Article Abstract

Introduction: Blood glucose monitoring meters (BGM) have not become redundant yet. The accuracy and precision of "GLUCOCARD S onyx," a new BGM with Bluetooth function, has been evaluated and proven to exceed the actual ISO 15197:2013/EN ISO 15197:2015 guidelines besides offering features for better patient safety and telemedicine.

Methods: 100 finger-prick whole blood samples from subjects with diabetes and 32 without diabetes were collected and measured with GLUCOCARD S onyx. Plasma blood glucose levels were measured using YSI2300 STAT PLUS as reference analyzer for comparison. The evaluation followed ISO 15197:2013, section 6.3 accuracy criteria. Furthermore, the MARD factor was calculated for the overall clinical important range (with n=132 samples).

Results: The performance of GLUCOCARD S onyx was evaluated according to ISO 15197:2013, revealing that 99.7% (598/600) of the results fell within ±15% or ±0.8 mmol/L (± 15 mg/dL) of difference over the total clinically relevant glucose range compared to the YSI2300 STAT PLUS. 100% (600/600) of the measurement results over the total range fell within Clark Error Grid Zone A. An overall mean absolute relative difference (MARD) factor of 4.15% was obtained; 5.05% for glucose <5.6 mmol/L (<100 mg/dL), and 3.65% for glucose ≥5.6 mmol/L (≥100 mg/dL).

Discussion: GLUCOCARD S onyx shows clinically satisfactory accuracy and reliability, even exceeding the ISO 15197:2013 criteria, for hypoglycemic cases with glucose critically low as <3.9 mmol/L (<70 mg/dL) and hyperglycemic cases with glucose ≥10.0 mmol/L (≥180 mg/dL). Healthcare organizations as well as manufacturers are aiming to offer new BGM systems that go beyond the ISO criteria and offer systems that can be consulted instead or besides CGM (Continuous Glucose Monitoring) in case of e.g. severe hypo- and/or hyperglycemic episodes. A MARD factor of 4.15% revealed an excellent system accuracy over the total clinically relevant glucose range. With additional user-friendly features, this BGM can be seen as a useful tool for efficient diabetes therapy, especially in the event of severe blood glucose fluctuations.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC12003385PMC
http://dx.doi.org/10.3389/fcdhc.2025.1465732DOI Listing

Publication Analysis

Top Keywords

blood glucose
12
glucose monitoring
8
"glucocard onyx"
8
iso 151972013/en
8
151972013/en iso
8
iso 151972015
8
patient safety
8
glucocard onyx
8
ysi2300 stat
8
iso 151972013
8

Similar Publications

Background: Experience with icodextrin use in children on long-term peritoneal dialysis is limited. We describe international icodextrin prescription practices and their impact on clinical outcomes: ultrafiltration, blood pressure control, residual kidney function (RKF), technique and patient survival.

Methods: We included patients under 21 years enrolled in the International Pediatric Peritoneal Dialysis Network (IPPN) between 2007 and 2024, on automated PD with a daytime dwell.

View Article and Find Full Text PDF

Acute ischemic stroke (AIS) remains a leading cause of mortality and long-term disability globally, with survivors at high risk of recurrent stroke, cardiovascular events, and post-stroke dementia. Statins, while widely used for their lipid-lowering effects, also possess pleiotropic properties, including anti-inflammatory, endothelial-stabilizing, and neuroprotective actions, which may offer added benefit in AIS management. This article synthesizes emerging evidence on statins' dual mechanisms of action and evaluates their role in reducing recurrence, improving survival, and mitigating cognitive decline.

View Article and Find Full Text PDF

Objectives: To evaluate the predictive role of carotid stiffening, quantified using ultrafast pulse wave velocity (ufPWV), for assessing cardiovascular risk in young populations with no or elevated cardiovascular risk factors (CVRFs).

Materials And Methods: This study enrolled 180 young, apparently healthy individuals who underwent ufPWV measurements. They were classified into three groups: the CVRF-free group (n = 60), comprising current non-smokers with untreated blood pressure < 140/90 mmHg, fasting blood glucose (FBG) < 7.

View Article and Find Full Text PDF

Background: Hemoglobin A1c (HbA1c) is a stable compound, which is an important indicator for diagnosing glycemia, evaluating blood glucose control in patients with diabetes, and guiding the formulation of treatment plans. However, the detection of HbA1c is easily affected by many factors, resulting in a false increase or decrease, which affects the accuracy of test results.

Methods: In this paper, two abnormal HbA1c results were detected in the laboratory, which did not match the blood glucose detected at the same time.

View Article and Find Full Text PDF

Background: Diabetes mellitus is still a major health problem affecting individuals all over the world. Type 1 diabetes mellitus occurs due to insulin deficiency resulting from the destruction of pancreatic β-cells. This study aimed to investigate how vitamin D reduces blood glucose levels and HbA1c.

View Article and Find Full Text PDF