Background And Objectives: Robot-assisted biopsies have gained popularity in the last years. Most robotic procedures are performed with a floor-based robotic arm. Recently, Medtronic Stealth Autoguide, a miniaturized robotic arm that work together with an optical neuronavigation system, was launched. Its application in pediatric cases is relatively unexplored. In this study, we retrospectively report our experience using the Stealth Autoguide, for frameless stereotactic biopsies in pediatric patients.
Methods: Pediatric patients who underwent stereotactic biopsy using the Stealth Autoguide cranial robotic platform from July 2020 to May 2023 were included in this study. Clinical, neuroradiological, surgical, and histological data were collected and analyzed.
Results: Nineteen patients underwent 20 procedures (mean age was 9-year-old, range 1-17). In four patients, biopsy was part of a more complex surgical procedure (laser interstitial thermal therapy - LITT). The most common indication was diffuse intrinsic brain stem tumor, followed by diffuse supratentorial tumor. Nine procedures were performed in prone position, eight in supine position, and three in lateral position. Facial surface registration was adopted in six procedures, skull-fixed fiducials in 14. The biopsy diagnostic tissue acquisition rate was 100% in the patients who underwent only biopsy, while in the biopsy/LITT group, one case was not diagnostic. No patients developed clinically relevant postoperative complications.
Conclusion: The Stealth Autoguide system has proven to be safe, diagnostic, and highly accurate in performing stereotactic biopsies for both supratentorial and infratentorial lesions in the pediatric population.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00381-024-06349-0 | DOI Listing |
Front Robot AI
December 2024
Department of Neurosurgery, Medical University of Graz, Graz, Styria, Austria.
Background: Accurate histological analysis is crucial for confirming intracerebral neoplasia due to the diverse array of potential diagnoses presented by imaging. In the realm of biopsy techniques, the use of robot-based systems is on the rise, primarily owing to their heightened targeting accuracy. The objective of this study was to elucidate the practicality, learning curve and workflow associated with robot-guided biopsies upon their introduction to a neurosurgical centre.
View Article and Find Full Text PDFWorld Neurosurg
December 2024
Department of Neurosurgery, Santobono-Pausilipon Children's Hospital, AORN, Naples, Italy.
Magnetic resonance-guided laser interstitial thermal therapy (MRgLITT) is a new minimally invasive treatment for tuberous sclerosis complex (TSC)-associated epilepsy in children. This video describes a case of a 17-year-old girl with TSC-associated drug-resistant epilepsy treated with robotic-assisted MRgLITT. In our case, MRgLITT was safe and effective in simultaneous targeting of multiple epileptic tubers in 1 single procedure, leading to a marked decrease in seizure frequency.
View Article and Find Full Text PDFWorld Neurosurg
December 2024
Neurosurgery Department, ARNAS Civico Di Cristina Benfratelli Hospital, Palermo, Italy.
Background: The Stealth Autoguide (Medtronic, Minneapolis, MN) is a robotic auto-targeting device for stereotactic brain biopsy, placement of stereoelectroencephalography electrodes, and laser ablation therapy. This paper evaluates its off-label application as robotic endoscope holder in transnasal and intraventricular neurosurgery. The potential to enhance stability, reduce manual adjustments, and improve surgical precision is discussed, alongside cost-effectiveness and feasibility.
View Article and Find Full Text PDFOper Neurosurg
April 2025
Department of Pediatrics, Herman B. Wells Center for Pediatric Research, Indiana University School of Medicine, Indianapolis , Indiana , USA.
Background And Objectives: In the molecular era of neuro-oncology, it is increasingly necessary to obtain tissue for next-generation sequencing and methylome profile for prognosis and targeted oncological management. Brainstem tumors can be technically challenging to biopsy in the pediatric population. Frame-based and frameless techniques have previously been described and proven to be safe and efficacious in children.
View Article and Find Full Text PDFObjective: Stereotactic techniques play an important role in neurosurgery. The development of a miniaturized cranial robot with an efficient workflow and accurate surgical execution is an important step in a broader application of these techniques. Herein, the authors describe their experience with the Medtronic Stealth Autoguide miniaturized cranial robot.
View Article and Find Full Text PDF