98%
921
2 minutes
20
Objective: To analyze the hip joint biomechanics of the acetabular anatomical reconstruction and nonanatomical reconstruction in total hip arthroplasty (THA) for Crowe type Ⅲ developmental dysplasia of the hip (DDH) by finite element method, which provided theoretical foundation and experimental basis for the anatomical acetabular reconstruction during THA in clinical practice.
Methods: One patient with left end-stage hip arthritis secondary to Crowe type Ⅲ DDH was selected in this study, who underwent total hip arthroplasty in the orthopedic department of the First Affiliated Hospital of Bengbu Medical College in April 2020. This patient was female, 57 years old. The preoperative and postoperative three dimentional CT scan of the patient's pelvis were performed. Fourteen acetabular cup models with different anteversion, inclination and rotation center height were established in Mimics and 3-Matic software. The boundary and load conditions were set in Abaqus software. The Von Mises and stress distribution of the hip joint were calculated and observed.
Results: In the Crowe type Ⅲ DDH THA, if the hip rotation center was restored anatomically and the acetabular cup's inclination was set as 40°, the cup's anteversion varied from 5° to 25°, the lowest Von Mises value of acetabular cup and polyethylene liner occured in 20°anteversioin;if the hip rotation center was restored anatomically and the acetabular cup's anteversion was set as 15°, the cup's inclination varied from 35° to 55°, the lowest Von Mises value of acetabular cup and polyethylene liner occured in 35° inclination;if the acetabular cup's anteversion and inclination were set as 15°and 40°respectively, the up migration of hip rotaion center varied from 0 mm to 20 mm, the lowest Von Mises value of acetabular cup and polyethylene liner occured in 10 mm up migration. In all fourteen models, the Von Mises value of the acetabulum, acetabulum cup and polyethylene liner were lowest when the acetabular cup's anteversion and inlcination were 15°, 35° respectively, as well as the rotation center was restored anatomically.
Conclusion: In total hip arthroplasty for Crowe type Ⅲ DDH, the anatomical restoration of hip rotation center with 15° anteversion and 35° inclination of the acetabular cup are suggested, bone graft above the acetabular cup and additional screws are recommended simultaneously to further reduce the Von Mises of hip joint.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.12200/j.issn.1003-0034.20221143 | DOI Listing |
J Orthop Sci
September 2025
American Hip Institute Research Foundation, Chicago, IL 60018, USA; American Hip Institute, Chicago, IL 60018, USA. Electronic address:
Background And Objective: With obese individuals experiencing osteoarthritis (OA) at early stages of life, hip resurfacing (HR) has emerged as an alternative to arthroplasty. The purpose is to conduct a short-term analysis on patients with obesity who underwent primary HR for OA compared to a benchmark control group of non-obese patients.
Materials And Methods: Patients with a body mass index (BMI) ≥ 30 kg/m2 who underwent HR from 2010 to 2021 were eligible for inclusion.
Int J Surg
September 2025
Shanghai Key Laboratory of Orthopaedic Implants, Department of Orthopaedic Surgery, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, People's Republic of China.
Background: Precise acetabular cup placement in total hip arthroplasty (THA) heavily relies on surgeons' visual judgment of angles. However, whether inherent visual angle misperception among surgeons affects surgical outcomes remains unclear. This study is the first to reveal that surgeons universally exhibit visual angle misperception, a key factor causing the cup implant positioning deviations in THA.
View Article and Find Full Text PDFActa Ortop Mex
September 2025
Servicio de Ortopedia y Traumatología, Hospital de San Rafael, Hospitales Pascual. Cádiz, España.
Introduction: anatomical deformities such as developmental dysplasia of the hip (DDH) and Perthes disease represent a challenge for reconstruction. The use of 3D-printed models can be helpful for assessing the deformity, bone mass, implant size, and orientation.
Objectives: to prospectively evaluate the outcomes of 3D simulation in primary total hip arthroplasty.
Arch Orthop Trauma Surg
September 2025
Division of Orthopaedics and Traumatology, Cantonal Hospital Winterthur, Winterthur, Switzerland.
Background: Accurate acetabular cup orientation in total hip arthroplasty (THA) is crucial for successful outcomes. Intraoperative fluoroscopy may be used to evaluate acetabular cup placement. This study aimed to evaluate the accuracy of purely visual estimation of cup inclination and anteversion using intraoperative fluoroscopy, considering different surgeon experience levels and cup designs.
View Article and Find Full Text PDFFront Surg
August 2025
Department of Orthopedics, The First Hospital of Jilin University, Changchun, Jilin, China.
Background: Acetabular reconstruction is often challenging in revision hip arthroplasty, especially in the face of moderate to severe acetabular bone deficiency. In some severe bone defects, double-metal tantalum cups can improve the contact area between bone and implants, increase the surface area for bone ingrowth, and better restore the anatomical position of the acetabulum. Furthermore, with a good press-fit, the auxiliary screw has a minimal effect on acetabular cup stability.
View Article and Find Full Text PDF