Background: To overcome deficiencies of the traditional von Willebrand factor (VWF) ristocetin cofactor activity assay (VWF:RCo), several automated assays for VWF platelet-binding activity have been developed. Information on the performance of these assays and their diagnostic utility remains limited.
Objectives: To validate the VWF:glycoprotein IbM assay INNOVANCE VWF Ac and compare it with an automated VWF:RCo assay as well as with an automated assay and a manual VWF:Ab assay and to generate reference ranges and analyze reproducibility of the VWF:glycoprotein IbM assay.
Methods: Clinical sites enrolled healthy subjects and patients representing the intended use population; VWF activity assays were performed, and results were analyzed. The performance of the INNOVANCE VWF Ac assay was also compared between the BCS XP System and the CS-2500 and CS-5100 analyzers.
Results: The INNOVANCE VWF Ac assay correlated well with the VWF:RCo assay and the automated HemosIL VWF:Ab assay, with Pearson coefficients of >.9 and a predicted bias of ≤5.0 IU/dL at VWF levels of 30 IU/dL and ≤5.8 IU/dL at the levels of 50 IU/dL, but correlation and bias were not as good when compared with the REAADS manual VWF:Ab assay. Reference ranges observed for healthy subjects correlated well with previously published findings. Reproducibility of the INNOVANCE VWF Ac assay on the BCS XP System and the CS analyzers was excellent, as was correlation among devices.
Conclusion: The characteristics of the INNOVANCE VWF Ac assay regarding comparability with other VWF activity assays, reference ranges, and precision support the use of this assay for evaluation of patients with concern for von Willebrand disease.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11152683 | PMC |
http://dx.doi.org/10.1016/j.rpth.2024.102422 | DOI Listing |
J Appl Lab Med
September 2024
Quest Diagnostics, Secaucus, NJ, United States.
Background: Historically, von Willebrand factor (VWF) activity assays utilized ristocetin despite limitations including poor limits of detection and high imprecision. Newer VWF activity assays such as the INNOVANCE® VWF Ac assay, however, do not rely on ristocetin to measure platelet-dependent VWF function. The purpose of this study was to evaluate the analytical and clinical performance of the Siemens Healthineers INNOVANCE VWF Ac Assay on the Siemens BCS® XP and the Sysmex® CS-2500 systems in a large reference laboratory setting.
View Article and Find Full Text PDFJ Neurol Sci
August 2024
Vascular Neurology Research Foundation, c/o Department of Neurology, Tallaght University Hospital (TUH) / The Adelaide and Meath Hospital, Dublin, incorporating the National Children's Hospital (AMNCH), Dublin, Ireland; Department of Neurology, TUH / AMNCH, Dublin, Ireland; Stroke Service, TUH / AMN
Data are limited on the impact of commencing antiplatelet therapy on von Willebrand Factor Antigen (VWF:Ag) or von Willebrand Factor propeptide (VWFpp) levels and ADAMTS13 activity, and their relationship with platelet reactivity following TIA/ischaemic stroke. In this pilot, observational study, VWF:Ag and VWFpp levels and ADAMTS13 activity were quantified in 48 patients ≤4 weeks of TIA/ischaemic stroke (baseline), and 14 days (14d) and 90 days (90d) after commencing aspirin, clopidogrel or aspirin+dipyridamole. Platelet reactivity was assessed at moderately-high shear stress (PFA-100® Collagen-Epinephrine / Collagen-ADP / INNOVANCE PFA P2Y assays), and low shear stress (VerifyNow® Aspirin / P2Y12, and Multiplate® Aspirin / ADP assays).
View Article and Find Full Text PDFRes Pract Thromb Haemost
May 2024
Pathology and Internal Medicine (Hematology/Oncology), University of Texas Southwestern Medical Center, Dallas, Texas, USA.
Background: To overcome deficiencies of the traditional von Willebrand factor (VWF) ristocetin cofactor activity assay (VWF:RCo), several automated assays for VWF platelet-binding activity have been developed. Information on the performance of these assays and their diagnostic utility remains limited.
Objectives: To validate the VWF:glycoprotein IbM assay INNOVANCE VWF Ac and compare it with an automated VWF:RCo assay as well as with an automated assay and a manual VWF:Ab assay and to generate reference ranges and analyze reproducibility of the VWF:glycoprotein IbM assay.
Georgian Med News
November 2022
6Haematology Laboratory-Blood Bank, Aretaieion Hospital, School of Medicine, National and Kapodistrian University of Athens, Athens, Greece.
The laboratory detection of factors that participate in coagulation mechanisms in patients with coronary heart disease may lead to important findings regarding the contribution of endothelial function to atherosclerotic lesions of coronary arteries. The main purpose of this study was to investigate the role of high-sensitivity C-reactive protein (hs-CRP), von Willebrand factor (vWF) activity, thrombomodulin (TM), ADAMTS13 activity and myeloperoxidase (MPO) in patients undergoing coronary angiography due to non-ST-elevation myocardial infarction (NSTEMI), unstable angina (UA) and stable angina pectoris with positive stress testing-induced myocardial ischemia (controls). Furthermore, the measured biomarkers were examined among patients with classical cardiovascular risk factors.
View Article and Find Full Text PDFBackground: Von Willebrand disease requires laboratory confirmation with quantitative and qualitative measurements of von Willebrand factor (VWF). Qualitative VWF-activity (VWF-Ac) tests have poor inter- and intra-laboratory reproducibility with coefficients of variation (CVs) as high as 64%, often lacking accuracy at low VWF-Ac levels.
Methods: This study evaluated the recently launched immunoturbidometric STAGO® STA-VWF:RCo® reagent for VWF-Ac.