Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: Navigation is becoming more useful in percutaneous pedicle screw fixation (PPSF). The aim of this study was to compare the efficiency, fluoroscopic time, accuracy, and clinical outcomes of PPSF with a novel electromagnetic navigation (EMN) system for thoraco-lumbar (TL) fractures with those of PPSF with conventional C-arm fluoroscopic (CF) guidance.

Methods: A retrospective study was conducted. A total of 162 screws were implanted in 29 patients with the assistance of the EMN system (EMN group), and 220 screws were inserted in 40 patients by using CF guidance (CF group). The duration of surgery, placement time per screw, fluoroscopic time per screw, accuracy of pedicle screw placement, and clinical outcomes were compared between the two groups.

Results: The duration of surgery and placement time per screw in the EMN group were significantly lower than those in the CF group (P < 0.05). The fluoroscopic time per screw in the CF group was significantly longer than that in the EMN group (P < 0.05). The learning curve of PPSF in the EMN group was steeper than that in the CF group. The accuracy of pedicle screw placement in the EMN group was more precise than that in the CF group (P < 0.05). The VAS scores in the EMN group were significantly lower than those in the CF group at one-week postoperatively (P < 0.05).

Conclusion: Compared with PPSF by using conventional fluoroscopic guidance, PPSF with the aid of the EMN system can increase the efficiency and accuracy of pedicle screw placement and reduce the fluoroscopic time.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00586-021-06948-4DOI Listing

Publication Analysis

Top Keywords

pedicle screw
12
fluoroscopic time
12
time screw
12
electromagnetic navigation
8
percutaneous pedicle
8
screw fixation
8
thoraco-lumbar fractures
8
efficiency fluoroscopic
8
time accuracy
8
clinical outcomes
8

Similar Publications

Study Design: This is a retrospective single-center study.

Purpose: The purpose is to investigate the incidence of distal junctional kyphosis (DJK) when fused proximal to the stable sagittal vertebra (SSV) in adolescent idiopathic scoliosis (AIS) patients undergoing selective thoracic fusion.

Methods: We retrospectively reviewed a consecutive cohort of surgically treated AIS patients with Lenke 1-2 A/B curves between 2011 and 2022 with a minimum of 2 years of follow-up.

View Article and Find Full Text PDF

The cervicothoracic junction (CTJ) presents a surgical challenge due to its transitional nature from mobile to rigid segments. Therefore, the biomechanical characteristics of this transitional zone must be taken into consideration during instrumentation. This study aimed to determine the efficacy of the cervical pedicle screw placement (CPS) combined with 5.

View Article and Find Full Text PDF

Posterior lumbar screw fixation is the most common surgical method for lumbar disc herniation, but patients often face multiple complications postoperatively. The occurrence of screw track loosening can lead to fusion failure and even life-threatening screw track extrusion. However, there is currently a lack of animal models specifically targeting changes in the screw track following lumbar screw fixation.

View Article and Find Full Text PDF

Thoracic aortic injury as a complication of spinal surgery: A new case and systematic review (1991-2024).

J Vasc Surg Cases Innov Tech

December 2025

Faculdade de Ciências Médicas de Alagoas, Vascular and Endovascular Surgery Division, Alagoas, Alagoas, Brazil.

Background: Iatrogenic thoracic aortic injury (TAI) is a rare but well-recognized complication of spine surgery, lacking standardized treatment guidelines due to its rarity and variability of manifestations.

Methods: We present a new case of TAI successfully managed with endovascular repair and systematically reviewed 52 articles (1991-2024) reporting 64 cases, including demographics, surgical indications, injury patterns, and treatments.

Results: A 53-year-old man with a T7 fracture underwent posterior spinal instrumentation and developed chest pain due to a combination of impingement and screw penetration into the thoracic aorta and was treated with thoracic endovascular aortic repair (TEVAR) and removal of pedicle screws.

View Article and Find Full Text PDF