A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

An in silico analysis of unsteady flow structures in a microaxial blood pump under a pulsating rotation speed. | LitMetric

An in silico analysis of unsteady flow structures in a microaxial blood pump under a pulsating rotation speed.

Comput Methods Programs Biomed

Applied Mechanics Laboratory, Department of Engineering Mechanics, School of Aerospace Engineering, Tsinghua University, Beijing, PR China. Electronic address:

Published: January 2024


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background And Objective: Ventricular assist devices (VADs) are generally designed to perform continuous flow. However, it has been proven that continuous flow, which is not a physiological hemodynamic state, may cause severe complications such as gastrointestinal bleeding, pulmonary hypertension, and ventricular suction. For these reasons, many pulsating blood pump control strategies have been proposed and have the potential for application in percutaneous ventricular assist devices (pVADs) or microaxial blood pumps. A few cases report extra hemolysis when introducing pulsating speed, while none involve blood pumps. This research's primary purpose is to evaluate the potential hemolysis of pVAD under pulsating flow conditions.

Methods: First, the pulsating flow state is deduced using a heart failure model and varying speed. The heart model is established according to the pathology state collected from a clinical check. The rotation speed and boundary physical state are set to fit the heart failure model. The computational fluid dynamics (CFD) method with the hemolysis prediction model is performed. Furthermore, we used proper orthogonal decomposition (POD) analysis to reconstruct the flow field and obtain more details about shearing and transporting effects.

Results: (1) As a variable rotational speed was introduced, no significant gain in hemolysis accumulation appeared in pVAD. This is quite different from long-term implantable VADs. (2) Pulsation affects hemolysis mainly through pressure (or normal stress). Variable rotational speed affects hemolysis mainly through flow instability. (3) Variable rotational speed will increase the instability and influence hemolysis by transporting and shearing effects, while the transporting effect is more significant.

Conclusions: The unsteady flow state will affect the spatial distribution of hemolysis, which should be taken into account during control strategy and impeller shape design.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cmpb.2023.107919DOI Listing