98%
921
2 minutes
20
Background: Recent multicenter, multivendor MRI-based R2* vs. liver iron concentration (LIC) calibrations (i.e., MCMV calibrations) may facilitate broad clinical dissemination of R2*-based LIC quantification. However, these calibrations are based on a centralized offline R2* reconstruction, and their applicability with vendor-provided R2* maps is unclear.
Purpose: To determine R2* ranges of agreement between the centralized and three MRI vendors' R2* reconstructions.
Study Type: Prospective.
Subjects: Two hundred and seven subjects (mean age 37.6 ± 19.6 years; 117 male) with known or suspected iron overload from four academic medical centers.
Field Strength/sequence: Standardized multiecho spoiled gradient echo sequence at 1.5 T and 3.0 T for R2* mapping and a multiple spin-echo sequence at 1.5 T for LIC quantification. MRI vendors: GE Healthcare, Philips Healthcare, and Siemens Healthineers.
Assessment: R2* maps were generated using both the centralized and vendor reconstructions, and ranges of agreement were determined. R2*-LIC linear calibrations were determined for each site, field strength, and reconstruction and compared with the MCMV calibrations.
Statistical Tests: Bland-Altman analysis to determine ranges of agreement. Linear regression, analysis of covariance F tests, and Tukey's multiple comparison testing to assess reproducibility of calibrations across sites and vendors. A P value <0.05 was considered significant.
Results: The upper limits of R2* ranges of agreement were approximately 500, 375, and 330 s for GE, Philips, and Siemens reconstructions, respectively, at 1.5 T and approximately 700 and 800 s for GE and Philips, respectively, at 3.0 T. Within the R2* ranges of agreement, vendor R2*-LIC calibrations demonstrated high reproducibility (no significant differences between slopes or intercepts; P ≥ 0.06) and agreed with the MCMV calibrations (overlapping 95% confidence intervals).
Data Conclusion: Based on the determined upper limits, R2* measurements obtained from vendor-provided R2* maps may be reliably and practically used to quantify LIC less than approximately 8-13 mg/g using the MCMV calibrations and similar acquisition parameters as this study.
Evidence Level: 1 TECHNICAL EFFICACY: Stage 3.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11502507 | PMC |
http://dx.doi.org/10.1002/jmri.29401 | DOI Listing |
ERJ Open Res
September 2025
Cambridge University Hospitals NHS Foundation Trust, Cambridge, UK.
Background: Measurement of total lung capacity (TLC) requires large and expensive equipment. We aimed to investigate whether spirometric restriction and low alveolar volume measured by single breath gas transfer ( ) can be used to identify those with a low TLC.
Methods: We retrospectively analysed data from adults referred to Cambridge University Hospitals between January 2016 and December 2023.
Front Oncol
August 2025
Department of Colorectal Surgery, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Objective: The retrieval of 12 lymph nodes (LNs) remains a crucial criterion for accurate staging and prognosis evaluation in rectal cancer (RC). However, some patients fail to meet this threshold after surgery. This study developed a nomogram model based on clinical variables to predict the probability of retrieving 12 LNs postoperatively.
View Article and Find Full Text PDFMed Phys
September 2025
School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, P.R. China.
Background: In catheter-based radiofrequency ablation (RFA), energy is delivered to heterogeneous thin-walled tissues to induce therapeutic heating. Variations in electrical and mechanical properties of tissue contents have a great effect on outcomes.
Purpose: The objective of this study is to develop models that replicate tissue heterogeneity and visualize ablation zones for effective evaluation and optimization.
Anal Methods
September 2025
Institut de recherche Robert-Sauvé en santé et en sécurité du travail (IRSST), Qc, Canada.
Toluene diisocyanate (TDI) is an irritant (skin, eye and respiratory) and a sensitizer. This compound is used to manufacture polyurethane materials such as flexible foams. The use of isocyanates may lead to exposure by inhalation and/or skin contact and isocyanates are recognized as a cause of occupational asthma.
View Article and Find Full Text PDFAppl Radiat Isot
September 2025
Université Paris-Saclay, CEA, LIST, Laboratoire National Henri Becquerel (LNE-LNHB), Palaiseau, F-91120 France.
Reliable X-ray emission intensities are essential for quantitative material analysis using X-ray spectrometry and for the efficiency calibration of energy-dispersive spectrometers. In order to improve the reliability of data, reference-free measurements were performed to determine X-ray emission intensities, along with their associated uncertainties, for a set of standard radionuclides in the energy range from 5.4 keV to 53.
View Article and Find Full Text PDF