A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Validation of Noninvasive Methemoglobin and Carboxyhemoglobin Measurements Using Pulse Co-Oximeter in Healthy Dogs. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: To assess the agreement between pulse co-oximeter and blood co-oximeter in measuring methemoglobin (MetHb) and carboxyhemoglobin (COHb) in healthy, awake dogs.

Patients And Methods: Forty-five healthy dogs were enrolled in the study. Pulse co-oximetry was performed using the Masimo Radical 7 Pulse Co-Oximeter with a Rainbow adhesive sensor. Simultaneously, venous blood samples were collected, and MetHb and COHb were immediately measured using a Stat Profile Prime Plus VET Critical Care Analyzer. Paired measurements of MetHb and COHb were evaluated via Spearman correlation, intra-class correlation (ICC), and Bland-Altman plots to evaluate the degree of agreement between the pulse co-oximeter and the blood co-oximeter.

Results: A total of 45 paired MetHb and COHb measurements were collected. There was a weak correlation between the pulse co-oximeter and the blood co-oximeter readings. The correlation coefficients for MetHb and COHb were 0.0 (95% CI, -0.3 to 0.3) and 0.03 (95% CI, -0.27 to 0.32), respectively. The ICC indicated poor agreement between the pulse and blood co-oximeter for MetHb (ICC = 0.00, 95% CI: -0.12 to 0.15) and COHb (ICC = 0.03, 95% CI: -0.27 to 0.33). Bland-Altman plots revealed low mean bias but wide limits of agreement, indicating that the pulse co-oximeter overestimated MetHb by on average of 0.7% (P < 0.0001) (95% LoA: -0.5 to 2.0) and COHb by on average 0.2% (P = 0.59) (95% LoA: -4.6 to 5.0).

Conclusion: Obtaining MetHb and COHb measurements with the Masimo Radical 7 Pulse Co-Oximeter is straightforward in healthy, awake dogs. However, the device does not provide accurate measurements compared to the blood co-oximeter, specifically in the range of MetHb and COHb in healthy dogs, based on the wide LoA.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11334929PMC
http://dx.doi.org/10.2147/VMRR.S465945DOI Listing