An extended phase graph-based framework for DANTE-SPACE simulations including physiological, temporal, and spatial variations.

Magn Reson Med

Wellcome Centre for Integrative Neuroimaging, FMRIB Division, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.

Published: July 2024


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: The delay alternating with nutation for tailored excitation (DANTE)-sampling perfection with application-optimized contrasts (SPACE) sequence facilitates 3D intracranial vessel wall imaging with simultaneous suppression of blood and CSF. However, the achieved image contrast depends closely on the selected sequence parameters, and the clinical use of the sequence is limited in vivo by observed signal variations in the vessel wall, CSF, and blood. This paper introduces a comprehensive DANTE-SPACE simulation framework, with the aim of providing a better understanding of the underlying contrast mechanisms and facilitating improved parameter selection and contrast optimization.

Methods: An extended phase graph formalism was developed for efficient spin ensemble simulation of the DANTE-SPACE sequence. Physiological processes such as pulsatile flow velocity variation, varying flow directions, intravoxel velocity variation, diffusion, and effects were included in the framework to represent the mechanisms behind the achieved signal levels accurately.

Results: Intravoxel velocity variation improved temporal stability and robustness against small velocity changes. Time-varying pulsatile velocity variation affected CSF simulations, introducing periods of near-zero velocity and partial rephasing. Inclusion of diffusion effects was found to substantially reduce the CSF signal. Blood flow trajectory variations had minor effects, but differences along the trajectory reduced DANTE efficiency in low- areas. Introducing low-velocity pulsatility of both CSF and vessel wall helped explain the in vivo observed signal heterogeneity in both tissue types.

Conclusion: The presented simulation framework facilitates a more comprehensive optimization of DANTE-SPACE sequence parameters. Furthermore, the simulation framework helps to explain observed contrasts in acquired data.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.30071DOI Listing

Publication Analysis

Top Keywords

velocity variation
16
vessel wall
12
simulation framework
12
extended phase
8
sequence parameters
8
vivo observed
8
observed signal
8
dante-space sequence
8
intravoxel velocity
8
diffusion effects
8

Similar Publications

Acute Effects of Mechanical Insufflation-Exsufflation on Cough Peak Flow, Chest Wall Volumes, and Breathing Pattern of Patients With Amyotrophic Lateral Sclerosis.

Respir Care

September 2025

Ms. Vieira, Dr. Lima, Dr. Pondofe, Ms. Maciel, Dr. da Fonsêca, Dr. Resqueti, and Dr. Fregonezi, are affiliated with the Departamento de Fisioterapia, PneumoCardioVascular Lab/HUOL and the Laboratório de Inovação Tecnológica em Reabilitação, Universidade Federal do Rio Grande do Norte, Natal,

Mechanical insufflation-exsufflation (MI-E) consists of increasing expiratory air flow, thereby promoting an increase in cough peak flow (CPF) and secretion clearance. Respiratory impairment, characterized by reduced lung volumes and ineffective cough, is the major cause of morbidity and mortality in patients with amyotrophic lateral sclerosis (ALS). This study aimed to assess the acute effects of MI-E on CPF and chest wall compartmental and operational volumes in patients with ALS.

View Article and Find Full Text PDF

Canonical stochastic models of decision-making treats decision and action as independent and sequential processes. However, studies involving limb movements consistently show that movement duration and kinematics are influenced by the quality of evidence. We tested whether saccade velocity varies with the quality of evidence in monkeys performing a visual search GO/NOGO task in which singleton elongation cued the GO/NOGO stimulus-response rule and the location of a color singleton specified saccade endpoint.

View Article and Find Full Text PDF

Background: International consensus recommends use of kinematic metrics of movement during standardized functional tasks after stroke to ascertain whether rehabilitation is driving behavioral restitution or compensation. Quality of human movement can be characterized by fluency metrics including smoothness and hesitation. Before using these metrics in stroke rehabilitation it is important to find whether 'reference values', from healthy adults, are repeatable.

View Article and Find Full Text PDF

This study investigates the sensitivity of ultrasonic phase velocity and attenuation statistics to grain size distribution types, specifically monomodal and bimodal, in cubic finite-sized polycrystals with identical overall arithmetic or volumetric grain size statistics. A large ensemble of synthetic polycrystalline microstructures was generated, and their ultrasonic responses were computed using a modified spectral function method. The latter enables estimation of both effective phase velocities and attenuation coefficients while rigorously accounting for finite specimen dimensions and statistical variabilities across different samples.

View Article and Find Full Text PDF

This study presents a comprehensive framework combining Selective Laser Melting (SLM) of Titanium (Ti64) alloys, finite element simulation, and artificial intelligence (AI) to advance orthopedic implants' design and predictive evaluation. Dense Ti64 specimens are fabricated using ten distinct SLM parameter sets to explore the effects of volumetric energy density (VED) on mechanical behavior, porosity distribution, and microstructural integrity. Optimal VED ranges are identified to balance defect minimization and mechanical performance, with porosity levels strongly influencing tensile strength and Young's modulus.

View Article and Find Full Text PDF