Article Abstract

Purpose: For the first time, three currently available 360° CZT-SPECT/CT cameras were compared under clinical conditions using phantom-based measurements.

Methods: A Tc- and a Lu-customized NEMA IEC body phantom were imaged with three different cameras, StarGuide (GE Healthcare), VERITON-CT versions 200 (V200) and 400 (V400) (Spectrum Dynamics Medical) under the same clinical conditions. Energy resolution and volumetric sensitivity were evaluated from energy spectra. Vendors provided the best reconstruction parameters dedicated to visualization and/or quantification, based on their respective software developments. For both Tc- and Lu-phantoms, noise level, quantification accuracy, and recovery coefficient (RC) were performed with 3DSlicer. Image quality metrics from an approach called "task-based" were computed with iQMetrix-CT on Tc visual reconstructions to assess, through spatial frequencies, noise texture in the background (NPS) and contrast restitution of a hot insert (TTF). Spatial resolution indices were calculated from frequencies corresponding to TTF and TTF.

Results: Despite the higher sensitivity of VERITON cameras and the enhanced energy resolution of the V400 (3.2% at 140 keV, 5.2% at 113 keV, and 3.6% at 208 keV), StarGuide presents comparable image quality. This highlights the need to differentiate sensitivity from count quality, which is influenced by hardware design (collimator, detector block) and conditions image quality as well as the reconstruction process (algorithms, scatter correction, noise regulation). For Tc imaging, the quantitative image optimization approach based on RC for StarGuide versus RC for V200 and V400 systems (RC/RC: 0.9/1.8; 0.5/0.9; 0.5/0.9 respectively-Ø37 mm). SR showed nearly equivalent spatial resolution performances across the different reconstructed images. For Lu imaging, the 113 keV imaging of the V200 and V400 systems demonstrated strong performances in both image quality and quantification, while StarGuide and V400 systems offer even better potential due to their ability to exploit signals from both the 113 and 208 keV peaks. Lu quantification was optimized according to RC for all cameras and reconstructions (1.07 ± 0.09-Ø37 mm).

Conclusions: The three cameras have equivalent potential for Tc imaging, while StarGuide and V400 have demonstrated higher potential for Lu. Dedicated visual or quantitative reconstructions offer better specific performances compared to the unified visual/quantitative reconstruction. The task-based approach appears to be promising for in-depth comparison of images in the context of system characterization/comparison and protocol optimization.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11502619PMC
http://dx.doi.org/10.1186/s40658-024-00684-6DOI Listing

Publication Analysis

Top Keywords

image quality
16
clinical conditions
12
v400 systems
12
360° czt-spect/ct
8
czt-spect/ct cameras
8
three cameras
8
energy resolution
8
spatial resolution
8
113 kev
8
208 kev
8

Similar Publications

The aim of the study was to evaluate the toxic metals (TMs) pollution, bioaccumulation and its potential health risk via consumption of different vegetables irrigated by different water sources released from industrial estates of Khyber Pakhtunkhwa. Water (fresh and waste), soil and vegetables samples were collected in triplicates and acid digested. Digestion of samples were followed by evaporation and filtration and then assessed for TMs via atomic absorption spectrophotometer.

View Article and Find Full Text PDF

Myocarditis is an inflammation of heart tissue. Cardiovascular magnetic resonance imaging (CMR) has emerged as an important non-invasive imaging tool for diagnosing myocarditis, however, interpretation remains a challenge for novice physicians. Advancements in machine learning (ML) models have further improved diagnostic accuracy, demonstrating good performance.

View Article and Find Full Text PDF

[The TAVI heart team].

Herz

September 2025

Department of Cardiology, Heart Center Leipzig, University of Leipzig, Strümpellstraße 39, 04289, Leipzig, Deutschland.

Since the earliest studies on transcatheter aortic valve implantation (TAVI), the heart team concept has been an integral component of treatment planning for patients with aortic valve stenosis (AS). The primary objective is to ensure patient-specific, guideline-based treatment through the structured involvement of all relevant medical disciplines. The TAVI heart team is strongly recommended with a class I indication in both European and US clinical guidelines.

View Article and Find Full Text PDF

Diagnostic tips for multi-phase post-mortem computed tomography angiography interpretation in upper gastro-intestinal bleeding.

Int J Legal Med

September 2025

University Center of Legal Medicine Lausanne-Geneva, University of Geneva, Geneva University Hospitals, Rue Michel-Servet 1, 1211, Geneva 4, Switzerland.

In the past 10 years, the Multi-phase Post-mortem Computed Tomography Angiography (MPMCTA) has considerably improved the quality and precision of postmortem diagnoses, particularly in cases with vascular implication. MPMCTA is known to have higher sensitivity for detecting the source of a hemorrhage than autopsy. Death by upper gastro-intestinal (GI) bleeding is not so uncommon in forensic practice.

View Article and Find Full Text PDF

Background: Stroke, frequently associated with carotid artery disease, is evaluated using carotid computed tomography angiography (CTA). Dual-energy CTA (DE-CTA) enhances imaging quality but presents challenges in maintaining high image clarity with low-dose scans.

Objectives: To compare the image quality of 50 keV virtual monoenergetic images (VMI) generated using Deep Learning Image Reconstruction (DLIR) and Adaptive Statistical Iterative Reconstruction-V (ASIR-V) algorithms under a triple-low scanning protocol in carotid CTA.

View Article and Find Full Text PDF