Feasibility of facial nerve monitoring using adhesive surface electrodes during parotidectomy: a comparative study with needle electrodes.

Eur Arch Otorhinolaryngol

Department of Otorhinolaryngology-Head and Neck Surgery, Pusan National University School of Medicine, Pusan National University and Biomedical Research Institute, Pusan National University Hospital, 179, Gudeok-Ro, Seo-Gu, Busan, 49241, Republic of Korea.

Published: April 2023


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: Electrophysiological monitoring of the facial nerve during parotidectomy has been reported as an adjunctive method to prevent facial nerve injury. Classically, a needle electrode is used to obtain electromyographic (EMG) signals from facial muscles during facial nerve monitoring (FNM) of parotid surgery, likewise adhesive surface electrodes. This study aimed to investigate the feasibility of performing FNM with surface electrodes during parotid surgery and to compare EMG values with needle electrodes.

Methods: Thirty patients who underwent parotidectomy under FNM using adhesive surface and needle electrodes were included. Two pairs of adhesive surface electrodes and needle electrodes were used for FNM during parotid surgery. Mean amplitudes were collected after electrical facial nerve stimulation at 1 mA after specimen removal.

Results: The mean amplitude of the adhesive surface electrodes was 226.50 ± 118.44 μV (orbicularis oculi muscle) and 469.6 ± 306.06 μV (orbicularis oris muscle), respectively. The mean amplitude of the needle electrodes was 449.85 ± 248.10 μV (orbicularis oculi muscle) and 654.66 ± 395.71 μV (orbicularis oris muscle), respectively. The mean amplitude of the orbicularis oris muscle was significantly greater than that of the orbicularis oculi. The amplitude values measured in the orbicularis oculi muscle showed significant differences between the needle and skin electrodes.

Conclusions: Facial nerve monitoring (FNM) using adhesive surface electrodes is feasible in parotid surgery. Although the mean amplitude value of the surface electrode was relatively lower than that of the needle electrode, the surface electrode is considered a feasible and safe EMG recording device for FNM in parotid surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00405-022-07731-xDOI Listing

Publication Analysis

Top Keywords

facial nerve
24
adhesive surface
24
surface electrodes
24
parotid surgery
20
needle electrodes
16
orbicularis oculi
16
nerve monitoring
12
fnm parotid
12
oculi muscle
12
orbicularis oris
12

Similar Publications

Fungal cerebral aneurysms, particularly those resulting from direct invasion by fungal sinusitis, are rare and often fatal when involving the cavernous segment of the internal carotid artery (ICA). We present a case of a ruptured fungal ICA aneurysm caused by sinusitis, successfully treated with parent artery occlusion (PAO). In this case, an 80-year-old woman presented with right ptosis, facial pain, and cranial nerve III, IV, and VI palsies.

View Article and Find Full Text PDF

Facial nerve pathology: emerging strategies for regeneration and functional restoration.

J Mater Chem B

September 2025

Nebraska Translational Research Center (NTRC), Department of Growth and Development, College of Dentistry, University of Nebraska Medical Center, Joseph D. & Millie E. Williams Science Hall, 525 S 42nd St, Room No 3.0.010, Omaha, NE 68105-6040, USA.

Facial nerve injuries cause significant functional impairments, affect facial expressions, speech, and overall quality of life. This article explores advances in facial nerve regeneration, encompassing both conventional and emerging therapeutic strategies. The regenerative process involves Wallerian degeneration, axonal regrowth, and target muscle reinnervation, where the distal axon degrades and the proximal axon initiates sprouting to restore connectivity.

View Article and Find Full Text PDF

Enhancing submandibular gland resection: A retrospective study on the efficacy of the ORBEYE 3D exoscope.

Oral Maxillofac Surg

September 2025

Department of Otolaryngology, Head and Neck Surgery, Kansai Medical University, Shinmachi 2-5-1, Hirakata-city, Osaka, Japan.

Purpose: For submandibular gland resection, conventional surgery with the naked eye remains the standard. With its excellent automatic focus and high magnification, the ORBEYE 3D exoscope enables precise submandibular gland resection with less stress. Therefore, we aimed to examine the usefulness of the exoscope in submandibular gland resection.

View Article and Find Full Text PDF

Background: Intracranial solitary fibrous tumors (SFTs) are rare mesenchymal tumors often presenting with dural-based lesions. These tumors can exhibit aggressive characteristics with high recurrence rates and extracranial metastasis. While SFTs occasionally invade venous sinuses, cases where the tumor arises within the venous sinus are rare.

View Article and Find Full Text PDF

Free Gracilis Muscle Transfer in the Nerve-Depleted Patient: Reconsidering the Deep Temporal Nerve.

Laryngoscope

September 2025

Division of Facial Plastic and Reconstructive Surgery, Department of Otolaryngology/Head and Neck Surgery, University of North Carolina, North Carolina, USA.

The choice of a donor nerve to innervate a free gracilis muscle transfer (FMGT) is more complicated in patients with a history of previous facial paralysis surgeries. This case report describes the use of the deep temporal nerve to successfully power a FMGT in a nerve-depleted patient.

View Article and Find Full Text PDF