Importance: Infection transmission following endoscopic retrograde cholangiopancreatography (ERCP) can occur due to persistent contamination of duodenoscopes despite high-level disinfection to completely eliminate microorganisms on the instrument.
Objective: To determine (1) contamination rates after high-level disinfection and (2) technical performance of duodenoscopes with disposable elevator caps compared with those with standard designs.
Design, Setting, And Participants: In this parallel-arm multicenter randomized clinical trial at 2 tertiary ERCP centers in Canada, all patients 18 years and older and undergoing ERCP for any indication were eligible.
Intervention: The intervention was use of duodenoscopes with disposable elevator caps compared with duodenoscopes with a standard design.
Main Outcomes And Measures: Coprimary outcomes were persistent microbial contamination of the duodenoscope elevator or channel, defined as growth of at least 10 colony-forming units of any organism or any growth of gram-negative bacteria following high-level disinfection (superiority outcome), and technical success of ERCP according to a priori criteria (noninferiority outcome with an a priori noninferiority margin of 7%), assessed by blinded reviewers.
Results: From December 2019 to February 2022, 518 patients were enrolled (259 disposable elevator cap duodenoscopes, 259 standard duodenoscopes). Patients had a mean (SD) age of 60.7 (17.0) years and 258 (49.8%) were female. No significant differences were observed between study groups, including in ERCP difficulty. Persistent microbial contamination was detected in 11.2% (24 of 214) of standard duodenoscopes and 3.8% (8 of 208) of disposable elevator cap duodenoscopes (P = .004), corresponding to a relative risk of 0.34 (95% CI, 0.16-0.75) and number needed to treat of 13.6 (95% CI, 8.1-42.7) to avoid persistent contamination. Technical success using the disposable cap scope was noninferior to that of the standard scope (94.6% vs 90.7%, P = .13). There were no differences between study groups in adverse events and other secondary outcomes.
Conclusions And Relevance: In this randomized clinical trial, disposable elevator cap duodenoscopes exhibited reduced contamination following high-level disinfection compared with standard scope designs, without affecting the technical performance and safety of ERCP.
Trial Registration: ClinicalTrials.gov Identifier: NCT04040504.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9871945 | PMC |
http://dx.doi.org/10.1001/jamainternmed.2022.6394 | DOI Listing |
Am J Gastroenterol
February 2025
Center for Endoscopic Research and Therapeutics, University of Chicago, Chicago, Illinois, USA.
Introduction: There are limited data comparing the contamination rates of reusable and novel duodenoscopes with disposable components. The clinical implications of novel duodenoscope bacterial contamination are also unknown. This study aimed to evaluate the rate of pathogenic bacterial contamination of novel and reusable duodenoscopes and assess the clinical implications for patients exposed to contaminated duodenoscopes.
View Article and Find Full Text PDFJ Gastroenterol Hepatol
February 2025
Division of Gastroenterology, Department of Medicine, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
Background And Aim: To reduce bacterial contamination after reprocessing, various new designs of duodenoscopes have been developed to better expose the elevator complex for cleaning. We compared the rates of bacterial contamination and organic residue in disposable distal cap duodenoscopes and detachable elevator duodenoscopes after manual cleaning and high-level disinfection (HLD), as well as their cost-effectiveness.
Methods: A total of 162 duodenoscopes were randomly assigned to either Group A (disposable distal caps; n = 81) or Group B (detachable elevator; n = 81).
J Cutan Aesthet Surg
December 2022
Mukhtar Skin Centre, KMCH Road, Katihar, Bihar, India.
Ingrown toenails is a painful foot problem seen in primary care among young men. To provide immediate pain relief, some types of splints are placed in the gutter to separate it from the ingrown lateral nail. There are time-consuming and effort-intensive procedures that use available nail separators and elevators to separate the nail plate.
View Article and Find Full Text PDFDent J (Basel)
June 2024
Department of Bioengineering, R-Crio Criogenia S.A., Campinas 13098-324, Brazil.
The aim of this study is to validate a minimally invasive surgical procedure to harvest palate periosteum as a source of tissue for mesenchymal stromal/stem cells. We performed a standardized procedure to harvest the palate periosteum in ten subjects, which consisted of a 3 mm disposable punch and a Molt periosteal elevator to harvest a small full-thickness fragment of soft tissue at the hard palate area, between the upper bicuspids, 3 to 4 mm apical to the cement enamel junction. The one-third inner portion was fragmented, and following standard cell culture procedures, the adherent cells were cultured for three passages, after obtaining 70-90% confluence.
View Article and Find Full Text PDFGastroenterology
September 2023
Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, Minnesota.