Article Abstract

Background: In MR-guided in-bore percutaneous needle interventions, typically 2D interactive real-time imaging is used for navigating the needle into the target. Misaligned 2D imaging planes can result in losing visibility of the needle in the 2D images, which impedes successful targeting. Necessary iterative manual slice adjustment can prolong interventional workflows. Therefore, rapid automatic alignment of the imaging planes with the needle would be preferable to improve such workflows.

Purpose: To investigate rapid 3D localization of needles in MR-guided interventions via a convolutional neural network (CNN)-based localization algorithm using an undersampled white-marker contrast acquisition for the purpose of automatic imaging slice alignment.

Methods: A radial 3D rf-spoiled gradient echo MR pulse sequence with white-marker encoding was implemented and a CNN-based localization algorithm was employed to extract position and orientation of an aspiration needle from the undersampled white-marker images. The CNN was trained using porcine tissue phantoms (257 needle trajectories, four-fold data augmentation, 90%/10% split into training and validation dataset). Achievable localization times and accuracy were evaluated retrospectively in an ex vivo study (109 needle trajectories) for a range of needle orientations between 78° and 90° relative to the B field. A proof-of-concept in vivo experiment was performed in two porcine animal models and feasibility of automatic imaging slice alignment was evaluated retrospectively.

Results: Ex vivo needle localization was achieved with a median localization accuracy of 1.9 mm (distance needle tip to detected needle axis) and a median angular deviation of 2.6° for needle orientations between 86° and 90° to the B field from fully sampled WM images (resolution of (4 mm), 6434 acquired radial k-space spokes, acquisition time of 80.4 s) in a field-of-view of (256 mm). Localization accuracy decreased with increasing undersampling and needle trajectory increasingly aligned with B. For needle orientations between 86° and 90° to the B field, a highly accelerated acquisition of only 32 k-space spokes (acquisition time of 0.4 s) yielded a median localization accuracy of 3.1 mm and a median angular deviation of 4.7°. For needle orientations between 78° and 82° to the B field, a median accuracy and angular deviation of 3.5 mm and 6.8° could still be achieved with 64 sampled spokes (acquisition time of 0.8 s). In vivo, a localization accuracy of 1.4 mm and angular deviation of 3.4° was achieved sampling 32 k-space spokes (acquisition time of 0.48 s) with the needle oriented at 87.7° to the B field. For a needle oriented at 77.6° to the B field, localization accuracy of 5.3 mm and angular deviation of 6.8° were still achieved sampling 128 k-space spokes (acquisition time of 1.92 s), allowing for retrospective slice alignment.

Conclusion: The investigated approach enables passive biopsy needle localization in 3D. Acceleration of the localization to real-time applicability is feasible for needle orientations approximately perpendicular to B. The method can potentially facilitate MR-guided needle interventions by enabling automatic imaging slice alignment with the needle.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.17376DOI Listing

Publication Analysis

Top Keywords

needle
22
needle orientations
20
localization accuracy
20
angular deviation
20
spokes acquisition
20
acquisition time
20
k-space spokes
16
localization
13
needle localization
12
slice alignment
12

Similar Publications

Background: Interventions aimed to increase healthcare provider empathy and capacity to deliver person-centered care have been shown to improve healthcare seeking and outcomes. In the context of self-injectable contraception, empathetic counseling and coaching may be promising approaches for addressing "fear of the needle" among clients interested in using subcutaneous depot medroxyprogesterone (DMPA-SC). In Nigeria, the Delivering Innovation in Self-Care (DISC) project developed and evaluated an empathy-based in-service training and supportive supervision intervention for public sector family (FP) planning providers implemented in conjunction with community-based mobilization.

View Article and Find Full Text PDF

Rationale And Objectives: To investigate the performance of CT-guided percutaneous CNB for the diagnosis of pancreatic tumors using a blunt-tip needle technique.

Materials And Methods: This is a retrospective cohort study included 103 consecutive patients (64 males and 39 females; mean age 61±12.3 years;range 21-86) who underwent CT-guided percutaneous CNB of pancreatic lesions using a blunt-tip needle technique between October 2021 and October 2023.

View Article and Find Full Text PDF

A numerical investigation of the kinematic and fluid dynamic behaviour of an intramuscular autoinjector designed for optimising injection efficiency.

Med Eng Phys

October 2025

Department of Mechanical Engineering, University of Cape Town, 7701, South Africa; Centre for Research in Computational and Applied Mechanics (CERECAM), University of Cape Town, 7701, South Africa.

The usability and versatility of autoinjectors in managing chronic and autoimmune diseases have made them increasingly attractive in medicine. However, investigations into autoinjector designs require an understanding of the kinematic properties and fluid behaviour during injection. To optimise injection efficiency, this study develops a mathematical and computational fluid dynamics (CFD) model of an IM autoinjector by investigating the effects of viscosity, needle length, needle diameter, and medication volume on the injection process.

View Article and Find Full Text PDF

Objective: The purpose of this study was to compare nitrous oxide (N2O) vs virtual reality (VR) as methods for reducing pain and anxiety during a dental injection. The primary objectives were to assess acute changes in stress responses by comparing salivary cortisol levels between the 2 groups and differences in injection pain scores.

Methods: A total of 132 female subjects serving as their own control received maxillary lateral incisor infiltration injections with the use of either N2O or a VR headset during separate appointments spaced at least 2 weeks apart.

View Article and Find Full Text PDF