Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Objective: Aortic valve neocuspidization (AVNeo) using autologous pericardium is a promising technique. Expected advantages are reduced immune response, appropriate biomechanics and lower treatment expenses. Nevertheless, autologous pericardium can be affected by patient's age and comorbidities. Usually, glutaraldehyde (GA) - fixed bovine pericardium is the basic material for aortic valve prostheses, easy available and carefully pre-examined in a standardized fabrication process. Aim of the study is the verification of autologous pericardial tissue homogeneity by analysing tissue thickness, biomechanics and extracellular matrix (ECM) composition.

Methods: Segments of human GA-fixed pericardium selected by the surgeon based on visual criteria for cusp pre-cut and remaining after surgical AV replacement were investigated in comparison to bovine standard tissue treated equivalently. Pericardium sampling was performed at up to three positions of each sutured cusp for histological or biomechanical analysis, according to tissue availability.

Results And Conclusions: Human pericardia exhibited a higher heterogeneity in collagen content, density of vessel structures and elastic moduli. Thickness, vessel density and collagen and elastin content differed significantly between the species. In contrast, significant interindividual differences were detected in most properties investigated for human pericardial samples but only for tissue thickness in bovine tissues. Higher heterogeneity of human pericardium, differing vessel and collagen content compared to bovine state-of-the-art material might be detrimental for long term AV functionality or deterioration and have to be intensely investigated in patients follow up after autologous cusp replacement.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11058761PMC
http://dx.doi.org/10.1007/s10856-024-06790-2DOI Listing

Publication Analysis

Top Keywords

aortic valve
12
valve neocuspidization
8
autologous pericardium
8
tissue thickness
8
higher heterogeneity
8
collagen content
8
pericardium
7
tissue
6
tissue requirements
4
requirements application
4

Similar Publications

Background: Chronic kidney disease (CKD) is associated with a higher prevalence of valvular diseases and increased mortality from cardiovascular causes. Factors that influence the genesis of cardiac valve calcification (CVC) in these patients are not well-defined.

Objective: To determine the risk factors for valvular calcification in patients with CKD.

View Article and Find Full Text PDF

Objectives: Failure of vascular closure device (VCD) is the most common cause of access-site vascular complications in transfemoral transcatheter aortic valve implantation (TAVI). The authors sought to determine if the systematic use of arteriotomy-site ballooning with concomitant manual compression following the delivery of a plug-based VCD (MANTA, Teleflex) can optimize toggle-plug assembly apposition to the common femoral artery (CFA) wall and improve the final hemostatic efficacy.

Methods: In this prospective, observational, single-center study, 323 consecutive patients undergoing transfemoral TAVI from October 2021 to December 2024 underwent access closure with the MANTA VCD.

View Article and Find Full Text PDF

Objectives: The authors hypothesized that the origin of the right coronary artery (RCA) is a direct continuation of the major aortic arch branches (MAAB) takeoff plane, which may have implications for brachiocephalic interventions and next generation transcatheter aortic valve intervention (TAVI) embolic protection devices (EPDs).

Methods: In this single-center, retrospective, cross-sectional study, the authors analyzed computed tomographic angiography (CTA) images from 92 patients undergoing TAVI evaluation to determine the spatial relationship between the origin of the RCA and the MAAB takeoff plane. Patients with prior cardiothoracic or aortic interventions and those with anomalous RCA origin were excluded.

View Article and Find Full Text PDF

Comparison of short- and long-term outcomes between transcatheter and surgical aortic valve replacement for bicuspid aortic valve stenosis: a systematic review and meta-analysis.

Int J Surg

September 2025

Department of Cardiovascular Surgery, The Affiliated Hospital, Southwest Medical University, Metabolic Vascular Diseases Key Laboratory of Sichuan Province, Key Laboratory of Cardiovascular Remodeling and Dysfunction, Luzhou, Sichuan, PR China.

Objective: This meta-analysis aimed to compare the perioperative safety and efficacy of transcatheter aortic valve replacement (TAVR) versus surgical aortic valve replacement (SAVR) in bicuspid aortic valve (BAV) stenosis.

Methods: We systematically analyzed studies from PubMed, Cochrane Library, Embase, and CNKI comparing TAVR and SAVR in BAV stenosis. Outcomes included postoperative mortality, complications, all-cause survival, and freedom from stroke.

View Article and Find Full Text PDF