Purpose: A single treatment planning system (TPS) model for matched linacs provides flexible clinical workflows from patient treatment to intensity-modulated radiation therapy (IMRT) quality assurance (QA) measurement. Since general guidelines for building a single TPS model and its validation for matched linacs are not well established, we present our RayStation photon TPS modeling strategy for matched Elekta VersaHD linacs.

Method: The four linacs installed from 2013 to 2020 were matched in terms of Percent Depth Dose (PDD), profile, output factor and wedge factors for 6-MV, 10-MV, 15-MV, and 6-MV-FFF, and maintained following TG-142 recommendations until RayStation commissioning. The RayStation single model was built to represent all four linacs within the tolerance limits recommended by MPPG-5.a. The comprehensive validation tests were performed for one linac following MPPG-5.a and TG-119 guidelines, and spot checks for the other three. Our TPS modeling/validation method was evaluated by re-analyzing the previous 103 patient-specific IMRT/volumetric modulated arc therapy (VMAT) QA measurements with the calculated planar doses by the single model in comparison with the analysis results using four individual Pinnacle TPS models.

Results: For all energies, our single model PDDs were within 1% agreement of the four-linac commissioning measurements. The MPPG-5.a validation tests from 5.1 through 7.5 and all TG-119 measurements passed within the recommended tolerance limits. The IMRT QA results (mean ± standard deviation) for RayStation single model versus Pinnacle individual models were 98.9% ± 1.3% and 98.0% ± 1.4% for 6-MV, 99.9% ± 0.1% and 99.1% ± 1.9% for 10-MV, and 98.2% ± 1.3% and 97.9% ± 1.8% for 6-MV-FFF, respectively.

Conclusion: We successfully built and validated a single photon beam model in RayStation for four Elekta Linacs. The proposed new validation methods were proven to be both efficient and effective.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11466464PMC
http://dx.doi.org/10.1002/acm2.14485DOI Listing

Publication Analysis

Top Keywords

single model
16
single
8
single photon
8
photon beam
8
model
8
beam model
8
model raystation
8
matched elekta
8
elekta linacs
8
tps model
8

Similar Publications

Major ABO-Incompatible Platelet Transfusions Are Associated With Brain Ischemia After Intracerebral Hemorrhage.

Stroke

September 2025

Department of Neurology, Vagelos College of Physicians and Surgeons, Columbia University, New York. (F.C.P., M.R., M.S., A.K., S.G., S.A., S.P., J.C., D.J.R.).

Background: Major ABO-incompatible platelet transfusions are associated with poor intracerebral hemorrhage (ICH) outcomes, yet drivers for this relationship remain unclear. Brain magnetic resonance imaging (MRI) ischemic lesions after ICH are neuroimaging biomarkers of secondary brain injury and are associated with poor outcomes. Given that ABO-incompatible platelet transfusions can induce immune complex formation, thrombo-inflammation, and endothelial barrier disruption, factors that could exacerbate cerebral ischemia, we explored whether major ABO-incompatible platelet transfusions are risk factors for ischemic lesions on brain MRI after ICH.

View Article and Find Full Text PDF

A flexible linear circular bipolarization conversion metasurface based on graphene.

Phys Chem Chem Phys

September 2025

School of Electrical and Automation Engineering, Suzhou University of Technology, Suzhou, 215506, China.

A flexible bipolarization conversion metasurface based on graphene is proposed in this paper, which can achieve single-band linear-to-linear (LTL) and dual-band linear-to-circular (LTC) polarization conversion. The polarization conversion ratio (PCR) and axial ratio (AR) are dynamically regulated by varying the sheet resistance () of graphene. When = 1400 Ω Sq, the designed metasurface achieves a single-band LTL polarization conversion of 7.

View Article and Find Full Text PDF

Cefepime (FEP), a fourth-generation cephalosporin combined with tazobactam (TAZ), a β-lactamase inhibitor, is being developed by Wockhardt as a pharmacodynamically optimized fixed dose combination (FEP-2 g + TAZ-2 g) for the treatment of multidrug-resistant Gram-negative infections. To undertake an exposure-response analysis for establishing pharmacokinetic (PK)/pharmacodynamic (PD) targets, it is crucial to characterize the PK profile of compounds in surrogate compartments, such as plasma and lung, in clinically relevant animal infection models used to evaluate efficacy. In the current study, PKs of FEP and TAZ were assessed in plasma and in epithelial lining fluid (ELF) of neutropenic noninfected, lung-infected, and thigh-infected mice.

View Article and Find Full Text PDF

Objective: To describe the self-report instruments used to measure well-being in children with disabilities, investigate their psychometric quality, cognitive accessibility and alignment with Keyes's operationalization of well-being, including emotional, psychological and social aspects.

Methods: MEDLINE, ProQuest, PubMed and CINAHL were searched for articles published from 2011 to March 2023, identifying 724 studies. Synonyms provided by thesaurus on the main constructs: 'children', 'measure', 'disability' and 'mental health' were employed in the search strategy.

View Article and Find Full Text PDF

Stochastic model analysis of luminescence switching in hybrid systems of CdSe/CdS quantum dots-spiropyran molecular photoswitches.

Phys Chem Chem Phys

September 2025

Department of Chemistry, Graduate School of Science and Technology, Kwansei Gakuin University, 1 Gakuen Uegahara, Sanda, Hyogo 669-1330, Japan.

Hybrid systems (HSs) of quantum dots (QDs) and molecular photoswitches exhibit luminescence switching of QDs based on energy transfer and have garnered attention for their potential applications in sensors and optical memories. In HSs, the chemical composition, such as the number of attached ligands, is inherently distributed, posing challenges for extracting the energy transfer process from the QDs to a single acceptor molecule. The stochastic model, assuming a Poisson distribution for the number of acceptors, proves to be an effective approach for extracting the process.

View Article and Find Full Text PDF