Article Abstract

Objective: To understand possible reasons for poor durability of the Nellix (Endologix Inc., Irvine, USA) endovascular aneurysm sealing (EVAS) device.

Materials And Methods: 21 Nellix endoprostheses explanted for endoleaks and migration underwent visual examinations of stent structures and instrumental examinations of the polymer endobags on 4 devices. We harvested 2.0-gram polymer slices out of each of them and tested the samples in an in vitro implantation replication that included wet and dry exposures. During the wet phase, we placed samples in a beaker with saline, mimicking the filling of the endobags during implantation. An exposure to a 37°C environment with 60% humidity during the dry phase replicated the postimplantation conditions inside the aneurysmal sac.

Results: Iatrogenic defects affected 16 (76%) metal stents and 20 (95%) endobags. The polymer was disintegrated owing to degradation in 15 (71%) cases. The polymer could lose more than 70% of its initial weight when partially dehydrated and regain 80% when placed in saline. We observed volume decrease and polymer fragmentation during these study phases.

Conclusions: The polymer can lose weight and volume while it dehydrates. This structural degradation of the polymer could lead to the development of endoleaks and/or migration of the device.

Clinical Impact: Based on the results of previous investigations, due to possible endovascular device degradation, patients with endografts should be offered life-long surveillance, and the Nellix device is no exception. Herein we suggest polymer degradation as one of the possible reasons for the device failure. Although Nellix has been withdrawn from the market, there are numerous patients with this type of endograft. Due to its unpredictable performance in the medium and long term, these patients should be recommended enhanced life-long surveillance every 6 months. Any suspicious conditions during the follow-up must be taken seriously and explantation should be considered.

Download full-text PDF

Source
http://dx.doi.org/10.1177/15266028241274736DOI Listing

Publication Analysis

Top Keywords

nellix device
8
device failure
8
polymer
8
polymer lose
8
life-long surveillance
8
nellix
5
failure mechanisms
4
mechanisms analysis
4
analysis explanted
4
explanted grafts
4

Similar Publications

Objective: The Nellix endovascular aneurysm sealing (EVAS) system demonstrated promising short and midterm outcomes following its introduction in 2011. However, long term results have shown a high incidence of graft failure. This study aimed to compare the long term outcomes of EVAS compared with standard endovascular aneurysm repair (EVAR).

View Article and Find Full Text PDF

Introduction: Endovascular aneurysm sealing (EVAS) appeared to be an innovative alternative to conventional endovascular abdominal aortic aneurysm repair. However, high rates of midterm failure of EVAS led to withdrawal of the device from the market. The study aim was to report midterm outcomes of patients treated with EVAS alone or associated with chimneys (Ch-EVAS) and the management of their complications.

View Article and Find Full Text PDF

Objective: To understand possible reasons for poor durability of the Nellix (Endologix Inc., Irvine, USA) endovascular aneurysm sealing (EVAS) device.

Materials And Methods: 21 Nellix endoprostheses explanted for endoleaks and migration underwent visual examinations of stent structures and instrumental examinations of the polymer endobags on 4 devices.

View Article and Find Full Text PDF

Purpose: This technical note presents a case of a patient with a failed Nellix device (Endologix, Irvine, Calif) who was not deemed fit for open conversion. Our planned approach for repair involved an endovascular procedure utilizing a custom-made branched device.

Technique: An endovascular repair was performed via a custom-made four outer branched device in conjunction with a custom-made bifurcated graft featuring inverted limbs (Cook Inc.

View Article and Find Full Text PDF

Endovascular Repair of a Failed Nellix Endograft Proximal Sealing Zone Using the Altura Stent-Graft: A Case Report and Literature Review.

Vasc Specialist Int

December 2023

Division of Vascular Surgery, 2nd Department of Surgery, Faculty of Medicine, "G. Gennimatas" General Hospital of Thessaloniki, School of Health Sciences, Aristotle University of Thessaloniki, Thessaloniki, Greece.

Endovascular aortic aneurysm sealing (EVAS) with the Nellix endograft was initially considered a groundbreaking and acceptable alternative to conventional endovascular aortic aneurysm repair, with encouraging initial results. However, long-term follow-up has revealed a high incidence of endograft-related complications, such as caudal migration and type Ia endoleaks, indicating the need for reintervention. Managing failed EVAS remains challenging and is an ongoing topic of discussion, especially for high-risk patients.

View Article and Find Full Text PDF