A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Effects of different ablation settings on lesion dimensions in an ex vivo swine heart model: Baseline impedance, irrigant, and electrode configuration. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Introduction: Intramural or epicardial locations of the arrhythmogenic substrate are regarded as one of the main reasons for radiofrequency (RF) catheter ablation failure. This study aims to conduct a comprehensive analysis of various factors including baseline impedance, irrigant and electrode configuration at similar ablation index (AI) value.

Methods: In 12 ex vivo swine hearts, RF ablation was performed at a target AI value of 500 and a multistep impedance load (100-180 Ω) in 4 settings: (1) conventional unipolar configuration with an irrigant of normal saline (NS); (2) conventional unipolar configuration with an irrigant of half normal saline (HNS); (3) bipolar configuration with an irrigant of NS; (4) sequential unipolar configuration with an irrigant of NS. The relationships between lesion dimensions and above factors were examined.

Results: Baseline impedance had a strong negative linear correlation with lesion dimensions at a certain AI. The correlation coefficient between baseline impedance and depth, width, and volume were R = -0.890, R = -0.755 and R = -0.813, respectively (p < .01). There were 10 (total: 10/100, 10%; bipolar: 10/25, 40%) transmural lesions during the whole procedure. Bipolar ablation resulted in significantly deeper lesion than other electrode configurations. Other comparisons in our experiment did not achieve statistical significance.

Conclusion: There is a strong negative linear correlation between baseline impedance and lesion dimensions at a certain AI value. Baseline impedance has an influence on the overall lesion dimensions among irrigated fluid and ablation configurations. Over a threshold impedance of 150 Ω, the predictive accuracy of AI can be compromised.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.15752DOI Listing