Article Abstract

Objectives: Photon-counting detector computed tomography (PCD-CT) is a promising new technique for CT imaging. The aim of the present study was the in vitro comparison of coil-related artifacts in PCD-CT and conventional energy-integrating detector CT (EID-CT) using a comparable standard brain imaging protocol before and after metal artifact reduction (MAR).

Methods: A nidus-shaped rubber latex, resembling an aneurysm of the cerebral arteries, was filled with neurovascular platinum coils and inserted into a brain imaging phantom. Image acquisition and reconstruction were repeatedly performed for PCD-CT and EID-CT (n = 10, respectively) using a standard brain imaging protocol. Moreover, linear interpolation MAR was performed for PCD-CT and EID-CT images. The degree of artifacts was analyzed quantitatively (standard deviation in a donut-shaped region of interest) and qualitatively (5-point scale analysis).

Results: Quantitative and qualitative analysis demonstrated a lower degree of metal artifacts in the EID-CT images compared to the total-energy PCD-CT images (e.g., 82.99 ± 7.89 Hounsfield units (HU) versus 90.35 ± 6.28 HU; p < 0.001) with no qualitative difference between the high-energy bin PCD-CT images and the EID-CT images (4.18 ± 0.37 and 3.70 ± 0.64; p = 0.575). After MAR, artifacts were more profoundly reduced in the PCD-CT images compared to the EID-CT images in both analyses (e.g., 2.35 ± 0.43 and 3.18 ± 0.34; p < 0.001).

Conclusion: PCD-CT in combination with MAR have the potential to provide an improved option for reduction of coil-related artifacts in cerebral imaging in this in vitro study.

Key Points: • Photon-counting detector CT produces more artifacts compared to energy-integrating detector CT without metal artifact reduction in cerebral in vitro imaging after neurovascular coil-embolization. • Spectral information of PCD-CT provides the potential for new post-processing techniques, since the coil-related artifacts were lower in PCD-CT images compared to EID-CT images after linear interpolation metal artifact reduction in this in vitro study.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9889475PMC
http://dx.doi.org/10.1007/s00330-022-09073-yDOI Listing

Publication Analysis

Top Keywords

eid-ct images
20
artifact reduction
16
brain imaging
16
pcd-ct images
16
photon-counting detector
12
energy-integrating detector
12
standard brain
12
imaging protocol
12
coil-related artifacts
12
metal artifact
12

Similar Publications

Background: Pulmonary manifestations of systemic diseases represent a complex and diagnostically challenging field. While pulmonary involvement in immunological and hematological disorders is well described, lung involvement in rare genetic and congenital systemic diseases-such as systemic storage diseases, neuromuscular disorders, and phacomatoses-is only gradually gaining increased attention in radiological diagnostics.

Results: The pulmonary component often manifests secondarily, frequently after years of progression and worsening of the underlying disease.

View Article and Find Full Text PDF

Background: Photon-counting detector computed tomography (PCD CT) offers higher dose efficiency than conventional energy-integrating detector CT (EID CT), which is particularly beneficial for children. Broad evidence is missing whether frequently acquired pediatric low-dose lung imaging can be further improved using PCD CT.

Objective: To compare radiation exposure, quantitative and qualitative image quality of pediatric low-dose chest PCD CT versus EID CT examinations.

View Article and Find Full Text PDF

Objectives: This study compared three-dimensional myocardial extracellular volume (ECV) quantification among single-phase and dual-phase photon-counting detector computed tomography (PCD-CT) and dual-phase energy-integrating detector computed tomography (EID-CT), using cardiac magnetic resonance (CMR) imaging as the reference.

Materials And Methods: This retrospective study included 80 patients who underwent both CMR and cardiac CT (40 patients underwent PCD-CT and 40 underwent EID-CT). Pearson correlation coefficients and intraclass correlation coefficients (ICCs) were used to evaluate the correlation and reliability of CT-ECV with CMR-ECV.

View Article and Find Full Text PDF

Background: This study aimed to evaluate the imaging performance and diagnostic value of a photon-counting detector (PCD) computed tomography (CT) compared to an energy-integrating detector (EID) and flat panel detector (FPD) for cadaveric wrist arthrographies.

Methods: Following ethics committee approval, ten cadaveric wrists were injected with diluted iodinated contrast agent. CT arthrographies using PCD-, EID-, and FPD-CT were performed.

View Article and Find Full Text PDF

Objective: To directly compare the image quality of photon-counting detector CT (PCD-CT) and conventional energy-integrating detector CT (EID-CT) for temporal bone imaging within the same patient cohort.

Methods: This retrospective study included seven patients who underwent both EID-CT and PCD-CT for clinical indications. To ensure a valid comparison of the imaging technologies, images of the non-operated, contralateral temporal bone were evaluated.

View Article and Find Full Text PDF