Purpose: To assess the quantification accuracy of pulmonary nodules using virtual monoenergetic images (VMIs) derived from spectral-detector computed tomography (CT) under an ultra-low-dose scan protocol.
Methods: A chest phantom consisting of 12 pulmonary nodules was scanned using spectral-detector CT at 100 kVp/10 mAs, 100 kVp/20 mAs, 120 kVp/10 mAs, and 120 kVp/30 mAs. Each scanning protocol was repeated three times. Each CT scan was reconstructed utilizing filtered back projection, hybrid iterative reconstruction, iterative model reconstruction (IMR), and VMIs of 40-100 keV. The signal-to-noise ratio and air noise of images, absolute differences, and absolute percentage measurement errors (APEs) of the diameter, density, and volume of the four scan protocols and ten reconstruction images were compared.
Results: With each fixed reconstruction image, the four scanning protocols exhibited no significant differences in APEs for diameter and density (all > 0.05). Of the four scan protocols and ten reconstruction images, APEs for nodule volume had no significant differences (all > 0.05). At 100 kVp/10 mAs, APEs for density using IMR were the lowest (APE: 6.69), but no significant difference was detected between VMIs at 50 keV (APE: 11.69) and IMR ( = 0.666). In the subgroup analysis, at 100 kVp/10 mAs, there were no significant differences between VMIs at 50 keV and IMR in diameter and density (all > 0.05). The radiation dose at 100 kVp/10 mAs was reduced by 77.8% compared with that at 120 kVp/30 mAs.
Conclusion: Compared with IMR, reconstruction at 100 kVp/10 mAs and 50 keV provides a more accurate quantification of pulmonary nodules, and the radiation dose is reduced by 77.8% compared with that at 120 kVp/30 mAs, demonstrating great potential for ultra-low-dose spectral-detector CT.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10679552 | PMC |
http://dx.doi.org/10.4274/dir.2023.232233 | DOI Listing |
Radiol Med
June 2025
Department of Neuroscience, Imaging and Clinical Sciences, "G. d'Annunzio" University of Chieti-Pescara, Chieti, Italy.
Aims: CT acquisition parameters and reconstruction techniques may affect the accuracy of calcium scoring measurements with a potential impact on clinical decision making. We evaluated the agreement of half- versus standard-dose protocols for assessment of aortic valve (AVCS HD and AVCS SD protocols) and coronary artery calcium scoring (CACS HD and CACS SD protocols) with and without the application of iterative reconstruction.
Methods: We enrolled 144 consecutive patients (mean age 83 ± 9 years) with known aortic stenosis undergoing 128-row prospective sequential CT with standard (120 kVp/20 mAs) and half-dose (120 kVp/10 mAs) protocols for both AVCS and CACS evaluation.
Insights Imaging
January 2025
Department of Medical Imaging Center, Nanfang Hospital, Southern Medical University, Guangzhou, 510515, China.
Objectives: To investigate the image quality and diagnostic performance with ultra-low dose dual-layer detector spectral CT (DLSCT) by various reconstruction techniques for evaluation of pulmonary nodules.
Materials And Methods: Between April 2023 and December 2023, patients with suspected pulmonary nodules were prospectively enrolled and underwent regular-dose chest CT (RDCT; 120 kVp/automatic tube current) and ultra-low dose CT (ULDCT; 100 kVp/10 mAs) on a DLSCT scanner. ULDCT was reconstructed with hybrid iterative reconstruction (HIR), electron density map (EDM), and virtual monoenergetic images at 40 keV and 70 keV.
Diagn Interv Radiol
September 2023
Department of Radiology, Second Affiliated Hospital of Naval Medical University, Shanghai, China.
Purpose: To assess the quantification accuracy of pulmonary nodules using virtual monoenergetic images (VMIs) derived from spectral-detector computed tomography (CT) under an ultra-low-dose scan protocol.
Methods: A chest phantom consisting of 12 pulmonary nodules was scanned using spectral-detector CT at 100 kVp/10 mAs, 100 kVp/20 mAs, 120 kVp/10 mAs, and 120 kVp/30 mAs. Each scanning protocol was repeated three times.
Korean J Radiol
July 2019
Department of Radiology, Anam Hospital, Korea University College of Medicine, Seoul, Korea.
Objective: To investigate the accuracy of model-based iterative reconstruction (MIR) for volume measurement of part-solid nodules (PSNs) and solid nodules (SNs) in comparison with filtered back projection (FBP) or hybrid iterative reconstruction (HIR) at various radiation dose settings.
Materials And Methods: CT scanning was performed for eight different diameters of PSNs and SNs placed in the phantom at five radiation dose levels (120 kVp/100 mAs, 120 kVp/50 mAs, 120 kVp/20 mAs, 120 kVp/10 mAs, and 80 kVp/10 mAs). Each CT scan was reconstructed using FBP, HIR, or MIR with three different image definitions (body routine level 1 [IMR-R1], body soft tissue level 1 [IMR-ST1], and sharp plus level 1 [IMR-SP1]; Philips Healthcare).
Eur J Radiol
April 2015
Department of Radiology, Seoul National University Hospital, Republic of Korea; Institute of Radiation Medicine, Seoul National University Hospital, Republic of Korea.
Purpose: To assess the effect of different reconstruction algorithms on computer-aided diagnosis (CAD) performance in ultra-low-dose CT colonography (ULD CTC).
Materials And Methods: IRB approval and informed consents were obtained. Thirty prospectively enrolled patients underwent non-contrast CTC at 120 kVp/10 mAs in supine and 100 kVp/10 mAs in prone positions, followed by same-day colonoscopy.