Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Objective: This study investigates the feasibility of non-contact retrospective respiratory gating and cardiac sensing using continuous wave Doppler radar deployed in an MRI system. The proposed technique can complement existing sensors which are difficult to apply for certain patient populations.

Methods: We leverage a software-defined radio for continuous wave radar at 2.4 GHz to detect in-vivo respiratory and cardiac time-scrolled signals. In-bore radar signal demodulation is verified with full electromagnetic simulations, and its functionality is validated on a test bench and within the MR bore with four normal subjects. Radar sensing was compared against well-known references: electrocardiography on a test bench, system bellows, and pulsed plethysmography sensors within the MRI bore.

Results: The feasibility of non-contact cardiac rate sensing, dynamic breathing sequence synchronization, and in-bore motion correction for retrospective respiratory gating applications was demonstrated. Optimal radar front-end system arrangement, along with spectral isolation and narrow bandwidth of operation, enable MRI-compatible and interference-free motion sensing. The signal-to-noise-ratio degradation by the radar integration was within 4.5% on phantom images.

Conclusion: We confirmed that in-bore retrospective motion correction using CW Doppler radar is feasible without MRI system constraints.

Significance: Non-contact motion correction sensing in MRI may provide better patient handling and throughput by complementing existing system sensors and motion correction algorithms.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11806077PMC
http://dx.doi.org/10.1109/TBME.2024.3440317DOI Listing

Publication Analysis

Top Keywords

motion correction
16
retrospective respiratory
12
respiratory gating
12
doppler radar
12
gating cardiac
8
cardiac sensing
8
radar
8
feasibility non-contact
8
continuous wave
8
mri system
8

Similar Publications

The accuracy of surface imaging in detecting secondary patient motion caused by pitch and roll corrections in pelvic radiotherapy.

Phys Med

September 2025

Department of Biomedical Technology, Faculty of Medicine and Health Technology, Tampere University, Korkeakoulunkatu 8, 33720 Tampere, Finland; School of Electrical and Information Engineering, University of the Witwatersrand, Johannesburg, Jan Smutslaan 1, 2050 Braamfontein, South Africa.

Background And Objective: Correction of rotational setup errors by tilting the treatment couch improves target dose accuracy and prevents healthy tissue overdosage in external beam radiotherapy. However, couch tilts may cause secondary patient motion. This study aimed to quantify the secondary motion caused by pitch and roll corrections and to evaluate the feasibility of surface imaging for detecting the secondary motion in pelvic radiotherapy.

View Article and Find Full Text PDF

Clinical evaluation of motion robust reconstruction using deep learning in lung CT.

Phys Eng Sci Med

September 2025

Department of Radiology, Otaru General Hospital, Otaru, Hokkaido, Japan.

In lung CT imaging, motion artifacts caused by cardiac motion and respiration are common. Recently, CLEAR Motion, a deep learning-based reconstruction method that applies motion correction technology, has been developed. This study aims to quantitatively evaluate the clinical usefulness of CLEAR Motion.

View Article and Find Full Text PDF

Tool use is a complex motor planning problem. Prior research suggests that planning to use tools involves resolving competition between different tool-related action representations. We therefore reasoned that competition may also be exacerbated with tools for which the motions of the tool and the hand are incongruent (e.

View Article and Find Full Text PDF

Background: Evidence supporting surgery in elderly patients with distal radius fractures is limited, but displaced fractures may benefit from surgery. This study aimed to determine whether casting is noninferior to surgery for patients aged 65 years or older with substantially displaced intra-articular (AO type C) distal radius fractures.

Methods: This multicenter randomized controlled noninferiority trial included 138 patients (mean age 76 years, SD 6.

View Article and Find Full Text PDF

Full Free-Breathing Cardiac MRI: Enhancing Efficiency and Image Quality in Clinical Practice.

J Cardiovasc Magn Reson

September 2025

Department of Magnetic Resonance Imaging, Fuwai Hospital and National Center for Cardiovascular Diseases, Peking Union Medical College and Chinese Academy of Medical Sciences, Beijing 100037, China; Key Laboratory of Cardiovascular Imaging, Chinese Academy of Medical Sciences, Beijing 100730, China.

Background: Conventional cardiac magnetic resonance (CMR) examinations require patients to repeatedly hold their breath, which can reduce examination efficiency and pose challenges for patients unable to do so. This study aimed to demonstrate the feasibility and effectiveness of a full free-breathing CMR protocol in clinical practice.

Methods: Patients prospectively enrolled in this study underwent a full free-breathing CMR exam on a 3T scanner between June 1 and June 30, 2024.

View Article and Find Full Text PDF