A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Long-term performance of subcutaneous implantable defibrillators in athletes: A multicenter, real-world analysis of sport activities from the iSUSI registry. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: No data regarding subcutaneous-implantable cardioverter defibrillator (S-ICD) technology in patients actively engaging in sports activities are available.

Objective: This study aims to compare S-ICD performance between athletes and nonathletes.

Methods: The primary outcome of the study was the comparison of overall device-related complications between athletes and nonathletes. Appropriate shocks, inappropriate shocks, and individual device-related complications were secondary outcomes.

Results: A total of 1493 patients were extracted from the International Subcutaneous Implantable Cardioverter Defibrillator Registry (iSUSI) registry, of whom 152 (10.2%) were athletes, mostly engaging in dynamic sports (54.2%). Brugada syndrome, myocarditis, and arrhythmogenic right ventricular cardiomyopathy (ARVC) were more common in athletes (11.2% vs 3.3%, P < .001; 19.1% vs 9.0%, P < .001; 8.6% vs 2.8%, P < .001, respectively). During a median follow-up time of 25.5 (12.0-41.2) months, athletes were more likely to experience appropriate shocks (yearly rate: 7.2 [4.9-10.7] % vs 4.3 [3.6-5.1] %, P = .028), occurring more frequently during exercise (3.9% vs 0.6%, P < .001). This finding lost significance when adjusting for confounders (adjusted hazard ratio [aHR] 1.440 [0.909-2.281], P = .120). No differences were found in overall device-related complications (yearly rate: 3.3% vs 3.4%, P = .448) and inappropriate shocks (yearly rate: 5.3% vs 3.7%, P = 0.111). Myopotential oversensing (4.0% vs 1.3%, P = .011) was more common in athletes, as were lead infections (3.3% vs 0.9%, P =.008), with the latter clustering in the early postimplantation period.

Conclusion: The S-ICD is a valid therapeutic option for preventing sudden cardiac death in athletes. Sports practice was not associated with an increased risk of complications or inappropriate shocks, although athletes are exposed to a higher risk of S-ICD infections in the early postoperative period.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2024.09.039DOI Listing

Publication Analysis

Top Keywords

device-related complications
12
inappropriate shocks
12
yearly rate
12
athletes
9
subcutaneous implantable
8
isusi registry
8
cardioverter defibrillator
8
appropriate shocks
8
common athletes
8
shocks yearly
8

Similar Publications