Article Abstract

Background: Pedicle screw loosening and breakage are common causes of revision surgery after lumbar fusion. Thus, there remains a continued need for supplemental fixation options that offer immediate stability without the associated failure modes. This finite element analysis compared the biomechanical properties of a novel cortico-pedicular posterior fixation (CPPF) device with those of a conventional pedicle screw system (PSS).

Methods: The CPPF device is a polyetheretherketone strap providing circumferential cortical fixation for lumbar fusion procedures via an arcuate tunnel. Using a validated finite element model, we compared the stability and load transfer characteristics of CPPF to intact conditions under a 415 N follower load and PSS conditions under a 222 N preload. Depending on the instrumented levels, two different interbody devices were used: a lateral lumbar interbody device at L4-5 or an anterior lumbar interbody device at L5-S1. Primary outcomes included range of motion of the functional spinal units and anterior load transfer, defined as the total load through the disk and interbody device after functional motion and follower load application.

Results: Across all combinations of interbody devices and lumbar levels evaluated, CPPF consistently demonstrated significant reductions in flexion (ranging from 90 to 98%), extension (ranging from 88 to 94%), lateral bending (ranging from 75 to 80%), and torsion (ranging from 77 to 86%) compared to the intact spine. Stability provided by the CPPF device was comparable to PSS in all simulations (range of motion within 0.5 degrees for flexion-extension, 0.6 degrees for lateral bending, and 0.5 degrees for torsion). The total anterior load transfer was higher with CPPF versus PSS, with differences across all tested conditions ranging from 128 to 258 N during flexion, 89-323 N during extension, 135-377 N during lateral bending, 95-258 N during torsion, and 82-250 N during standing.

Conclusion: Under the modeled conditions, cortico-pedicular fixation for supplementing anterior or lateral interbody devices between L4 and S1 resulted in comparable stability based on range of motion measures and less anterior column stress shielding based on total anterior load transfer measures compared to PSS. Clinical studies are needed to confirm these finite element analysis findings.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10636999PMC
http://dx.doi.org/10.1186/s13018-023-04349-5DOI Listing

Publication Analysis

Top Keywords

finite element
16
load transfer
16
element analysis
12
lumbar fusion
12
cppf device
12
interbody devices
12
interbody device
12
range motion
12
anterior load
12
lateral bending
12

Similar Publications

Molecular Plasmonic Cavities.

Nano Lett

September 2025

Department of Physics, Columbia University, New York, New York 10027, United States.

Graphene-based photonic structures have emerged as fertile ground for the controlled manipulation of surface plasmon polaritons (SPPs), providing a two-dimensional platform with low optoelectronic losses. In principle, nanostructuring graphene can enable further confinement of nanolight─enhancing light-matter interactions in the form of SPP cavity modes. In this study, we engineer nanoscale plasmonic cavities composed of self-assembled C arrays on graphene.

View Article and Find Full Text PDF

Microswimmer locomotion in non-Newtonian fluids is crucial for biological processes, including infection, fertilization and biofilm formation. The behaviour of microswimmers in these media is an area with many conflicting results, with swimmers displaying varying responses depending on their morphology, actuation and the complex properties of the surrounding fluid. Using a hybrid computational approach, we numerically investigate the effect of shear-thinning rheology and viscoelasticity on a simple conceptual microswimmer consisting of three linked spheres.

View Article and Find Full Text PDF

A practical approach to determining screw preload in dental implant systems.

J Prosthet Dent

September 2025

Professor, Washington Dental Service Endowed Chair in Dentistry, and Chair, Department of Restorative Dentistry, School of Dentistry, University of Washington, Seattle, Wash.

A practical and novel technique used both in testing and in practice when tightening dental implant screw systems is described for determining the preload implant screw systems, which differs from traditional tightening procedures that are based solely on the application of a predetermined manufacturer specified torque value. Preload is the critical quantity for a reliable joint: about 90% of tightening torque goes into overcoming friction and the remaining amount, approximately 10%, goes into preload. Because of the heavy dependence of torque on friction, the actual preload achieved is subject to large variability of up ±35%, depending on surface conditions (dry, wet, or contaminated).

View Article and Find Full Text PDF

Statement Of Problem: Although custom temporomandibular joint (TMJ) prostheses manufactured via computer-aided design and manufacturing (CAD-CAM) and produced through 3-dimensional (3D) printing or computer numerical control (CNC) allow for sagittal curvature adjustments in the glenoid fossa, their design remains unregulated by the Food and Drug Administration. Consequently, the geometry is determined largely by the engineer's discretion, with limited biomechanical evidence to guide these decisions. The lack of validation regarding how sagittal curvature influences joint stress distribution under various anatomical and functional conditions represents a gap in current knowledge that warrants investigation.

View Article and Find Full Text PDF

Finite Element Analysis of Mandibular Distraction Osteogenesis With a New Partially Bioabsorbable Distractor.

J Craniofac Surg

September 2025

Department of Craniomaxillofacial Surgery, Peking Union Medical College, Chinese Academy of Medical Sciences, Plastic Surgery Hospital, Beijing, China.

Objective: We designed a new distractor pairing a bioabsorbable upper fixing plate fixed by bioabsorbable screws with a traditional titanium distractor to simplify the second surgery removing the distractor after mandibular distraction osteogenesis. The present study aims to evaluate its biomechanical properties using finite element method.

Materials And Methods: Ten computer-aided designed models simulating mandibles of 5 patients under 2 working conditions, the instance of distraction and mastication, were produced.

View Article and Find Full Text PDF