Mandibular Implants: A Metamaterial-Based Approach to Reducing Stress Shielding.

Adv Healthc Mater

Engineering and Technology Institute Groningen, University of Groningen, Groningen, 9747AG, The Netherlands.

Published: May 2025


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Biomechanical complications, such as stress shielding, bone resorption, and reconstruction failure, are prevalently associated with solid titanium mandible reconstruction plates. This study evaluates the potential of metamaterial designs with porous gyroid microarchitectures, to enhance biomechanical stimulation and mitigate these complications. A novel metamaterial reconstruction plate is compared with solid titanium plates, both patient-specifically designed and fabricated from Ti6Al4 V alloy. Stress shielding is assessed through photoelasticity experiments and validated with finite element analysis (FEA). Transparent mandible models are loaded incrementally (0-1000 N) to analyze stress distributions in the implants, screws, and mandible segments. The metamaterial plate reduces stress concentrations in the distant mandibular regions from the defect, while increasing stress around the screws near the defect, favoring local mechanical stimulation. FEA confirms improved load distribution (p = 0.003). However, the metamaterial plate exhibited a lower load-bearing capacity, failing at 775 N, while the solid plate withstood 1800 N without failure. Yet, the metamaterial design effectively reduced stress shielding, thereby enhancing biomechanical function near critical mandibular regions. Hence, despite their reduced load-bearing capacity, they can, potentially, preserve bone integrity and prevent implant failure that should be validated in future (pre-)clinical studies.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC12083438PMC
http://dx.doi.org/10.1002/adhm.202500405DOI Listing

Publication Analysis

Top Keywords

stress shielding
16
solid titanium
8
metamaterial plate
8
mandibular regions
8
load-bearing capacity
8
stress
7
metamaterial
5
mandibular implants
4
implants metamaterial-based
4
metamaterial-based approach
4

Similar Publications

Evaluation of lung oxidative stress and inflammatory state using exhaled breath condensate analysis in early-life arsenic exposure.

J Breath Res

September 2025

Department of Anatomy, Physiology, and Cell Biology, , University of California Davis, School of Veterinary Medicine, Davis, California, 95616-5270, UNITED STATES.

Millions of people worldwide are exposed to environmental arsenic in drinking water, resulting in both malignant and nonmalignant diseases. Interestingly, early life exposure by itself is sufficient to produce higher incidences of these diseases later in life. Based on the delayed onset of disease, we hypothesized that early life arsenic exposure would also induce long-term alterations in the metabolic profile.

View Article and Find Full Text PDF

Sectionally nonlinearly functionally graded (SNFG) structures with triply periodic minimal surface (TPMS) are considered ideal for bone implants because they closely replicate the hierarchical, anisotropic, and porous architecture of natural bone. The smooth gradient in material distribution allows for optimal load transfer, reduced stress shielding, and enhanced bone ingrowth, while TPMS provides high mechanical strength-to-weight ratio and interconnected porosity for vascularization and tissue integration. Wherein, The SNFG structure contains sections with thickness that varies nonlinearly along their length in different patterns.

View Article and Find Full Text PDF

With the rapid advancement in autonomous vehicles, 5G and future 6G communications, medical imaging, spacecraft, and stealth fighter jets, the frequency range of electromagnetic waves continues to expand, making electromagnetic interference (EMI) shielding a critical challenge for ensuring the safe operation of equipment. Although some existing EMI shielding materials offer lightweight construction, high strength, and effective shielding, they struggle to efficiently absorb broadband electromagnetic waves and mitigate dimensional instability and thermal stress caused by temperature fluctuations. These limitations significantly reduce their service life and restrict their practical applications.

View Article and Find Full Text PDF

Background: Patients with secondary hip osteoarthritis due to developmental dysplasia of the hip (DDH) often have abnormal femoral morphology, making stem design critical for long-term outcomes. The FMS-anatomic stem previously demonstrated favourable mid-term results. Its successor, the Anatomic Fit stem, was developed with a reduced hydroxyapatite-coated area to enhance proximal load transfer and a narrower lateral flare to facilitate insertion.

View Article and Find Full Text PDF

Active films displayed substantial prospects to maintain quality of tropical fruits during storage and transportation. This study developed multifunctional composite films loaded with melatonin/carvacrol nanoemulsions (MCNE) in guar gum/pullulan polysaccharide (GP) matrixes. The SEM analysis showed that MCNE was uniformly dispersed in GP film matrixes, and formed dense and continuous phase structure.

View Article and Find Full Text PDF