Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1075
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3195
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
Background: Clinical decision-making for risk stratification for possible myocardial infarction (MI) uses high-sensitivity cardiac troponin (hs-cTn) thresholds that range from the limit of detection to several-fold higher than the upper reference limit (URL). To establish a minimum analytical variation standard, we can quantify the effect of variation on the population clinical measures of safety (sensitivity) and effectiveness [proportion below threshold, or positive predictive value (PPV)].
Methods: From large datasets of patients investigated for possible MI with the Abbott hs-cTnI and Roche hs-cTnT assays, we synthesized datasets of 1 000 000 simulated patients. Troponin concentrations were randomly varied several times based on absolute deviations of 0.5 to 3 ng/L and relative changes of 2% to 20% around the low-risk threshold (5 ng/L) and URLs, respectively.
Results: For both assays at the low-risk thresholds, there were negligible differences in sensitivity (<0.3%) with increasing analytical variation. The proportion of patients characterized as low risk reduced by 30% to 29% (Roche) and 53% to 44% (Abbott). At the URL, increasing analytical variation also did not change sensitivity; the PPV fell by less than 3%. For risk stratification, increased delta thresholds (change between serial troponin concentrations) increased sensitivity at the cost of a decreased percentage of patients below the delta threshold, with the largest changes at the greatest analytical variation.
Conclusions: At the low-risk threshold, analytical variation up to 3 ng/L minimally impacted the safety metric (sensitivity) but marginally reduced effectiveness. Similarly, at the URL even relative variation up to 25% minimally impacted safety metrics and effectiveness. Analytical variation for delta thresholds did not negatively impact sensitivity but decreased effectiveness.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/clinchem/hvae059 | DOI Listing |