Article Abstract

Background: Due to the wide variety of morphology, size, and dynamics, selecting an optimal valve size and location poses great difficulty in percutaneous pulmonary valve implantation (PPVI). This study aimed to report our experience with in vitro bench testing using patient-specific three-dimensional (3D)-printed models for planning PPVI with the Venus P-valve.

Methods: Patient-specific 3D soft models were generated using PolyJet printing with a compliant synthetic material in 15 patients scheduled to undergo PPVI between July 2018 and July 2020 in Central China Fuwai Hospital of Zhengzhou University.

Results: 3D model bench testing altered treatment strategy in all patients (100%). One patient was referred for surgery because testing revealed that even the largest Venus P-valve would not anchor properly. In the remaining 14 patients, valve size and/or implantation location was altered to avoid valve migration and/or compression coronary artery. In four patients, it was decided to change the point anchoring because of inverted cone-shaped right ventricular outflow tract (RVOT) ( n = 2) or risk of compression coronary artery ( n = 2). Concerning sizing, we found that an oversize of 2-5 mm suffices. Anchoring of the valve was dictated by the flaring of the in- and outflow portion in the pulmonary artery. PPVI was successful in all 14 patients (absence of valve migration, no coronary compression, and none-to-mild residual pulmonary regurgitation [PR]). The diameter of the Venus P-valve in the 3D simulation group was significantly smaller than that of the conventional planning group (36 [2] vs. 32 [4], Z = -3.77, P <0.001).

Conclusions: In vitro testing indicated no need to oversize the Venus P-valve to the degree recommended by the balloon-sizing technique, as 2-5 mm sufficed.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11046019PMC
http://dx.doi.org/10.1097/CM9.0000000000002793DOI Listing

Publication Analysis

Top Keywords

bench testing
12
venus p-valve
12
vitro bench
8
testing patient-specific
8
percutaneous pulmonary
8
pulmonary valve
8
valve implantation
8
valve size
8
valve migration
8
compression coronary
8

Similar Publications

Introduction: Benchtop and animal models have traditionally been used to study the propagation of Onyx Liquid Embolic Systems (Onyx) used in the treatment of brain arteriovenous malformations (AVM). However, such models are costly, do not provide sufficient detail to elucidate how variations in Onyx viscosity alter flow dynamics, and rely on some trial-and-error, resulting in elongated timelines for product development.

Objectives: The goal of this study was to leverage Computational Fluid Dynamics (CFD) simulations to predict the behavior of different Onyx formulations.

View Article and Find Full Text PDF

Neuroprostheses capable of providing Somatotopic Sensory Feedback (SSF) enables the restoration of tactile sensations in amputees, thereby enhancing prosthesis embodiment, object manipulation, balance and walking stability. Transcutaneous Electrical Nerve Stimulation (TENS) represents a primary noninvasive technique for eliciting somatotopic sensations. Devices commonly used to evaluate the effectiveness of TENS stimulation are often bulky and main powered.

View Article and Find Full Text PDF

BackgroundA stable guiding system is essential for successful carotid artery stenting (CAS), particularly when navigating tortuous aortic or supra-aortic anatomy. However, data on the mechanical behavior of stent delivery systems remain scarce.ObjectiveTo assess and compare the bending stiffness and trackability of five commercially available carotid stent delivery systems using bench-top experiments.

View Article and Find Full Text PDF

Objectives: A framework of high-reliability principles was used to identify, investigate, and mitigate infusion pump safety concerns at a large, multisite health care system. We developed a systematic approach to address challenges associated with overinfusions, underinfusions, and the inability to clear upstream occlusion alarms. We identified 112,875 upstream occlusion events for 389,604 infusion starts (failure rate, 29%) within 6 months.

View Article and Find Full Text PDF

Purpose: Hereditary Leiomyomatosis and Renal Cell Cancer (HLRCC) is a rare cancer susceptibility syndrome exclusively attributable to pathogenic variants in FH (HGNC:3700). This paper quantitatively weights the phenotypic context (PP4/PS4) of such very rare variants in FH.

Methods: We collated clinical diagnostic testing data on germline FH variants from 387 individuals with HLRCC and 1,780 individuals with renal cancer, and compared the frequency of 'very rare' variants in each phenotypic cohort against 562,295 population controls.

View Article and Find Full Text PDF