Article Abstract

MALDI mass spectrometry imaging (MALDI MSI) is a powerful analytical method for achieving 2D localization of compounds from thin sections of typically but not exclusively biological samples. The dynamically harmonized ICR cell (ParaCell) was recently introduced to achieve extreme spectral resolution capable of providing the isotopic fine structure of ions detected in complex samples. The latest improvement in the ICR technology also includes 2ω detection, which significantly reduces the transient time while preserving the nominal mass resolving power of the ICR cell. High-resolution MS images acquired on FT-ICR instruments equipped with 7T and 9.4T superconducting magnets and the dynamically harmonized ICR cell operating at suboptimal parameters suffered severely from the pixel-to-pixel shifting of / peaks due to space-charge effects. The resulting profile average mass spectra have depreciated mass measurement accuracy and mass resolving power under the instrument specifications that affect the confidence level of the identified ions. Here, we propose an analytical workflow based on the monitoring of the total ion current to restrain the pixel-to-pixel / shift. Adjustment of the laser parameters is proposed to maintain high spectral resolution and mass accuracy measurement within the instrument specifications during MSI analyses. The optimized method has been successfully employed in replicates to perform high-quality MALDI MS images at resolving power (FWHM) above 1,000,000 in the lipid mass range across the whole image for superconducting magnets of 7T and 9.4T using 1 and 2ω detection. Our data also compare favorably with MALDI MSI experiments performed on higher-magnetic-field superconducting magnets, including the 21T MALDI FT-ICR prototype instrument of the NHMFL group at Tallahassee, Florida.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9260710PMC
http://dx.doi.org/10.1021/acs.analchem.2c00754DOI Listing

Publication Analysis

Top Keywords

resolving power
16
icr cell
16
mass resolving
12
dynamically harmonized
12
harmonized icr
12
2ω detection
12
superconducting magnets
12
mass spectrometry
8
spectrometry imaging
8
mass
8

Similar Publications

The ion-optical properties of the second stability region () formed by the square wave shape potential with a duty cycle of 50% are studied as applied to the operation of a linear ion trap. The stability diagram is presented in detail, the stability parameters and , which determine the spectrum of ion oscillations, are calculated; the pseudopotential well-depth for this zone is given. The LIT acceptances for sinusoidal and rectangular wave forms are shown for comparison.

View Article and Find Full Text PDF

Objective: To explore the feasibility of establishing a mouse stellate ganglion (SG) regulation model through infrared polarized light (IPL) irradiation of the SG, and preliminarily evaluate its effects on SG function and related physiological indicators. Surgery, and IPL groups, with 8 mice in each group. A ZZIR-ID therapeutic device was used to directly irradiate bilateral SG regions of IPL group mice, with wavelength 980 nm, power density 1000 mW/cm2, 10 min per session (5 min per side), every other day for 6 times.

View Article and Find Full Text PDF

An electrostatic linear ion trap (ELIT) is used to trap ions between two ion mirrors with image current detection by central detection electrode. Transformation of the time-domain signal to the frequency-domain via Fourier transform (FT) yields an ion frequency spectrum that can be converted to a mass-to-charge scale. Injection of ions into an ELIT from an external ion source leads to a time-of-flight ion separation that ultimately determines the range of over which ions can be collected from a given ion injection step.

View Article and Find Full Text PDF

Metal-organic frameworks (MOFs) are transformative platforms for heterogeneous catalysis, but distinguishing atomically dispersed metal sites from subnanometric clusters remains a major challenge. This often demands the integration of multiple characterization techniques, many of which either lack the resolving power to distinguish active sites from their surrounding environments (e.g.

View Article and Find Full Text PDF

Bifunctional integration of indoor organic photovoltaics (OPVs) and photodetectors (OPDs) faces fundamental challenges because of incompatible interfacial thermodynamics: indoor OPVs require unimpeded charge extraction under low-light conditions (200-1000 lx), whereas OPDs require stringent suppression of noise current. Conventional hole transport layers (HTLs) fail to satisfy these opposing charge-dynamic requirements concurrently with commercial practicality (large-area uniformity, photostability, and cost-effective manufacturability). This study introduces benzene-phosphonic acid (BPA)-a minimalist self-assembled monolayer (SAM)-based HTL with a benzene core and phosphonic acid anchoring group-enabling cost-effective synthesis and excellent ITO interfacial properties such as energy alignment, uniform monolayer, and stability.

View Article and Find Full Text PDF