed complications after hospital discharge and readmission remain in an issue following TAVR. We aimed to evaluate the impact of remote monitoring systems on clinical outcomes after TAVR.

View Article and Find Full Text PDF