Statement Of Problem: Intraoral scanners are promising options for removable prosthodontics. However, analog aids, including occlusion rims, are still used, as a completely digital workflow is challenging and scientific evidence on the topic is scarce.
Purpose: The purpose of this in vitro study was to assess and compare the trueness and precision of scans obtained from a reference typodont of a completely edentulous maxilla by using an intraoral scanner (TRIOS 3 Pod; 3Shape A/S) with scans obtained by using a laboratory scanner (DScan 3; EGS S.R.L.) from both Type IV stone casts and polysulfide impressions.
Material And Methods: The polyurethane resin reference typodont was replicated from a clinical cast and was scanned with a metrological machine to obtain a reference scan. Ten digital casts were obtained by applying standardized scanning strategies to the reference typodont with the intraoral scanner. A device was created to make 10 consistent polysulfide impressions, and a scan of each impression was made with the laboratory scanner and then digitally reversed to obtain 10 digital reversed casts. Ten Type IV stone casts were poured and then scanned with the laboratory scanner to obtain 10 digital extraoral scanner casts. The scans in standard tessellation language (STL) format were imported into a dedicated software program, and the trueness and precision were calculated in μm. In addition to descriptive statistics (confidence interval 95%), 1-way ANOVA followed by the Bonferroni test or the Kruskal-Wallis and the Dunn tests were used to analyze differences among groups (α=.05).
Results: The trueness values (95% confidence interval) were digital intraoral scanner cast=48.7 (37.8-59.5), digital reversed cast=249.9 (121.3-378.5), and digital extraoral scanner cast=308.8 (186.6-430.9); significant differences were detected between digital intraoral scanner cast and digital reversed cast (P<.001) and between digital IOS casts and digital extraoral scanner cast (P<.001). The precision values (95% confidence interval) were digital intraoral scanner cast=46.7 (29.7-63.7), digital reversed cast=271.2 (94.6-447.8), and digital extraoral scanner cast=341.4 (175.5-507.3); significant differences were detected between digital intraoral scanner cast and digital reversed cast (P=.003) and between digital intraoral scanner cast and digital extraoral scanner cast (P=.001).
Conclusions: Directly scanning a solid typodont of a completely edentulous maxilla with the intraoral scanner produced better trueness and precision than scanning the polysulfide impressions or the stone casts with a laboratory scanner.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.prosdent.2020.07.018 | DOI Listing |
J Prosthodont Res
September 2025
Department of Health Sciences, School of Dentistry, University of Araraquara, UNIARA, Araraquara, Sao Paulo, Brazil.
Purpose: This in vitro study evaluated the adaptation and fracture resistance of screw- and cement-retained implant-supported restorations using lithium disilicate crowns cemented onto titanium bases and universal abutments of different heights.
Methods: Thirty crowns were designed, milled, and divided into three groups: TBA (titanium base abutment - 4.7-mm-height and 4.
J Dent Educ
September 2025
Department of Prosthodontics, Oral Science Research Center, Yonsei University Dental College, Seoul, South Korea.
Purpose: Crown preparation is a fundamental procedure in restorative dentistry. This study aimed to evaluate quantitative and color-coded assessment of tooth reduction using various 3D-printed tooth reduction guides in comparison to conventional guiding methods during crown preparation.
Methods: Twelve prosthodontic residents from a single prosthodontic graduate program (first year: n = 6; second year: n = 6) participated in this IRB-approved study (Yonsei University Dental Hospital IRB 2-2024-0026).
Korean J Orthod
September 2025
Private Practice, Brisbane, Australia.
Objective: This study aimed to investigate the effects of low-nicotine cigarette smoke (CS) inhalation on orthodontic tooth movement (OTM) in rats.
Methods: Forty 4-week-old male Wistar rats, were randomly divided into four groups: three experimental and one control. Group 1 (control group) had dental impressions taken at week 9.
J Prosthodont
September 2025
Department of Graphic Design and Engineering Projects, Faculty of Engineering Gipuzkoa, University of the Basque Country UPV/EHU, San Sebastian, Spain.
Purpose: The aim of this in vitro study was to assess whether the accuracy (trueness and precision) of full-arch implant digital impressions can be improved by replacing standard digital impressions (StDIs) with segmented digital impressions (SgDIs), obtained by superimposing short-span digital impressions using a geometric pattern.
Materials And Methods: The study was conducted using a commercial edentulous maxillary model with four implants-positioned at the sites of the canines and second molars-and four commercial intraoral scan bodies (ISBs) screwed to them. Ten StDIs and SgDIs of the model were obtained, and their accuracy was evaluated and compared in terms of trueness and precision, based on measurements of three distances and three angles between the ISBs.
J Adv Prosthodont
August 2025
Department of Prosthodontics, School of Dentistry, Kyungpook National University, Daegu, Republic of Korea.
Purpose: Designing restorations remains challenging because the process is time-consuming and requires operator skill and experience. This clinical study evaluated the fit accuracy of polymerized complete crowns fabricated using a web-based 3D generative artificial intelligence design (GAID) method compared to crowns fabricated using a conventional computer-aided design (CCAD) method.
Materials And Methods: Sixty-two patients requiring complete crowns in maxillary and mandibular premolars and molars were enrolled.