Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: For reliable DCE MRI parameter estimation, k-space undersampling is essential to meet resolution, coverage, and signal-to-noise requirements. Pseudo-spiral (PS) sampling achieves this by sampling k-space on a Cartesian grid following a spiral trajectory. The goal was to optimize PS k-space sampling patterns for abdomin al DCE MRI.

Methods: The optimal PS k-space sampling pattern was determined using an anthropomorphic digital phantom. Contrast agent inflow was simulated in the liver, spleen, pancreas, and pancreatic ductal adenocarcinoma (PDAC). A total of 704 variable sampling and reconstruction approaches were created using three algorithms using different parametrizations to control sampling density, halfscan and compressed sensing regularization. The sampling patterns were evaluated based on image quality scores and the accuracy and precision of the DCE pharmacokinetic parameters. The best and worst strategies were assessed in vivo in five healthy volunteers without contrast agent administration. The best strategy was tested in a DCE scan of a PDAC patient.

Results: The best PS reconstruction was found to be PS-diffuse based, with quadratic distribution of readouts on a spiral, without random shuffling, halfscan factor of 0.8, and total variation regularization of 0.05 in the spatial and temporal domains. The best scoring strategy showed sharper images with less prominent artifacts in healthy volunteers compared to the worst strategy. Our suggested DCE sampling strategy also showed high quality DCE images in the PDAC patient.

Conclusion: Using an anthropomorphic digital phantom, we identified an optimal PS sampling strategy for abdominal DCE MRI, and demonstrated feasibility in a PDAC patient.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.30213DOI Listing

Publication Analysis

Top Keywords

dce mri
12
sampling
10
pseudo-spiral sampling
8
dce
8
abdominal dce
8
k-space sampling
8
sampling patterns
8
anthropomorphic digital
8
digital phantom
8
contrast agent
8

Similar Publications

Aims: Cardiac tumors are aggressive and asymptomatic in early stages, causing late diagnosis and locoregional metastasis. Currently, the standard of care uses gadolinium-based contrast agents for MRI, and the associated hypersensitivity reactions are a significant concern, such as gadolinium deposition disease. In addition, the proximity of cardiac lesions closer to vital structures complicates surgical interventions.

View Article and Find Full Text PDF

We reveal contrasting behaviors in molecular motion between the two materials, including the identification of resonance-enhanced dynamic features in elastomers. We present a depth-resolved analysis of molecular dynamics in semicrystalline polytetrafluoroethylene (PTFE) and fully amorphous fluorinated elastomer (SIFEL) films using static-gradient solid-state F NMR imaging. By measuring spin-lattice relaxation rates ( ) at multiple frequencies and evaluating the corresponding spectral density functions, we reveal distinct dynamic behaviors between the two materials.

View Article and Find Full Text PDF

Background: Azygos vein aneurysm (AVA) is a rare thoracic pathology that is frequently misdiagnosed. While contrast-enhanced chest computed tomography (CT) or magnetic resonance imaging (MRI) can delineate AVA location and size, these techniques lack the capability for dynamic real-time assessment of internal architecture.

Case Presentation: We present a highly unusual case of a 67-year-old woman who had an isolated azygos vein aneurysm presenting with dysphagia.

View Article and Find Full Text PDF

Glioblastoma is characterized by aggressive infiltration into surrounding brain tissue, hindering complete surgical resection and contributing to poor patient outcomes. Identifying tumor-specific invasion patterns is essential for advancing our understanding of glioblastoma progression and improving surgical and radiotherapeutic strategies. Here, we leverage in vivo dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) to noninvasively quantify interstitial fluid velocity, direction, and diffusion within and around glioblastomas.

View Article and Find Full Text PDF

Male breast MRI: a review of different pathological conditions.

Radiol Med

September 2025

Department of Diagnostic and Public Health, Section of Radiology, University of Verona, P.le L.A. Scuro 10, 37134, Verona, Italy.

The male breast is predisposed to be affected by many of the same pathological processes as the female breast is. The diagnosis of male breast pathologies is generally achievable when clinical evaluation is combined with standard breast imaging methods such as mammography and ultrasound. Magnetic resonance imaging is also a valuable tool in diagnosing the main pathologies affecting the male breast, especially for evaluating pre- and post-surgical treatments and follow-up.

View Article and Find Full Text PDF