Article Abstract

Background: Ultra-high dose rate radiotherapy (UHDR) delivers therapeutic doses at rates >40 Gy/s in a fraction of a second, aiming to enhance the therapeutic ratio through the FLASH effect. The substantial increase in UHDR beam current poses serious challenges for conventional active dosimeters. Integrating current transformers (ICT) offer a nondestructive solution for accurate monitoring, enabling the type of fast transient readout that will be crucial for UHDR treatment verification.

Purpose: The aim of this study is to build and characterize a clinically deployable ICT system and to develop an accurate calibration methodology for absolute charge determination that is traceable to primary electrical standards.

Methods: The ICT was constructed from a Super MuMetal toroid, and its secondary winding was made from 50 Ω coaxial cable. A 3D-printed case with an internal conductive coating shields the toroid assembly from interference. The ICT readout involves a custom differential amplifier and a commercial flash analog-to-digital converter. The system was calibrated using a bespoke sub-µs current pulser in the range of (2 to 16) mA, which itself is traceable to electrical standards via an in-house built electrometer with a calibrated feedback capacitor. The performance of the ICT was evaluated as a milliampere-scale beam monitor against concurrent absorbed dose graphite calorimetry irradiation measurements acquired on a specially tuned medical accelerator for UHDR delivery.

Results: The ICT responses for nominal test pulses, generated by a function generator and a current pulser, exhibited accurate reproduction of rise and fall times within the 1 ns sampling frequency. A systematic droop effect of 0.6(1)%/µs was observed but is accounted for through the calibration chain. The calibration of the current pulser exhibited a repeatability typically better than 0.05%, with a slowly-varying leakage that can be subtracted using a linear regression of the leakage current. The ICT charge calibration demonstrated a repeatability in the range of (0.5 to < 0.05)% for charge per pulse values in the range of (0.5 to 50) nC, respectively. The ICT response showed a strong linear relationship (adj-R = 0.99997) to charge per pulse. The in-beam comparison with a graphite calorimeter demonstrated the effectiveness of the ICT as an online beam monitor, independent of pulse repetition frequency in the range of (25 to 200) Hz, reducing the mean excess (i.e., independent of accelerator output) calorimeter variation to 0.3% (0.1% standard error on the mean).

Conclusions: This work demonstrates the feasibility of accurately calibrating the ICT in terms of absolute charge and applying it as a clinically deployable monitoring system of mA-scale electron beams delivered by a medical accelerator.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11972034PMC
http://dx.doi.org/10.1002/mp.17653DOI Listing

Publication Analysis

Top Keywords

current pulser
12
medical accelerator
8
traceable primary
8
primary electrical
8
electrical standards
8
pulser exhibited
8
current
7
ict
7
electron beam
4
beam monitoring
4

Similar Publications

Coronary artery disease (CAD) is the third leading cause of disability and death globally. Intravascular ultrasound (IVUS) is the most commonly used imaging modality for the characterization of vulnerable plaques. The development of novel intravascular imaging and therapy devices requires dedicated open systems (e.

View Article and Find Full Text PDF

Background: Ultra-high dose rate radiotherapy (UHDR) delivers therapeutic doses at rates >40 Gy/s in a fraction of a second, aiming to enhance the therapeutic ratio through the FLASH effect. The substantial increase in UHDR beam current poses serious challenges for conventional active dosimeters. Integrating current transformers (ICT) offer a nondestructive solution for accurate monitoring, enabling the type of fast transient readout that will be crucial for UHDR treatment verification.

View Article and Find Full Text PDF

Electroporation, or the use of electric pulses to facilitate the intracellular delivery of DNA, RNA, and other molecules, is a well-established technique, that has been demonstrated to significantly augment the immunogenicity of DNA/mRNA vaccines and therapeutics. However, the clinical translation of traditional electroporators has been limited due to high costs, large size, complex user operation, and poor tolerability in humans due to nerve stimulation. In prior work, we introduced ePatch: an ultra-low-cost, handheld, battery-free electroporator employing a piezoelectric pulser coupled with a microneedle electrode array that showed enhanced immunogenic responses to an intradermal SARS-CoV-2 DNA vaccine in mice.

View Article and Find Full Text PDF

Cemented carbide used in the rotor of a mud pulser is subjected to the scouring action of solid particles and corrosive mud media for a long time, which causes abrasive wear and electrochemical corrosion. To improve the wear and corrosive resistance of cemented carbide, samples with different cobalt content (WC-5Co, WC-8Co, and WC-10Co) receive deep cryogenic treatment (DCT) at -196 °C for 2.5 h.

View Article and Find Full Text PDF

High-intensity nanosecond pulse electric fields (nsPEF) can preferentially induce various effects, most notably regulated cell death and tumor elimination. These effects have almost exclusively been shown to be associated with nsPEF waveforms defined by pulse duration, rise time, amplitude (electric field), and pulse number. Other factors, such as low-intensity post-pulse waveform, have been completely overlooked.

View Article and Find Full Text PDF