Classification of "Near-patient" and "Point-of-Care" SARS-CoV-2 Nucleic Acid Amplification Test Systems and a first approach to evaluate their analytical independence of operator activities.

J Clin Virol

INSTAND e.V. Society for Promoting Quality Assurance in Medical Laboratories, Düsseldorf, Germany; IQVD GmbH, Institut für Qualitätssicherung in der Virusdiagnostik, Berlin, Germany; GBD Gesellschaft für Biotechnologische Diagnostik mbH, Berlin, Germany.

Published: August 2023


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: European legislation defines as "near-patient testing" (NPT) what is popularly and in other legislations specified as "point-of-care testing" (POCT). Systems intended for NPT/POCT use must be characterized by independence from operator activities during the analytic procedure. However, tools for evaluating this are lacking. We hypothesized that the variability of measurement results obtained from identical samples with a larger number of identical devices by different operators, expressed as the method-specific reproducibility of measurement results reported in External Quality Assessment (EQA) schemes, is an indicator for this characteristic.

Materials And Methods: Legal frameworks in the EU, the USA and Australia were evaluated about their requirements for NPT/POCT. EQA reproducibility of seven SARS-CoV-2-NAAT systems, all but one designated as "POCT", was calculated from variabilities in Ct values obtained from the respective device types in three different EQA schemes for virus genome detection.

Results: A matrix for characterizing test systems based on their technical complexity and the required operator competence was derived from requirements of the European In Vitro Diagnostic Regulation (IVDR) 2017/746. Good EQA reproducibility of the measurement results of the test systems investigated implies that different users in different locations have no recognizable influence on their measurement results.

Conclusion: The fundamental suitability of test systems for NPT/POCT use according to IVDR can be easily verified using the evaluation matrix presented. EQA reproducibility is a specific characteristic indicating independence from operator activities of NPT/POCT assays. EQA reproducibility of other systems than those investigated here remains to be determined.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10249340PMC
http://dx.doi.org/10.1016/j.jcv.2023.105521DOI Listing

Publication Analysis

Top Keywords

test systems
16
eqa reproducibility
16
independence operator
12
operator activities
12
reproducibility measurement
8
eqa schemes
8
systems investigated
8
systems
7
eqa
6
reproducibility
5

Similar Publications

Tuning the Electronic Structure in the MoS/SrTiO Heterojunction via Phase Evolution of the SrTiO Substrate.

ACS Nano

September 2025

Department of Chemical Physics, Key Laboratory of Precision and Intelligent Chemistry, University of Science and Technology of China, Hefei, Anhui 230026, China.

The coupling between transition metal dichalcogenides (TMDCs) and SrTiO has recently emerged as a fertile platform for discovering interfacial phenomena, where particle interactions, lattice coupling, and dielectric screening give rise to interesting physical effects. These hybrid systems hold significant promise for two-dimensional (2D) electronics, ferroelectric state control, and metastable phase engineering. However, effective modulation of the interfacial electronic structure remains a critical challenge.

View Article and Find Full Text PDF

Preclinical stroke research faces a critical translational gap, with animal studies failing to reliably predict clinical efficacy. To address this, the field is moving toward rigorous, multicenter preclinical randomized controlled trials (mpRCTs) that mimic phase 3 clinical trials in several key components. This collective statement, derived from experts involved in mpRCTs, outlines considerations for designing and executing such trials.

View Article and Find Full Text PDF

Introduction: Anxiety and stress are prevalent mental health issues. Traditional drug treatments often come with unwanted side effects and may not produce the desired results. As an alternative, probiotics are being used as a treatment option due to their lack of specific side effects.

View Article and Find Full Text PDF

This study was designed to investigate the switch between the open-source automated insulin delivery (OS-AID) system AndroidAPS (AAPS) and commercially available AID systems Control-IQ (CIQ) and MiniMed 780G (780G) conducted in a new extended follow-up study. In this prospective open-label single-arm clinical trial, 41 adults with type 1 diabetes (age 35 ± 11 years, glycated hemoglobin [HbA1c] 6.4 ± 2.

View Article and Find Full Text PDF