Objective: To compare the effectiveness of disinfection protocols utilizing a ultraviolet (UV) Smart D60 light system with Impelux™ technology with a standard Cidex ortho-phthalaldehyde (OPA) disinfection protocol for cleaning flexible fiberoptic laryngoscopes (FFLs).
Methods: Two hundred FFLs were tested for bacterial contamination after routine use, and another 200 FFLs were tested after disinfection with one of four methods: enzymatic detergent plus Cidex OPA (standard), enzymatic detergent plus UV Smart D60, microfiber cloth plus UV Smart D60, and nonsterile wipe plus UV Smart D60. Pre- and post-disinfection microbial burden levels and positive culture rates were compared using Kruskal-Wallis ANOVA and Fisher's two-sided exact, respectively.
Results: After routine use, approximately 56% (112/200) of FFLs were contaminated, with an average contamination level of 9,973.7 ± 70,136.3 CFU/mL. The standard reprocessing method showed no positive cultures. The enzymatic plus UV, microfiber plus UV, and nonsterile wipe plus UV methods yielded contamination rates of 4% (2/50), 6% (3/50), and 12% (6/50), respectively, with no significant differences among the treatment groups (p > 0.05). The pre-disinfection microbial burden levels decreased significantly after each disinfection technique (p < 0.001). The average microbial burden recovered after enzymatic plus UV, microfiber plus UV, and nonsterile wipe plus UV were 0.40 CFU/mL ± 2, 0.60 CFU/mL ± 2.4, and 12.2 CFU/mL ± 69.5, respectively, with no significant difference among the treatment groups (p > 0.05). Micrococcus species (53.8%) were most frequently isolated, and no high-concern organisms were recovered.
Conclusion: Disinfection protocols utilizing UV Smart D60 were as effective as the standard chemical disinfection protocol using Cidex OPA.
Level Of Evidence: NA Laryngoscope, 133:3512-3519, 2023.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/lary.30869 | DOI Listing |
Front Plant Sci
March 2025
Honda Biotech. Laboratory, Utsunomiya-Shi, Tochigi, Japan.
Background: Breeding rice varieties that are resilient to climate change and optimizing the cultivation methods for each developed variety are challenging issues in global food demands.
Methods: In this study, the late-flowering gene of Koganebare was introduced into Koshihikari through backcrossing to create 'Koshihikari Hd16'. It was then crossed with 'Koshihikari d60' to develop an isogenic line of Koshihikari containing both and .
Laryngoscope
December 2023
Department of Otolaryngology-Head and Neck Surgery, NYU Langone Health, New York, New York, USA.
Objective: To compare the effectiveness of disinfection protocols utilizing a ultraviolet (UV) Smart D60 light system with Impelux™ technology with a standard Cidex ortho-phthalaldehyde (OPA) disinfection protocol for cleaning flexible fiberoptic laryngoscopes (FFLs).
Methods: Two hundred FFLs were tested for bacterial contamination after routine use, and another 200 FFLs were tested after disinfection with one of four methods: enzymatic detergent plus Cidex OPA (standard), enzymatic detergent plus UV Smart D60, microfiber cloth plus UV Smart D60, and nonsterile wipe plus UV Smart D60. Pre- and post-disinfection microbial burden levels and positive culture rates were compared using Kruskal-Wallis ANOVA and Fisher's two-sided exact, respectively.
Pharmaceuticals (Basel)
August 2022
Center for Bioinformatics and Integrative Biology (CBIB), Facultad de Ciencias de la Vida, Universidad Andres Bello, Av. República 330, Santiago 8370146, Chile.
The rapid emergence and spread of new variants of coronavirus type 2, as well as the emergence of zoonotic viruses, highlights the need for methodologies that contribute to the search for new pharmacological treatments. In the present work, we searched for new SARS-CoV-2 papain-like protease inhibitors in the PubChem database, which has more than 100 million compounds. Based on the ligand efficacy index obtained by molecular docking, 500 compounds with higher affinity than another experimentally tested inhibitor were selected.
View Article and Find Full Text PDFEur Arch Otorhinolaryngol
October 2021
Department of Otolaryngology, Head and Neck Surgery, University Hospital Marburg, Philipps-Universität Marburg, Baldingerstrasse, 35043, Marburg, Germany.
Background: Reprocessing of flexible endoscopes (FEs) is often expensive, time consuming, and becomes increasingly complex, due to rising demands of hygiene. After beneficial results in reprocessing of rigid endoscopes using Impelux™ UV-C light technology, we tested the same method for reprocessing of FEs without working channel.
Materials And Methods: Testing was performed on FEs without working channel after routine clinical use (transnasal flexible endoscopy).