Article Abstract

Background Context: In contrast to cervical discectomy and fusion, total disc replacement (TDR) aims at preserving the motion at the treated vertebral level. Spinal motion is commonly evaluated with the range of motion (ROM). However, more qualitative information about cervical kinematics before and after TDR is still lacking.

Purpose: The aim of this in vitro study was to investigate the influence of cervical TDR on ROM, instantaneous centers of rotation (ICR) and three-dimensional helical axes.

Study Design: An in vitro study with human spine specimens under pure moment loading was conducted to evaluate the kinematics of the intact cervical spine and compare it to cervical TDR.

Methods: Six fresh frozen human cervical specimens (C4-5, median age 28 years, range 19-47 years, two female and four male) were biomechanically characterized in the intact state and after implantation of a cervical disc prosthesis (MOVE-C, NGMedical, Germany). To mimic in vivo conditions regarding temperature and humidity, water steam was used to create a warm and humid test environment with 37°C. Each specimen was quasistatically loaded with pure moments up to ±2.5 Nm in flexion/extension (FE), lateral bending (LB) and axial rotation (AR) in a universal spine tester for 3.5 cycles at 1 °/s. For each third cycle of motion the ROM was evaluated and an established method was used to determine the helical axis and COR and to project them into three planar X-rays. Statistical analysis was conducted using a Friedman-test and post hoc correction with Dunn-Bonferroni-tests (p<.05).

Results: After TDR, total ROM was increased in FE from 19.1° to 20.1°, decreased in LB from 14.6° to 12.6° and decreased in AR from 17.7° to 15.5°. No statistical differences between the primary ROM in the intact condition and ROM after TDR were detected. Coupled rotation between LB and AR were also maintained. The position and orientation of the helical axes after cervical TDR was in good agreement with the results of the intact specimens in all three motion directions. The ICR in FE and AR before and after TDR closely matched, while in LB the ICR after TDR were more caudal. The intact in vitro kinematics we found also resembled in vivo results of healthy individuals.

Conclusion: The results of this in vitro study highlight the potential of artificial cervical disc implants to replicate the quantity as well as the quality of motion of the intact cervical spine.

Clinical Significance: Physiological motion preservation was a driving factor in the development of cervical TDR. Our results demonstrate the potential of cervical TDR to replicate in vivo kinematics in all three motion directions.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.spinee.2023.08.020DOI Listing

Publication Analysis

Top Keywords

range motion
8
motion rom
8
vitro study
8
cervical
7
motion
5
move-c cervical
4
cervical artificial
4
artificial disc
4
disc restore
4
restore intact
4

Similar Publications

Widefield acoustics heuristic: advancing microphone array design for accurate spatial tracking of echolocating bats.

BMC Ecol Evol

September 2025

Lehrstuhl für Zoologie, TUM School of Life Sciences, Technical University of Munich, Liesel-Beckmann Strasse 4, Freising, 85354, Germany.

Accurate three-dimensional localisation of ultrasonic bat calls is essential for advancing behavioural and ecological research. I present a comprehensive, open-source simulation framework-Array WAH-for designing, evaluating, and optimising microphone arrays tailored to bioacoustic tracking. The tool incorporates biologically realistic signal generation, frequency-dependent propagation, and advanced Time Difference of Arrival (TDoA) localisation algorithms, enabling precise quantification of both positional and angular accuracy.

View Article and Find Full Text PDF

Introduction: Military fast jet pilots face significant physical challenges, including high Gz accelerations during dynamic maneuvers. The objectives of this study were threefold: 1) to record pilot movements during real flights, 2) to quantify head and trunk movements under standardized Gz conditions and during basic fighter maneuvers (BFMs), and 3) to categorize compensatory strategies used to mitigate physical strain.

Methods: A total of 20 Eurofighter pilots (mean age: 28.

View Article and Find Full Text PDF

Introduction: This study investigated pilot cognitive engagement patterns across diverse flight conditions using electroencephalography (EEG)-based measurements in a high-fidelity rotary-wing simulation environment.

Methods: A total of 8 experienced U.S.

View Article and Find Full Text PDF

Introduction: The rapidly expanding commercial spaceflight (CSF) market has fueled increasing interest in spaceflight experiences among individuals without professional astronaut qualifications. Such individuals may present with a range of medical conditions that add uncertainties to medical preparation and risk assessment for spaceflight. As the ear, nose, and throat (ENT) working group of the Aerospace Medical Association Ad Hoc Committee on Commercial Spaceflight, we conducted a scoping review to assess the available biomedical literature for ENT and neuro-vestibular conditions and physiology pertinent to spaceflight for nonprofessional space travelers.

View Article and Find Full Text PDF

Sea surface temperature of the Red Sea has increased by up to 0.45 °C per decade over the last 30 years, and coral bleaching events are becoming more frequent. A reef bleaching event was observed in October 2020, whereby some parts of the Red Sea experienced more than 12 °C-weeks.

View Article and Find Full Text PDF