Article Abstract

Aim: This study aimed to establish the shear wave measurement (SWM) cut-off value for each fibrosis stage using magnetic resonance (MR) elastography values as a reference standard.

Methods: We prospectively analyzed 594 patients with chronic liver disease who underwent SWM and MR elastography. Correlation coefficients (were analyzed, and the diagnostic value was evaluated by the area under the receiver operating characteristic curve. Liver stiffness was categorized by MR elastography as F0 (<2.61 kPa), F1 (≥2.61 kPa, <2.97 kPa, any fibrosis), F2 (≥2.97 kPa, <3.62 kPa, significant fibrosis), F3 (≥3.62 kPa, <4.62 kPa, advanced fibrosis), or F4 (≥4.62 kPa, cirrhosis).

Results: The median SWM values increased significantly with increasing fibrosis stage (p < 0.001). The correlation coefficient between SWM and MR elastography values was 0.793 (95% confidence interval 0.761-0.821). The correlation coefficients between SWM and MR elastography values significantly decreased with increasing body mass index and skin-capsular distance; skin-capsular distance values were associated with significant differences in sensitivity, specificity, accuracy, or positive predictive value, whereas body mass index values were not. The best cut-off values for any fibrosis, significant fibrosis, advanced fibrosis, and cirrhosis were 6.18, 7.09, 8.05, and 10.89 kPa, respectively.

Conclusions: This multicenter study in a large number of patients established SWM cut-off values for different degrees of fibrosis in chronic liver diseases using MR elastography as a reference standard. It is expected that these cut-off values will be applied to liver diseases in the future.

Download full-text PDF

Source
http://dx.doi.org/10.1111/hepr.14026DOI Listing

Publication Analysis

Top Keywords

shear wave
8
wave measurement
8
diagnostic performance
4
performance shear
4
measurement detection
4
detection hepatic
4
hepatic fibrosis
4
fibrosis multicenter
4
multicenter prospective
4
prospective study
4

Similar Publications

Sectionally nonlinearly functionally graded (SNFG) structures with triply periodic minimal surface (TPMS) are considered ideal for bone implants because they closely replicate the hierarchical, anisotropic, and porous architecture of natural bone. The smooth gradient in material distribution allows for optimal load transfer, reduced stress shielding, and enhanced bone ingrowth, while TPMS provides high mechanical strength-to-weight ratio and interconnected porosity for vascularization and tissue integration. Wherein, The SNFG structure contains sections with thickness that varies nonlinearly along their length in different patterns.

View Article and Find Full Text PDF

The objective of this study was to assess the link between postpartum blood loss and placental elastography in multiparous pregnancies.In this prospective study, multiparous women who delivered at term between 2020 and 2024 were included. During the obstetric ultrasonography procedure, placental elastography was evaluated through the utilization of the shear wave elastography (SWE) technique.

View Article and Find Full Text PDF

Alterations in skeletal muscle morphology and composition are critical factors in cerebral palsy (CP), including changes in passive stiffness and in belly and fascicle lengths. In this study, we quantified the relative contributions of muscle and tendon to passive stiffness across the ankle range of motion in individuals with CP and typically developing (TD) peers. We also investigated morphological factors underlying increased muscle stiffness.

View Article and Find Full Text PDF

Background: Hepatic sinusoidal obstruction syndrome (SOS), or veno-occlusive disease (VOD), is a severe complication following hematopoietic stem cell transplantation (HSCT), often leading to liver dysfunction and poor outcomes if not detected early. Traditional diagnostic methods, including ultrasound and liver biopsy, have limitations in sensitivity and feasibility. Non-invasive elastography techniques, such as transient elastography (TE) and shear-wave elastography (SWE), offer a promising alternative by quantitatively assessing liver stiffness.

View Article and Find Full Text PDF

Multiparametric Ultrasound for Preoperative Assessment of Parotid Tumors: A Novel Diagnostic Pathway.

Head Neck

September 2025

Department of Radiology, Charité - Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität Zu Berlin, and Berlin Institute of Health, Berlin, Germany.

Background: Accurate preoperative differentiation of parotid gland tumors (PGTs) is essential for facial nerve preservation. This study evaluates a novel, real-time multiparametric ultrasound (mpUS) approach combining B-mode, shear wave elastography (SWE), and contrast-enhanced ultrasound (CEUS), based on qualitative image interpretation.

Methods: Eighty-nine patients with 91 PGTs underwent mpUS prior to surgery or ultrasound-guided biopsy.

View Article and Find Full Text PDF