Article Abstract

Objective: Strain has become a viable index for evaluating abdominal aortic aneurysm stability after endovascular aneurysm repair (EVAR). In addition, literature has shown that healthy aortic tissue requires a degree of strain to maintain homeostasis. This has led to the hypothesis that too much strain reduction conferred by a high degree of graft oversizing is detrimental to the aneurysm neck in the seal zone of abdominal aortic aneurysms after EVAR. We investigated this in a laboratory experiment by examining the effects that graft oversizing has on the pressure-normalized strain ( /pulse pressure [PP]) reduction using four different infrarenal EVAR endografts and our ultrasound elastography technique. Approximate graft oversizing percentages were 20% (30 mm phantom-graft combinations), 30% (28 mm phantom-graft combinations), and 50% (24 mm phantom-graft combinations).

Methods: Axisymmetric, 10% by mass polyvinyl alcohol phantoms were connected to a flow simulator. Ultrasound elastography was performed before and after implantation with the four different endografts: (1) 36 mm polyester/stainless steel, (2) 36 mm polyester/electropolished nitinol, (3) 35 mm polytetrafluoroethylene (PTFE)/nitinol, and (4) 36 mm nitinol/polyester/platinum-iridium. Five ultrasound cine loops were taken of each phantom-graft combination. They were analyzed over two different cardiac cycles (end-diastole to end-diastole), yielding a total of 10 maximum mean principal strain ( ) values. was divided by pulse pressure to yield pressure-normalized strain ( /PP). An analysis of variance was performed for graft comparisons. We calculated the average percent /PP reduction by manufacturer and percent oversizing. These values were used for linear regression analysis.

Results: Results from one-way analysis of variance showed a significant difference in /PP between the empty phantom condition and all oversizing conditions for all graft manufacturers ((3, 56) = 106.7 [graft A], 132.7 [graft B], 106.5 [graft C], 105.7 [graft D],  < .0001 for grafts A-D). There was a significant difference when comparing the 50% condition with the 30% and 20% conditions across all manufacturers by post hoc analysis ( < .0001). No significant difference was found when comparing the 20% and 30% oversizing conditions for any of the manufacturers or when comparing /PP values across the manufacturers according to percent oversize. Linear regression demonstrated a significant positive correlation between the percent graft oversize and the all-graft average percent /PP reduction (  = 0.84,  < .0001).

Conclusions: This brief report suggests that a 10% increase in graft oversizing leads to an approximate 5.9% reduction in /PP on average. Applied clinically, this increase may result in increased stiffness in axisymmetric vessels after EVAR. Further research is needed to determine if this is clinically significant.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11153908PMC
http://dx.doi.org/10.1016/j.jvssci.2024.100198DOI Listing

Publication Analysis

Top Keywords

graft oversizing
16
ultrasound elastography
12
average percent
12
pressure-normalized strain
12
strain reduction
8
abdominal aortic
8
phantom-graft combinations
8
analysis variance
8
percent /pp
8
/pp reduction
8

Similar Publications

Recurrent anterior instability with glenoid bone loss is a difficult problem with several surgical options. The Latarjet technique remains the gold standard for glenoid bone reconstruction in the setting of critical glenoid bone loss with excellent long-term outcomes. However, this technique has well known downsides including high rates of complications.

View Article and Find Full Text PDF

Thoracic endovascular aortic repair (TEVAR) involving the left subclavian artery (Ishimaru zone 2) presents technical challenges. This multicentre study evaluates the Ankura thoracic stent graft with in-situ fenestration using a dedicated needle system (Lifetech, Shenzhen, China) for various thoracic aortic pathologies. Between January 2020 and December 2025, 59 patients from three tertiary centres underwent in-situ fenestration TEVAR (if-TEVAR) for thoracoabdominal aortic pathologies.

View Article and Find Full Text PDF

Saphenous vein grafts (SVGs) remain the most commonly used conduits in coronary artery bypass grafting, but their high failure rates necessitate repeated revascularization in many cases. SVG perforation, although rare, is a severe complication that can result in cardiac tamponade, myocardial infarction, or death if not promptly managed. Three main stages describe SVG failure pathophysiology, starting with thrombosis, followed by intimal hyperplasia, and progressing to atherosclerosis later in the process.

View Article and Find Full Text PDF

Background: Blunt traumatic thoracic aortic injury (BTAI) is considered a life-threatening surgical emergency. Thoracic endovascular aortic repair (TEVAR) has significantly improved survival rates and reduced complications. The patients are typically young and present with normal aortic diameter.

View Article and Find Full Text PDF

Chest-Wall-Strapping (CWS), a technique forcing the lung to operate at low volumes, is a model for an oversized lung allograft. Dynamic-dysanapsis describes airway dilation mediated by increased elastic recoil from CWS. We hypothesized that Dynamic-dysanapsis can be quantified via the Dysanapsis-Ratio, defined as the airway-size to lung-volume ratio.

View Article and Find Full Text PDF