Article Abstract

Objective: To evaluate the cyclic fatigue fracture resistance, torsional fatigue, and angular deflection of Reciproc Blue (RB), WaveOne Gold (WOG), One Reci (OR), and Plex RC-One (RC-One) instruments for managing root curvatures in simulated canals.

Methods: A sample size calculation determined 10 samples per group (5 instruments for the cyclic fatigue test and 5 for the torsional fatigue test) for detecting significant differences. The study included 40 NiTi instruments (25 mm) from four reciprocating systems groups: RB R25, WOG 25, OR 25, and RC-One 25. Instruments were inspected for defects. Bending tests assessed memory control. Static cyclic fatigue tests used an artificial 60° curved canal with motorized activation. Torsional fatigue followed ISO 3630-1, measuring torque and angular deflection until failure. Fractured surfaces were analyzed using automated image analysis and measured with Rhinoceros 8.0 software. Statistical analysis included central tendency and dispersion measures, followed by ANOVA to identify statistically significant differences between groups.

Results: The bending test showed RC-One retained shape after pressure, while RB exhibited moderate resistance. WOG and OR did not demonstrate satisfactory bending resistance. Analysis of cyclic fatigue revealed RC-One had the highest resistance, followed by RB, OR, and WOG (p<0.001). RB required the highest torque for fracture, followed by RC-One, WOG, and OR (p<0.001). WOG had the lowest deflection angle at fracture, followed by OR, RB, and RC-One (p=0.0399). SEM and intelligent automated image processing analysis showed RC-One and RB had wear zones indicative of slow fracture, while WOG and OR displayed 100% rapid fracture dimples. RC-One had 41.86% slow fracture and 54.14% rapid fracture, while RB had 17.83% slow and 82.17% rapid fracture. Torsion tests revealed similar features in RC-One and RB, supporting their high and similar torsional resistance and deflection, unlike WOG's lower resistance and OR's lowest resistance with the highest deflection.

Conclusion: The best performance in the bending test was obtained by RC-One, followed by RB, while most failures were observed in WOG and OR. RC-One showed significantly higher cyclic fatigue resistance in severe curvatures than RB, WOG, and OR. RB and RC-One exhibited very similar torsional resistance values, significantly higher than OR and WOG. RC-One displayed superior angular deflection results compared to the other instruments. (EEJ-2024-08-135).

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11971697PMC
http://dx.doi.org/10.14744/eej.2024.08860DOI Listing

Publication Analysis

Top Keywords

cyclic fatigue
20
angular deflection
12
torsional fatigue
12
rc-one instruments
8
fatigue test
8
resistance wog
8
fatigue
7
cyclic
5
instruments
5
rc-one
5

Similar Publications

Introduction: Screw loosening remains a frequent mechanical complication in implant-supported prostheses, primarily caused by the gradual loss of abutment-screw preload. The aim of this study was to evaluate the mechanical performance of CAD-CAM custom and stock abutments by measuring removal torque values (RTV) at multiple time points and assessing surface morphology by scanning electron microscopy (SEM), following prolonged loading up to 2 × 10 cycles.

Methods: Forty-four implant-abutment assemblies with an internal conical-hex connection were divided into two groups: Stock abutments (SA) and custom abutments (CA).

View Article and Find Full Text PDF

Objectives: This study aimed to assess the effect of cross-section geometry on the mechanical properties of nickel-titanium (NiTi) instruments by comparing two instruments with identical tip size, taper, and thermal treatment but differing in cross-section design.

Methods: One hundred four NiTi rotary instruments, being S-shaped and triangular cross-section, manufactured with Blueish thermal treatment, were tested (n = 52 per group). Differential scanning calorimetry was employed, and the metal mass volume and cross-section area were assessed.

View Article and Find Full Text PDF

Are endodontic nickel-titanium instruments indefectible? A fractographic analysis of a compound core manufacturing defect.

Eur J Oral Sci

August 2025

Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Institute of Science Tokyo, Tokyo, Japan.

The aim of this study is to present a compound core manufacturing defect identified in a nickel-titanium (NiTi) instrument via fractographic analysis during additional retrospective analysis after dynamic cyclic fatigue resistance testing. This study raises awareness of core manufacturing defects and fosters the development of high-quality NiTi instruments without defects for clinical application. A scanning electron microscope with an acceleration voltage of 5 kV was used for fractographic analysis after cyclic fatigue testing.

View Article and Find Full Text PDF

Mechanical and structural changes to the annulus fibrosus in response to Sub-failure cyclic loading.

Acta Biomater

August 2025

Marquette University, Milwaukee, WI, USA; Medical College of Wisconsin, Milwaukee, WI, USA; Zablocki Veterans Affairs Medical Center, Milwaukee, WI, USA. Electronic address:

This study aimed to quantify how repetitive tensile loading alters the mechanical and structural properties of the annulus fibrosus (AF). Mechanical changes were evaluated through a three-step protocol involving pre-damage characterization of dynamic and viscoelastic properties, damage induction using predetermined loading cycles (n=400, 1600, 6400, 12,800) to a specified strain magnitude (11 %, 20 %, 28 %, 44 %), and post-damage characterization of the same properties. Structural changes were assessed by subjecting tissue to damage cycles and staining with hematoxylin and eosin or fluorescing collagen hybridizing peptides (F-CHP).

View Article and Find Full Text PDF

This paper presents the development, modeling, and analysis of an autonomous active ankle prosthesis with two degrees of freedom (2-DoF), designed to reproduce movements in the sagittal (dorsiflexion/plantarflexion) and frontal (inversion/eversion) planes in order to enhance the stability and naturalness of the user's gait. Unlike most commercial prostheses, which typically feature only one active degree of freedom, the proposed device combines a lightweight mechanical design, a screw drive with a stepper motor, and a microcontroller-based control system. The prototype was developed using CAD modeling in SolidWorks 2024, followed by dynamic modeling and finite element analysis (FEA).

View Article and Find Full Text PDF