Article Abstract

Purpose: A fully digital approach to oral prosthodontic rehabilitation requires the possibility of combining (i.e., registering) digital documentation from different sources. This becomes more complex in an edentulous jaw, as fixed dental markers to perform reliable registration are lacking. This validation study aimed to evaluate the reproducibility of 1) intraoral scanning and 2) soft tissue-based registration of an intraoral scan with a cone-beam computed tomography (CBCT) scan for a fully edentulous upper jaw.

Materials And Methods: Two observers independently performed intraoral scans of the upper jaw in 14 fully edentulous patients. The palatal vault of both surface models was aligned, and the inter-observer variability was assessed by calculating the mean inter-surface distance at the level of the alveolar crest. Additionally, a CBCT scan of all patients was obtained and a soft tissue surface model was generated using patient-specific gray values. This CBCT soft tissue model was registered with the intraoral scans of both observers, and the intraclass correlation coefficient (ICC) was calculated to evaluate the reproducibility of the registration method.

Results: The mean inter-observer deviation when performing an intraoral scan of the fully edentulous upper jaw was 0.10 ± 0.09 mm. The inter-observer agreement for the soft tissue-based registration method was excellent (ICC=0.94; 95% confidence interval, 0.81-0.98).

Conclusion: Even when teeth are lacking, intraoral scanning of the jaw and soft tissue-based registration of an intraoral scan with a CBCT scan can be performed with a high degree of precision.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10060752PMC
http://dx.doi.org/10.5624/isd.20220131DOI Listing

Publication Analysis

Top Keywords

intraoral scan
16
intraoral scanning
12
registration intraoral
12
soft tissue-based
12
tissue-based registration
12
cbct scan
12
fully edentulous
12
intraoral
9
edentulous jaw
8
validation study
8

Similar Publications

Enhancing the accuracy of full-arch digital implant impressions via segmentation: An in vitro study.

J Prosthodont

September 2025

Department of Graphic Design and Engineering Projects, Faculty of Engineering Gipuzkoa, University of the Basque Country UPV/EHU, San Sebastian, Spain.

Purpose: The aim of this in vitro study was to assess whether the accuracy (trueness and precision) of full-arch implant digital impressions can be improved by replacing standard digital impressions (StDIs) with segmented digital impressions (SgDIs), obtained by superimposing short-span digital impressions using a geometric pattern.

Materials And Methods: The study was conducted using a commercial edentulous maxillary model with four implants-positioned at the sites of the canines and second molars-and four commercial intraoral scan bodies (ISBs) screwed to them. Ten StDIs and SgDIs of the model were obtained, and their accuracy was evaluated and compared in terms of trueness and precision, based on measurements of three distances and three angles between the ISBs.

View Article and Find Full Text PDF

Purpose: To present a novel digital workflow (the Columbus Digital Bridge Protocol) for immediately loaded full-arch rehabilitations, integrating digital technologies throughout diagnostic, surgical and prosthetic phases, with a focus on the application of intraoral photogrammetry scanning.

Materials And Methods: The workflow presented in this article, successfully implemented in 14 patients, includes standardised clinical steps: digital diagnostic planning through matching of facial scans and CBCT data, surgical placement of four implants following tooth extraction, immediate post-surgical intraoral photogrammetry scanning using a three-step procedure (i.e.

View Article and Find Full Text PDF

The aim of the study is to present a case of a dermoid cyst in the oral floor of an 80-year-old female patient and discuss its clinical presentation, diagnostic approach, and treatment. A retrospective case study was conducted at the Department of Maxillofacial Surgery, Aulss 2 Marca Trevigiana. The patient, with a medical history of hypertension, presented with a swelling in the oral floor.

View Article and Find Full Text PDF

Purpose: Designing restorations remains challenging because the process is time-consuming and requires operator skill and experience. This clinical study evaluated the fit accuracy of polymerized complete crowns fabricated using a web-based 3D generative artificial intelligence design (GAID) method compared to crowns fabricated using a conventional computer-aided design (CCAD) method.

Materials And Methods: Sixty-two patients requiring complete crowns in maxillary and mandibular premolars and molars were enrolled.

View Article and Find Full Text PDF

Purpose: This study aimed to evaluate the effect of clinically practical auxiliary geometric devices (AGD) on measurement trueness and scan time in full-arch implant-supported prostheses, focusing on different intraoral scanners (IOS) and implant angulations.

Materials And Methods: Four implants were planned in an edentulous maxillary arch and divided into two groups based on posterior implant angulation: Model A (Parallel) and Model B (30°). Each model was evaluated under three auxiliary geometric devices (AGD) application types (std, agd1, agd2), and scanned using three different intraoral scanners (IOSs) [3Shape Trios 3 (T), Medit i700 (M), and Cerec Primescan (PS)], resulting in nine groups per model (n = 10).

View Article and Find Full Text PDF