98%
921
2 minutes
20
Objective: To assess the feasibility of collecting, examining and reporting observational, real-world evidence regarding the novel use of the GORE EXCLUDER Iliac Branch Endoprosthesis (IBE) in conjunction with the GORE VIABAHN VBX Balloon Expandable Endoprosthesis (IBE+VBX stent graft).
Design: Multicentre retrospective cohort study.
Setting: Four real-world data sources were used: a national quality improvement registry, a statewide clinical research network, a regional quaternary health system and two tertiary academic medical centres.
Participants: In total, 30 patients with 37 IBE+VBX stent graft were identified. Of those, the mean age was 72±10.2 years and 90% were male. The cohort was 77% white, 10% black, 3% Hispanic and 10% other.
Main Outcome Measures: Outcome measures included: proportion of percutaneous vs open surgical access, intensive care admission, intensive care unit (ICU) length-of-stay (LOS), total LOS, postoperative complications, discharge disposition and 30-day mortality.
Results: The majority (89%) of cases were performed percutaneously, 5% required surgical exposure following failed percutaneous access and 6% required open surgical exposure outright. Nearly half (43%) required intensive care admission with a median ICU LOS of 1 day (range: 1-2). Median total LOS was 1 day (IQR: 1-2). There were zero postoperative myocardial infarctions, zero reported leg embolisations and no reported reinterventions. Access site complications were described in 1 of 28 patients, manifesting as a haematoma or pseudoaneurysm. Ultimately, 97% were discharged to home and one patient was discharged to a nursing home or rehabilitation facility. There were no 30-day perioperative deaths.
Conclusions: This project demonstrates the feasibility of identifying and integrating real-world evidence, as it pertains to an unapproved combination of endovascular devices (IBE+VBX stent graft), for short-term outcomes analysis. This new paradigm of evidence has potential to be used for device monitoring, submission to regulatory agencies, or consideration in indication expansions and approvals with further efforts to systematise data collection and transmission mechanisms.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9345049 | PMC |
http://dx.doi.org/10.1136/bmjsit-2021-000085 | DOI Listing |
BMJ Surg Interv Health Technol
July 2022
Department of Vascular Surgery, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire, USA.
Objective: To assess the feasibility of collecting, examining and reporting observational, real-world evidence regarding the novel use of the GORE EXCLUDER Iliac Branch Endoprosthesis (IBE) in conjunction with the GORE VIABAHN VBX Balloon Expandable Endoprosthesis (IBE+VBX stent graft).
Design: Multicentre retrospective cohort study.
Setting: Four real-world data sources were used: a national quality improvement registry, a statewide clinical research network, a regional quaternary health system and two tertiary academic medical centres.
J Vasc Surg
September 2022
Department of Vascular Surgery, Heart Vascular and Thoracic Institute, Cleveland Clinic, Cleveland, OH. Electronic address:
Objective: The Gore Excluder iliac branch endoprosthesis (IBE; W.L. Gore & Associates, Flagstaff, AZ) is the only iliac branch device approved in the United States to preserve blood flow to the external and internal iliac arteries (IIAs).
View Article and Find Full Text PDFJ Vasc Surg
May 2022
The University of Texas Health Science Center at Houston, McGovern Medical School, Houston, Tex. Electronic address:
Purpose: The purpose of this study was to compare outcomes of internal iliac artery (IIA) stenting using balloon-expandable (BESG) or self-expandable stent grafts (SESG) during endovascular repair of aortoiliac aneurysms with iliac branch endoprosthesis (IBE; W. L. Gore, Flagstaff, Ariz).
View Article and Find Full Text PDFAnn Vasc Surg
November 2021
Division of Vascular Surgery and Endovascular Therapy, Yale University School of Medicine, New Haven, CT; VA Connecticut Healthcare System, West Haven, CT. Electronic address:
Background: Endovascular treatment of complex common iliac artery (CIA) and internal iliac artery (IIA) aneurysms using iliac branch endoprostheses (IBE) has proven safe and effective. Instructions for use (IFU) require deployment of current IBE technology with the corresponding manufacturer's modular bifurcated aortic endograft. Concomitant aortoiliac occlusive disease, inadequate renal artery-iliac bifurcation length, and unfavorable aortic anatomy preclude on-label IBE deployment.
View Article and Find Full Text PDFJ Endovasc Ther
June 2021
Division of Vascular Surgery & Endovascular Therapy, University of Pennsylvania, Philadelphia, PA, USA.
Purpose: To describe a novel, entirely ipsilateral femoral technique for distal endograft extension using the Gore Iliac Branch Endoprosthesis.
Technique: Femoral arterial access is obtained on the side of the intended repair, and a 16F sheath is inserted over a stiff wire. A looped wire is used to pre-cannulate the internal gate of the IBE device prior to insertion, and the device is then positioned and deployed.