Article Abstract

Unlabelled: is to determine physical and structural properties of woven synthetic prostheses depending on the type of the weave.

Materials And Methods: Ten vascular prostheses manufactured at the Science and Technology Park of the BNTU "Polytechnic" (Minsk, Republic of Belarus) have been analyzed. The prostheses differed in the type of weaving, duration and temperature of thermal fixation during crimping. Three samples had a single-layer structure and 7 samples had a double-layer structure. Tests for water permeability, resistance to radial bending, and porosity of the prostheses have been performed.

Results: The single-layer woven prostheses have demonstrated a low level of water permeability: the best result was shown by sample No.1: 80 [77.1; 80.5] ml/min/cm. A strong direct correlation was revealed for these prostheses: the larger the pore diameter, the greater permeability (r=0.778; p=0.05). The single-layer woven prostheses appeared to be most resistant to radial bending, samples No.1 and 3 had no deformations at the minimum radius of the cylinder (r<4 mm), sample No.2 showed deformation on the cylinder with r=5 mm. For the single-layer prostheses, a strong negative correlation was noted (r=‒0.97; p=0.04) between the density of the warp threads and the kinking radius.All double-layer prostheses have demonstrated higher water permeability and weak resistance to deformation during radial bending. Samples No.4 and 8 were found to have minimum and maximum water permeability of 276.5 [258.3; 288.4] and 538.8 [533.3; 564.3] ml/min/cm, respectively. The minimum kinking radius (7 mm) was shown by samples No.9 and 10. The worst results were demonstrated by sample No.6, which was deformed with minimal bending.

Conclusion: Samples with ordinary plain weave have a low level of water permeability and high resistance to radial deformation, which makes them look most promising for the application in vascular surgery.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10171052PMC
http://dx.doi.org/10.17691/stm2022.14.6.01DOI Listing

Publication Analysis

Top Keywords

prostheses
8
vascular prostheses
8
water permeability
8
radial bending
8
single-layer woven
8
woven prostheses
8
influence weaving
4
weaving technologies
4
technologies integral
4
integral characteristics
4

Similar Publications

This study assessed the effect of saliva exposure on roughness (Ra) and Vickers hardness (VHN) of two direct restorative materials, enamel, and dentin adjacent to the restorations. Enamel and dentin cavities in molars (n = 10) were restored with a) bulk-fill resin composite (Tetric N-Flow Bulk Fill, BF) with the application of a universal adhesive (Tetric N-Bond Universal) and b) alkasite restorative material (Cention N, CN) with and without the application of a universal adhesive. After 24 h (baseline), surface roughness and hardness of the restorative material and dental tissues were assessed at 100 μm from the tooth/restoration interface.

View Article and Find Full Text PDF

This study aimed to evaluate the longitudinal effect of dentition status on the perceived mobility limitation of community-dwelling Brazilian older adults. This cohort study used data from individuals who participated in the second (2006), third (2010), and fourth (2015) waves of the Health Well-being and Aging Study, conducted in the urban region of the city of São Paulo, Brazil, with adults aged 60 years and older. Mobility limitation was assessed in all waves according to reports of difficulty in performing seven activities, with higher scores representing a higher number of limitations.

View Article and Find Full Text PDF

This study evaluated the influence of a customized healing abutment (CHA) placed on immediate implants. It also assessed bone ridge volume, keratinized mucosal collar, and postoperative pain. Thirty-one patients needing tooth extraction and immediate implant were selected.

View Article and Find Full Text PDF

This study aimed to histomorphometrically evaluate the effect of guided bone regeneration (GBR) and two implant surfaces on the thickness and height of newly formed bone in dehiscence defects around titanium implants. Three premolars and the first bilateral molar were extracted from ten adult mongrel dogs, and 40 buccal bone dehiscences measuring 5 mm in height and 4 mm in width were created using a University of North Carolina (UNC) periodontal probe to confirm the dimensions. Forty implants were randomly assigned to one of four groups: oxidized implant surfaces (OIS, n = 10), turned/machined implant surfaces (TIS, n = 10), OIS + GBR (n = 10), and TIS + GBR (n = 10).

View Article and Find Full Text PDF

The human auditory system must distinguish relevant sounds from noise. Severe hearing loss can be treated with cochlear implants (CIs), but how the brain adapts to electrical hearing remains unclear. This study examined adaptation to unilateral CI use in the first and seventh months after CI activation using speech comprehension measures and electroencephalography recordings, both during passive listening and an active spatial listening task.

View Article and Find Full Text PDF