Article Abstract

Background And Purpose: Treating aneurysms with intra-saccular flow disruption is a feasible alternative to coil-embolization. Besides the established WEB device, the novel Contour Neurovascular System has emerged as a potentially easier alternative regarding sizing and deployment. We report the learning curve experienced at our center from the first 48 patients treated with Contour and compared it with 48 consecutive WEB cases.

Methods: Both groups were compared concerning intervention time, sizing failures leading to device changes and radiation dose. Additionally, we analyzed potential learning effects by comparing the first 24 Contour cases with our last 24 Contour cases and WEB cases respectively.

Results: Patient demographics, acute vs. incidental cases and aneurysm localization were comparable in both groups. The deployment time was faster in our 48 Contour cases (median: 22.0 ± 17.0 min), than in the WEB group (median: 27.5 ± 24.0 min). Total intervention time was similar for Contour (median: 68.0 ± 46.9 min) and WEB cases (median: 69.0 ± 38.0 min). Device implantation times in our WEB cases were slightly shorter in the later cases (median: 25.5 ± 24.1 min) than in the earlier (median: 28.0 ± 24.4 min) cases. In the Contour cohort, deployment times were similar for the first 24 cases (median: 22.0 ± 14.5 min) and the final 24 (median: 22.0 ± 19.4 min). Radiation dose was lower in the Contour group (1469.0 ± 1718 mGy*cm vs. 1788.0 ± 1506 mGy*cm using the WEB device). Less intra-procedural device changes were performed in the Contour cohort (6 of 48 cases, 12.5%), than in the WEB group (8 of 48 cases, 16.7%).

Conclusion: Aneurysm occlusion times and consequently radiation doses, as well as the amount of device changes were lower in the Contour group. Occlusion times did not differ in the first and last 24 Contour cases, leading to the assumption that the handling of Contour does not require extended training. A short training effect in occlusion times was noted, however, between the first and last WEB cases as shorter procedure times were seen in the latter cases.

Download full-text PDF

Source
http://dx.doi.org/10.1177/15910199231179512DOI Listing

Publication Analysis

Top Keywords

contour cases
16
web cases
16
cases median
16
cases
15
contour
13
device changes
12
occlusion times
12
web
10
web device
8
intervention time
8

Similar Publications

Giant cell tumor (GCT) of the bone, although benign, demonstrates local aggressiveness, a potential for recurrence, and, in rare instances, malignant transformation. Functional preservation is crucial in cases involving the articular surface, often utilizing the Sandwich Technique. We propose an enhanced reconstruction method using the inner table of the iliac crest in a reverse fashion, offering a more anatomically contoured proximal tibial plateau and reducing donor site morbidity compared to tricortical iliac crest grafting.

View Article and Find Full Text PDF

In therapy with Synchrony® mounted on Radixact®, the fiducial marker (FM) and adrenal gland metastasis, which shift with respiratory phase, require margin compensation for high-dose prescriptions. Although compensation is critical, no studies have examined the margin to compensate for the respiratory phase shift. Therefore, we aimed to suggest the compensating margin for the FM and adrenal metastasis shift along with respiratory phase.

View Article and Find Full Text PDF

Background: Low-field MRI provides superior soft-tissue contrast compared to CT while costing significantly less than high-field MRI, which makes it a more accessible option for MRI-guided radiation therapy planning. Four-dimensional MRI (4D-MRI) is a technique that has been increasingly adopted clinically for internal target volume (ITV) delineation in free-breathing liver radiotherapy planning, and it requires high spatial resolution and accurate respiratory phase differentiation to enable precise dose planning. The feasibility of 4D-MRI at low-field strength, specifically at 0.

View Article and Find Full Text PDF

Introduction: Gynecomastia is a common condition characterized by male breast enlargement, which can have a profound psychological impact on affected individuals. Surgical intervention is often sought to alleviate these concerns. This study evaluates the lateral port technique for gynecomastia surgery, which combines power-assisted liposuction (PAL) and ultrasound-assisted liposuction (UAL) to achieve optimal surgical outcomes with no scar on front of chest.

View Article and Find Full Text PDF