Study Design: Cadaveric study.

Objectives: The purpose of this study was to compare a novel, integrated 3D navigational system (NAV) and conventional fluoroscopy in the accuracy, efficiency, and radiation exposure of thoracolumbar percutaneous pedicle screw (PPS) placement.

Methods: Twelve skeletally mature cadaveric specimens were obtained for twelve individual surgeons. Each participant placed bilateral PS at 11 segments, from T8 to S1. Prior to insertion, surgeons were randomized to the sequence of techniques and the side (left or right). Following placement, a CT scan of the spine was obtained for each cadaver, and an independent reviewer assessed the accuracy of screw placement using the Gertzbein grading system. Outcome metrics of interest included a comparison of breach incidence/severity, screw placement time, total procedure time, and radiation exposure between the techniques. Bivariate statistics were employed to compare outcomes at each level.

Results: A total of 262 screws (131 using each technique) were placed. The incidence of cortical breaches was significantly lower with NAV compared to FG (9% vs 18%; = .048). Of breaches with NAV, 25% were graded as moderate or severe compared to 39% in the FG subgroup ( = .034). Median time for screw placement was significantly lower with NAV (2.7 vs 4.1 min/screw; = .012), exclusive of registration time. Cumulative radiation exposure to the surgeon was significantly lower for NAV-guided placement (9.4 vs 134 μGy, = .02).

Conclusions: The use of NAV significantly decreased the incidence of cortical breaches, the severity of screw breeches, screw placement time, and radiation exposure to the surgeon when compared to traditional FG.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11572174PMC
http://dx.doi.org/10.1177/21925682231224394DOI Listing

Publication Analysis

Top Keywords

screw placement
20
radiation exposure
16
accuracy efficiency
8
pedicle screw
8
placement time
8
time radiation
8
incidence cortical
8
cortical breaches
8
lower nav
8
exposure surgeon
8

Similar Publications

Objective: To determine if a novel robotic system has comparable positional and angular accuracy to that achievable with patient-specific guides (PSG) when used for transcondylar screw (TCS) placement in the canine humerus.

Study Design: Experimental laboratory study.

Sample Population: A total of 32 synthetic humeral models (16 per group).

View Article and Find Full Text PDF

Background: Lumbar spondylolisthesis (LS) is a spinal disorder that often necessitates surgical intervention. However, evidence on the comparative clinical value of robot-assisted full-endoscopic transforaminal lumbar interbody fusion (RA FE-TLIF) versus conventional FE-TLIF in early-grade (Grades I and II) LS remains limited, leaving uncertainty about its true clinical value in this patient population. This study aims to compare the clinical efficacy and safety of FE-TLIF with RA FE-TLIF in patients with Grade I and II LS.

View Article and Find Full Text PDF

The purpose of this study is to propose a standardized classification of minimally invasive cervical pedicle screw (MICEPS) fixation according to the levels instrumented and the extent of the construct, thereby facilitating reproducible surgical planning and technique. We developed a three-tiered MICEPS classification with a specific surgical algorithm based on anatomic levels and construct length: Type 1, subaxial cervical fixation; Type 2, subaxial cervical to proximal thoracic fixation; and Type 3, subaxial cervical to T3/4 cervicothoracic stabilization. All techniques employ O-arm intraoperative navigation and preserve posterior tension-band integrity.

View Article and Find Full Text PDF

Purpose: To evaluate current surgical and prosthetic concepts in adult patients requiring complete tooth extraction for immediate implant placement and restoration in full-arch situations.

Materials And Methods: A narrative review was conducted using a comprehensive search strategy on MEDLINE via PubMed based on the PIO question to identify studies published up to October 2024. Data on treatment strategies, survival and success rates, marginal bone loss (MBL), peri-implant health, patient satisfaction, and quality of life (QoL) were extracted and synthesized.

View Article and Find Full Text PDF

Purpose: This scoping review describes the current evidence on immediate implant placement (IIP) and immediate restoration (IR) in single-unit and multiple-unit fixed dental prostheses regarding anatomical, surgical, and prosthetic factors.

Materials And Methods: Electronic searches were conducted in PubMed, Embase, and Cochrane databases for systematic reviews on IIP and IR in the anterior and posterior regions of the jaws. Bibliometric data and study details were extracted.

View Article and Find Full Text PDF