Article Abstract

Background: The Harmony transcatheter pulmonary valve (TPV) is the first U.S. Food and Drug Administration-approved device for severe pulmonary regurgitation (PR) in the native or surgically repaired right ventricular outflow tract (RVOT).

Objectives: One-year safety and effectiveness of the Harmony TPV were evaluated in patients from the Harmony Native Outflow Tract Early Feasibility Study, Harmony TPV Pivotal Study, and Continued Access Study, representing the largest cohort to date of Harmony TPV recipients.

Methods: Eligible patients had severe PR by echocardiography or PR fraction ≥ 30% by cardiac magnetic resonance imaging and clinical indications for pulmonary valve replacement. The primary analysis included 87 patients who received a commercially available TPV22 (n = 42) or TPV25 (n = 45) device; 19 patients who received an early device iteration prior to its discontinuation were evaluated separately.

Results: In the primary analysis, median patient age at treatment was 26 years (IQR: 18-37 years) in the TPV22 group and 29 years (IQR: 19-42 years) in the TPV25 group. At 1 year, there were no deaths; 98% of TPV22 and 91% of TPV25 patients were free from the composite of PR, stenosis, and reintervention (moderate or worse PR, mean RVOT gradient >40 mmHg, device-related RVOT reoperation, and catheter reintervention). Nonsustained ventricular tachycardia occurred in 16% of patients. Most patients had none/trace or mild PR (98% of TPV22 patients, 97% of TPV25 patients). Outcomes with the discontinued device are reported separately.

Conclusions: The Harmony TPV device demonstrated favorable clinical and hemodynamic outcomes across studies and valve types through 1 year. Further follow-up will continue to assess long-term valve performance and durability.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcin.2023.03.002DOI Listing

Publication Analysis

Top Keywords

harmony tpv
16
pulmonary valve
12
patients
9
cohort harmony
8
harmony transcatheter
8
transcatheter pulmonary
8
outflow tract
8
primary analysis
8
patients received
8
years iqr
8

Similar Publications

Background: The Harmony transcatheter pulmonary valve (TPV) is designed to treat severe pulmonary regurgitation in the native or surgically repaired right ventricular (RV) outflow tract. Early outcomes after TPV replacement with the Harmony valve have been positive, but longer-term data are limited.

Methods: The study included patients who received a commercially available TPV22 or TPV25 device as part of the nonrandomized, prospective Native Outflow Tract Early Feasibility Study, Harmony Pivotal Trial, and Continued Access Study.

View Article and Find Full Text PDF

The Harmony™ transcatheter pulmonary valve (TPV) is specifically designed to treat severe pulmonary regurgitation in a native or surgically repaired right ventricular outflow tract. To date, no reports have described TPV implantation (TPVI) in patients with prior aortic valve replacement (AVR) using a mechanical valve. Therefore, the risks of technical challenges or complications, particularly coronary compression or device fracture associated with prior AVR, remain unknown.

View Article and Find Full Text PDF

Pulmonary disfunction is frequent in repaired congenital heart diseases. Both pulmonary regurgitation and pulmonary stenosis are possible complications over time. In the past, the surgical approach was the only feasible management but exposed the patient to a redo-surgery with its consequent risks.

View Article and Find Full Text PDF

Transcatheter pulmonary valve replacement is the first choice to treat residual or recurrent right ventricular outflow tract dysfunction. Surgery is an effective option when anatomy is not permissive for transcatheter procedures. When surgical risk is too high, hybrid procedures might be considered.

View Article and Find Full Text PDF

Acute Left Main Coronary Artery Compression After Harmony-TPV Implantation.

Catheter Cardiovasc Interv

April 2025

Toronto Congenital Cardiac Centre for Adults, Peter Munk Cardiac Centre, University Health Network, University of Toronto, Toronto, Ontario, Canada.

We present an 18-year-old male with Fallot's tetralogy (ToF) repaired with a transannular and main pulmonary artery (PA) patch, who suffered an acute coronary artery compression after a Harmony-TPV deployment. Left main coronary artery compression was confirmed with angiography and IVUS. He underwent emergent surgery, with a bioprosthetic valve replacement.

View Article and Find Full Text PDF